Endopath Electrosurgery Probe Plus II

K180031 · Ethicon Endo-Surgery, LLC · GEI · May 7, 2018 · General, Plastic Surgery

Device Facts

Record IDK180031
Device NameEndopath Electrosurgery Probe Plus II
ApplicantEthicon Endo-Surgery, LLC
Product CodeGEI · General, Plastic Surgery
Decision DateMay 7, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve.

Device Story

Monopolar electrosurgical instrument; facilitates tissue dissection, coagulation, irrigation, and fluid evacuation during minimally invasive surgery. Device features pencil-grip handle; connects to standard monopolar electrosurgical unit; operates through 5mm trocar sleeve. Surgeon controls suction, irrigation, and electrosurgical energy delivery via handle. Output includes thermal coagulation of tissue and mechanical fluid management. Benefits include integrated functionality for dissection and irrigation in single device, reducing instrument exchanges during surgery.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including flow rate, impedance, and mechanical force evaluations, alongside electrical safety (IEC 60601-1) and EMC (IEC 60601-1-2) compliance testing.

Technological Characteristics

Monopolar electrosurgical probe; pencil-grip handle configuration. Uses standard suction/irrigation modalities. Sterilized via gamma radiation. Complies with IEC 60601-1:2005+A1:2012 (electrical safety) and IEC 60601-1-2:2007 (EMC).

Indications for Use

Indicated for minimally invasive surgical procedures requiring tissue dissection, coagulation, irrigation, and fluid evacuation. Contraindicated for contraceptive coagulation of fallopian tissue (though permitted for hemostasis post-transection) and when minimally invasive techniques are contraindicated.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 7, 2018 Ethicon Endo-Surgery, LLC David Locke Regulatory Affairs Manager 4545 Creek Road Cincinnati, Ohio 45242 Re: K180031 Trade/Device Name: Endopath Electrosurgery Probe Plus II, Pencil Grip Handle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 5, 2018 Received: April 6, 2018 Dear David Locke: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely. Jennifer R. Stevenson For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) #### K180031 Device Name ENDOPATH Electrosurgery Probe Plus II (Pencil) Indications for Use (Describe) Indications The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve. Contraindications · These instruments are not intended for contraceptive coagulation of fallopian tissue but may be used to achieve hemostasis following transection of the fallopian tube. · These instruments are not intended for use when minimally invasive techniques are contraindicated. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary ## Company Ethicon Endo-Surgery 475 Calle C Guaynabo, PR 00969 ## Contact David Locke, DRSc, MS, RAC Manager, Regulatory Affairs Ethicon Endo-Surgery, LLC. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-1281 Fax: (513) 337- 4366 Email: dlocke1 @its.jnj.com Date Prepared January 2, 2018 ### Device Name Trade Name: Evacuation/Irrigation/Electrosurgical Device Classification Name: Laparoscope, General & Plastic Surgery Common Name: Endopath® Electrosurgery Probe Plus II, Pencil Grip Handle Catalog Number: EPH04 Device Class Class II Device Panel General & Plastic Surgery Product Code GEI # Predicate Device Endopath® Electrosurgery Probe Plus II cleared under K160128 {4}------------------------------------------------ ## Device Description The Endopath® Electrosurgery Probe Plus II device with the pencil handle configuration (product code EPH04) works with a standard monopolar unit and has applications for minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve. Previously cleared shafts (K160128) EPS01 through EPS07 all attach to the EPH04 pencil handle in the same manner as the previously cleared pistol handle, and are for use through a 5mm diameter surgical trocar or a larger trocar with a 5mm reducer. ## Indications for Use The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation and fluid evacuation through a common trocar sleeve. ### Contraindications - These instruments are not intended for contraceptive coagulation of fallopian tissue but ● may be used to achieve hemostasis following transection of the fallopian tube. - These instruments are not intended for use when minimally invasive techniques are . contraindicated. ### Technological Characteristics The subject ENDOPATH Electrosurgery Probe Plus II Pencil device (EPH04) and predicate ENDOPATH Electrosurgery Probe Plus II Pistol device (EPH02) use the same monopolar technology to perform their intended use. The subject and predicate devices also use the same standard suction and irrigation modalities to perform their intended use. The subject ENDOPATH Electrosurgery Probe Plus II Pencil device is similar to the predicate ENDOPATH Electrosurgery Probe Plus II Pistol device with respect to the primary modes of action which include suction, irrigation and hemostasis. As compared with the predicate devices, the subject device takes on a straight pencil configuration as to where the cleared predicate device takes on a pistol configuration. #### Performance Data Bench testing was conducted to demonstrate that the ENDOPATH Electrosurgery Probe Plus II Pencil device (product code EPH04) performed as intended. #### Sterilization The subject device and the predicate device are sterilized via gamma radiation; both devices are sterilized to the same sterility assurance level. #### Biocompatibility Biocompatibility was not required for this submission as no new materials were introduced on this device. {5}------------------------------------------------ # EMC Electrical safety and EMC testing were conducted on the ENDOPATH Electrosurgery Probe Plus II Pencil device; the system complies with IEC 60601-1-2:2007 for electromagnetic compatibility and IEC IEC 60601-2-2:2009 & IEC 60601-1:2005+A1:2012 for electrical safety. ## Bench Flow rate of suction and irrigation, impedance, minimum distal retraction force and sheath extension force were evaluated for the ENDOPATH Electrosurgery Probe Plus II Pencil device to support substantial equivalent to the predicate device. Data generated from these tests met the predetermined acceptance criteria. ### Animal This premarket notification does not rely on preclinical animal testing to demonstrate substantial equivalence. ### Clinical This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence. #### Conclusion The results of the bench testing demonstrate that the Endopath® Electrosurgery Proble Plus II Pencil device (product code EPH04) is as safe and effective and performs as well as the identified legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...