VALLEYLAB SUCTION COAGULATOR, MODEL E3305

K091223 · Covidien LP · GEI · Sep 11, 2009 · General, Plastic Surgery

Device Facts

Record IDK091223
Device NameVALLEYLAB SUCTION COAGULATOR, MODEL E3305
ApplicantCovidien LP
Product CodeGEI · General, Plastic Surgery
Decision DateSep 11, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Suction Coagulator is a single use device intended for use in surgical procedures (such as general and ENT procedures) where the coagulation of tissue and suction of fluids are desired.

Device Story

Hand-held, single-use electrosurgical device; provides monopolar tissue coagulation and fluid aspiration. Components: metal insulated French-sized tube, switching mechanism, molded plastic body, cord with plug. Operated by surgeon in OR; coagulation activated via handle-mounted plastic button; suction activated by covering handle-mounted vent hole. Connects to external electrosurgical generators. Coagulation occurs at uninsulated distal end of suction tube. Facilitates simultaneous tissue management and fluid removal during surgery.

Clinical Evidence

No clinical testing conducted. Safety and effectiveness supported by bench testing and verification to IEC 60601-1 and IEC 60601-2-2 standards, and historical literature regarding use in ENT procedures.

Technological Characteristics

Disposable, hand-held electrosurgical device. Materials: metal insulated tube, molded plastic body. Energy: monopolar electrosurgical current from external generator. Connectivity: wired connection to generator. Sterilization: supplied sterile. Standards: IEC 60601-1, IEC 60601-2-2.

Indications for Use

Indicated for patients undergoing surgical procedures, including general and ENT surgeries, requiring tissue coagulation and fluid aspiration.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ #### 510/k) Summary or 510(k) Statement 510(k) Summary Submitter Information: 1. Covidien LP Formerly known as Valleylab, a Division of Tyco Healthcare Group LP 5920 Longbow Drive Boulder, CO Contact: Philip E. Ake Senior Regulatory Associate Phone: 303-581-6934 Fax: 303-516-8516 Email: Philip.Ake@Covidien.com 2. Name of Device Trade name: Valleylab Suction Coagulators Common/Classification name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400) 3. Predicate Devices The Valleylab Suction Coagulators are substantially equivalent to the following legally marketed medical devices: - Valleylab Lectrovac (K791752) . - Megadyne Suction Coagulator (K072559) . Device Description 4. The Valleylab Suction Coagulators are hand held electrosurgical devices used for the coagulation of tissue and aspiration of fluids during electrosurgical procedures. The devices are activated by a plastic button with a spring contact located in the handle of the device. The suction is activated by the surgeon placing a finger over the suction vent hole also in the handle. #### 5. Intended Use The Suction Coagulator is a single use device intended for use in surgical procedures (such as general and ENT procedures) where the coagulation of tissue and suction of fluids are desired. Summary of Technology Characteristics 6. {2}------------------------------------------------ The Suction Coagulators are a disposable, hand held device used in electrosurgical procedures. It provides control of both monopolar coagulation of tissue and suction of fluids in a single handle. The coagulation mode of operation is activated by a depressing a plastic button located on the top of the body. The activation of suction is activated by the surgeon's finger covering the suction port. Coagulation takes place at the uninsulated end of the suction tube. The Suction coagulator consists of four main elements: -a metal, insulated, French sized tube - -a switching mechanism - -a two piece molded plastic body - -a cord with plug The Suction Coagulator is supplied sterile and can be used with various electrosurgical generators. - 7. Summary Non-clinical testing Bench Testing and verification testing was conducted to ensure proper device function. This included testing to the relevant electrical standards (IEC 60601-1 and IEC 60601-2-2). Summary of Clinical testing 8. No clinical testing was conducted. The use of Electrosurgical Suction Coagulators has been documented in the published literature and indicates safe and effective use in ENT procedures. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem. SEP 1 1 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 Covidien LP % Mr. Philip E. Ake Senior Regulatory Associate 5920 Longbow Drive Boulder, Colorado 80301 Re: K091223 Trade Name: Suction Coagulator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 27, 2009 Received: July 29, 2009 Dear Mr. Ake: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. Philip E. Ake forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fdagov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark A Wilkerson Mark N. Melkerson Director . Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): __K091223_ Device Name: Suction Coagulator Indications for Use: The Suction Coagulator is a single use device intended for use in surgical procedures (such as general and ENT procedures) where the coagulation of tissue and suction of fluids are desired. Over-The-Counter Use Prescription Use _____________________________________________________________________________________________________________________________________________________________ 11cscription osc _____________________________________________________________________________________________________________________________________________________________ D) C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K091223
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...