The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
Device Story
Sterile, single-use, nano-coated bipolar electrosurgical instrument; used in minimally invasive or open surgery. Input: RF energy from compatible Covidien generator. Operation: User clamps tissue between jaws via hand-actuated lever; activates RF energy via button to seal vessels/tissue; user-actuated trigger deploys blade to divide sealed tissue. Used by surgeons for vessel ligation, tissue grasping, and blunt dissection. Output: Sealed and divided tissue. Benefits: Hemostasis in surgical procedures; reduced thermal spread. No software component.
Clinical Evidence
No clinical data. Evidence includes bench performance testing (mechanical, electrical, functional, ex-vivo vessel burst pressure), acute animal study (sealing performance, lateral thermal spread), and chronic animal study (21-day seal quality/hemostasis). Usability engineering performed per IEC 62366-1.
Indicated for ligation and division of vessels (arteries, veins, pulmonary veins) up to 7 mm, tissue bundles, and lymphatics in general, urologic, vascular, thoracic, and gynecologic surgery. Contraindicated for tubal coagulation/sterilization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-coated (LF4418) (K162047)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA acronym and the agency's name are in blue.
June 21, 2018
Covidien Mr. James Davis Senior Specialist, Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301
Re: K181085
Trade/Device Name: LigaSure Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 24, 2018 Received: April 25, 2018
Dear Mr. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{2}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181085
#### Device Name
LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T)
#### Indications for Use (Describe)
The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc.
The LigaSure system has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the word "COVIDIEN" in a sans-serif font. To the left of the word is a blue square with a white rectangle in the middle. The square is divided into three vertical sections, with the middle section being a lighter shade of blue.
510(k) Summarv Date summary prepared: April 24, 2018
# 510(k) Submitter/Holder
Covidien, Ilc. 5920 Longbow Drive Boulder, CO 80301
#### Contact
James R. Davis Sr. Regulatory Affairs Specialist Telephone: 303-581-6964 Fax: 303-530-6313 Email: james.r.davis(@medtronic.com
# Name of Device
| Trade Name: | LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing,<br>Nano-coated |
|----------------------|----------------------------------------------------------------------------------------------------|
| Catalog Numbers: | LF1930T |
| Common Name: | Bipolar Vessel Sealing Device |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories<br>(21 CFR 878.4400, Class II, GEI) |
| Predicate Devices: | Primary |
| Trade Name: | LigaSure Impact™ Curved, Large Jaw, Open Sealer/Divider, Nano-<br>coated (LF4418) |
| Common Name: | Bipolar Vessel Sealing Device |
| Catalog Number: | LF4418 |
| 510(k) Number: | K162047 (cleared 9/15/16) |
| Manufacturer: | Covidien |
| Recalls: | This device has not been subject to a design-related recall |
| | Secondary |
| Trade Name: | LigaSure™ Maryland Jaw Sealer/Divider One-step Sealing, Nano-coate |
| Common Name: | Bipolar Vessel Sealing Device |
| Catalog Number: | LF1923, LF1937, LF1944 |
| 510(k) Number: | K133338 (cleared 12/20/2013), K141153 (cleared 8/14/2014), K170869<br>(cleared 04/21/2017) |
| Manufacturer: | Covidien |
| Recalls: | This device has not been subject to a design-related recall |
No reference devices were used in this submission.
# Device Description
The LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) device is a sterile, single-use, coated, hand-held bipolar vessel sealing device designed for use with a Covidien electrosurgical generator that include vessel sealing capabilities
{5}------------------------------------------------
to ligate (seal) and divide (cut) vessels (arteries, veins, pulmonary arteries, pulmonary veins), tissue bundles, and lymphatics clamped between the jaws using radio frequency (RF) energy. The LF1930T subject can also grasp tissue and perform blunt dissection during general surgical procedures (as indicated).
A hand actuated lever allows the user to open or close the instrument jaws during vessel sealing and cutting, with a single activation button incorporated into the body of the handle. The proposed device has a 30cm length shaft and does not contain software.
# Indications for use:
The LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries, veins, pulmonary arteries, pulmonary veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as urologic, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc.
The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
# Comparison of Technological Characteristics to the Predicate Devices
The application of radio frequency (RF) energy to target tissue is the fundamental technology for both the subject and predicate devices. A seal is created by application of RF energy to structures (vasculature, lymphatics, and tissue bundles) interposed between the jaws of the instrument. A blade within the instrument is user actuated to divide tissue after the seal is created.
At a high level, the subject and both the primary/secondary predicate devices are based on the following same technological elements:
- . Lever - closes the jaw to grasp the tissue
- Jaws Used to grasp tissue and to deliver RF energy to seal target structures ●
- Activation Button - Allows RF energy to be activated by the user
- Cutting Trigger Allows the user to divide (cut) sealed structures ●
- Device Materials - Materials are identical to the LF19XX secondary predicate devices only
The following technological differences exist between the subject and both the primary/secondary predicate devices:
- . Shaft Length – Provides the jaw assembly access to the surgical site
- Jaw Assembly Specifications Minor restrictions of the jaw assembly specification range ● within the specification range of the LF19XX secondary predicate devices
- . Device Identifiers - Changes to product identifiers on the device, packaging, and labeling
These changes in conjunction with the performance data outlined below support the changes to the indications for use of the subject device and have demonstrated to be substantially equivalent to the predicate devices.
#### Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
{6}------------------------------------------------
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety testing and EMC testing were conducted on the LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) or equivalent device as appropriate. The system complies with relevant portions of the IEC 60601-1 and IEC 60601-2-2 standards for electrical safety and IEC 60601-1-2 standard for EMC.
### Bench Performance Testing
Mechanical, electrical, functional, and ex-vivo vessel burst pressure testing was carried out on validated and representative conditions as appropriate to verify that the new device performs as expected. The subject device conforms to the requirements defined in the related design inputs and subsequent product specifications.
# Acute Animal Study
In the acute animal study conducted, subjects underwent procedures to assess acute sealing performance and lateral thermal spread by the subject and predicate devices as appropriate. These studies demonstrated that the subject device is as safe and effective as the predicate devices.
# Chronic Animal Study
A chronic study was conducted to assess seal quality over the course of 21 days. All animals enrolled in the study survived 21 days post-op and all seals maintained chronic hemostasis.
# Human Factors and Usability
The usability engineering process applied to the subject device was in compliance with the requirements of IEC 62366-1 "Medical devices-Application of usability engineering to medical devices". The process included analysis of user needs and potential use errors followed by testing to demonstrate that representative users can use the subject device in a safe and effective manner.
#### Clinical Studies
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
#### Summarv
Based on the pre-clinical performance as documented in the performance testing, the LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated (LF1930T) subject device was found to have a safety and effectiveness profile that is similar to the predicate devices.
#### Conclusions
The subject device has the same intended use as the predicate devices and the design changes made to the subject device does not raise different kinds of safety and effectiveness questions. The comparison of device characteristics and review of the performance verification/validation data demonstrate substantial equivalence of the LigaSure™ Maryland Jaw Thoracic Sealer/Divider One-step Sealing, Nano-coated subject device to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.