LigaSure Impact Curved, Large Jaw, Open Sealer/Divider, Nano-coated

K162047 · Covidien · GEI · Sep 15, 2016 · General, Plastic Surgery

Device Facts

Record IDK162047
Device NameLigaSure Impact Curved, Large Jaw, Open Sealer/Divider, Nano-coated
ApplicantCovidien
Product CodeGEI · General, Plastic Surgery
Decision DateSep 15, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, etc. The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.

Device Story

LigaSure Impact is a sterile, single-use, hand-held bipolar electrosurgical instrument for open surgery. Device connects to Covidien electrosurgical generators to deliver RF energy for vessel sealing and tissue division. User operates hand-actuated lever to close/latch jaws around target tissue (vessels/bundles up to 7mm); presses activation button to deliver RF energy for sealing; uses cutting trigger to actuate internal blade for division. Nano-coated jaws reduce tissue sticking. Used by surgeons in OR settings. Output is a sealed and divided tissue structure. Benefits include efficient hemostasis and reduced surgical time. Device contains no software.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing (biocompatibility per ISO 10993-1, electrical safety per ANSI AAMI ES 60601-1/IEC 60601-2-2, EMC per IEC 60601-1-2, mechanical/functional verification), ex-vivo vessel burst pressure testing on porcine tissue, thermal profile bench testing, and acute/chronic porcine animal studies evaluating seal quality and lateral thermal damage.

Technological Characteristics

Bipolar electrosurgical vessel sealing instrument. Materials: Nano-coated jaws. Energy: RF energy from compatible Covidien generators. Dimensions: Large jaw, curved. Connectivity: None (standalone). Sterilization: Sterile, single-use. Software: None.

Indications for Use

Indicated for open surgical procedures requiring ligation and division of vessels (arteries, veins, pulmonary veins up to 7mm), tissue bundles, and lymphatics. Applicable to general, urologic, vascular, thoracic, and gynecologic surgery. Contraindicated for tubal coagulation for sterilization.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 15, 2016 Covidien Mr. Celso Duran Specialist, Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301 Re: K162047 Trade/Device Name: LigaSure Impact Curved, Large Jaw, Open Sealer/divider, Nanocoated Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: August 31, 2016 Received: September 1, 2016 Dear Mr. Duran: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth {1}------------------------------------------------ in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Christopher J. Ronk-S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Page 1 of 1 #### 510(k) Number (if known) K162047 #### Device Name LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-coated Indications for Use (Describe) The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, etc. The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Covidien. The logo consists of a blue square with a white square in the center and a light blue vertical line. To the right of the square is the word "COVIDIEN" in a dark blue sans-serif font. #### 510(k) Summary Date summary prepared: 8/17/2016 ## 510(k) Submitter/Holder Covidien 5920 Longbow Drive Boulder, CO 80301 ## Contact Celso Duran Specialist, Regulatory Affairs Telephone: 303-530-6445 Fax: 303-530-6313 Email: celso.duran@medtronic.com #### Name of Device | Trade Name: | LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-<br>coated | |----------------------|----------------------------------------------------------------------------------------------------| | Catalog Numbers: | LF4418 | | Common Name: | Bipolar Vessel Sealing Device | | Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR<br>878.4400, Class II, GEI) | ## Predicate Device | Trade Name: | LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider | |-----------------|-------------------------------------------------------------| | Catalog Number: | LF4318 | | Common Name: | Bipolar Vessel Sealing Device | | 510(k) Number: | K123444 (cleared 1/30/2013), K141153 (cleared 8/14/2014) | | Manufacturer: | Covidien | | Recalls: | This device has not been subject to a design-related recall | No reference devices were used in this submission. # Device Description The LigaSure Impact™ Curved, Large Jaw. Open Sealer/Divider, Nano-coated (LF4418) is a sterile, single-use, coated, hand-held bipolar vessel sealing device designed for use with Covidien electrosurgical generators that include vessel sealing capabilities to ligate (seal) and divide (cut) vessels, tissue bundles, pulmonary arteries/veins, and lymphatics during open general surgical procedures (as indicated). A hand actuated lever allows the user to open or close the instrument jaws, and includes a latching mechanism that holds the jaws in the closed position during vessel sealing and cutting. The proposed devices do not contain software. # Indications for Use The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary arteries, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. {4}------------------------------------------------ The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures. # Comparison of Technological Characteristics with the Predicate Device The application of radio frequency (RF) energy is the principle for both the subject and predicate devices. A seal is created by application of RF energy to structures (vascular and lymphatics) interposed between the jaws of the instrument. A blade within the instrument is user actuated to divide tissue after the seal is created. At a high level, the subject and predicate devices are based on the following same technological elements: - Bilateral jaws – used to reach target structures (vascular and lymphatics) - Lever -- incorporates a latch mechanism to hold the jaws in the closed position - Activation button allows RF energy to be activated by the user ● - Cutting trigger allows user to divide (cut) sealed structures . The only difference between the subject and predicate devices is the addition of a non-stick coating to the jaws of the proposed device. # Performance Data The following performance data were provided in support of the substantial equivalence determination. # Biocompatibility The biocompatibility evaluation for the LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-coated device was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation Of Medical Devices - Part 1: Evaluation And Testing Within A Risk Management Process," as recognized by FDA. The battery of testing included the following tests: - Cytotoxicity - Sensitization - Irritation . - . Systemic Toxicity - o Hemolysis # Electrical Safety and Electromagnetic Compatibility (EMC) Electrical safety testing and EMC testing was conducted on the LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-coated device. The system complies with relevant clauses of the ANSI AAMI ES 60601-1 and IEC 60601-2-2 standards for electrical safety and IEC 60601-1-2 standard for EMC. # Mechanical / Functional Testing Mechanical, electrical, and functional testing was carried out to verify that the proposed device performed as expected. - Jaw force - Jaw gap - Shaft/Jaw Rotation ● - Device resistance, capacitance, and inductance ● - Lever latching/unlatching/opening force . # Ex-vivo Vessel Burst Pressure Ex-vivo burst pressure testing of excised fresh porcine renal, pulmonary arteries, and lymphatics was conducted on both the subject and predicate devices to demonstrate bipolar electrosurgical vessel sealing performance. {5}------------------------------------------------ ## Thermal Profile Bench thermal profile testing using porcine tissue was conducted to evaluate the thermal profile of the jaw and shaft of the subject device in comparison to the predicate. ## Acute Animal Study In the animal study conducted, two female porcine underwent various procedures to assess acute sealing performance and lateral thermal damage by the subject and predicate devices. The thermal safety of the new device was evaluated by macroscopic and histological evaluation of the tissue that was sealed and divided. These studies demonstrated that the subject devices are as safe and effective as the predicate device. ## Chronic Animal Study A chronic study on five animals to assess seal quality over the course of 21 days was conducted. All animals survived 21 days post-op and all seals maintained chronic hemostasis. ## Clinical Studies This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence. ## Summary Based on the preclinical performance as documented in the performance testing, the LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-coated device was found to have a safety and effectiveness profile that is similar to the predicate device. #### Conclusions The subject device has the same indications for use as the legally marketed predicate device. Furthermore, design change made to introduce the coated device does not raise different kinds of safety and effectiveness questions. Verification and validation data support substantial equivalence of the modified LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nanocoated devices to the legally marketed predicate device. 5-4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...