K132743 · Maquet Cardiovascular, LLC · GEI · Dec 3, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132743
Device Name
HARVESTING CANNULA
Applicant
Maquet Cardiovascular, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 3, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The VASOVIEW System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or the radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels and other tissues of the chest wall.
Device Story
Harvesting Cannula is a multi-lumen surgical instrument used in minimally invasive endoscopic vessel harvesting. Device features four lumens housing an endoscope, C-Ring, distal lens washer tube, and a VASOVIEW Harvesting Tool. Physician operates the device to perform blunt dissection of tissue and vessels; harvesting tool enables cutting and coagulation of vessel branches. C-Ring slider on handle controls C-Ring retraction and distal lens washing. Device provides visualization and mechanical access for surgeons during arterial bypass procedures. Benefits include reduced surgical trauma via minimally invasive approach compared to open surgery.
Clinical Evidence
Bench testing only. In-vitro performance and shelf-life testing demonstrated that the functionality and performance characteristics of the Harvesting Cannula are comparable to the currently marketed predicate device.
Technological Characteristics
Four-lumen cannula design; includes endoscope, C-Ring, lens washer, and harvesting tool ports. Mechanical actuation via C-Ring slider. Materials and sterilization processes are identical to the predicate device. Class II electrosurgical accessory (Product Code GEI).
Indications for Use
Indicated for patients undergoing endoscopic surgery for arterial bypass, including coronary artery bypass grafting (CABG) and peripheral artery bypass. Used for blunt dissection of blood vessels (saphenous vein, radial artery), ducts, and other structures in extremities, extraperitoneal, subcutaneous, and thoracic spaces.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Guidant VasoView HemoPro Endoscopic Vessel Harvesting System (K052274)
Related Devices
K041981 — GUIDANT VASOVIEW 6 HARVESTING CANNULA, MODEL VH-2000; GUIDANT VASOVIEW 6 ACCESSORY PACK, MODEL VH-2004 · Guidant Corporation · Aug 20, 2004
K091733 — VASOVIEW 6 PRO ENDOSCOPIC VESSEL HARVESTING SYSTEM, MODEL VH-2500 · Maquet Cardiovascular, LLC · Aug 28, 2009
K233879 — Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (VH-6000), Vasoview Hemopro 3 Endoscopic Vessel Harvesting System with Vasoshield (VH-6001), Vasoview Hemopro 3 Power Adapter (VH-6020) · Maquet Cardiovascular, LLC · Feb 27, 2024
{0}------------------------------------------------
# K132743
## Harvesting Cannula
# 510(k) Summary
# Prepared in accordance with 21 CFR Part 807.92
| 510(k) Number: | KXXXXXX K132743 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 30 August 2013 |
| Device Owner: | MAQUET Cardiovascular LLC<br>45 Barbour Pond Drive<br>Wayne, New Jersey 07470 |
| Contact Personnel: | Mark Dinger |
| Title: | Regulatory Affairs Specialist II |
| Email: | mark.dinger@maquet.com |
| Phone: | 973-709-7691 |
| | Fax: 973-909-9954 |
| Trade Name: | Harvesting Cannula |
| Device Generic Name: | Harvesting Cannula |
| Classification: | According to 21 CFR 878.4400 of the Federal Food, Drug<br>and Cosmetic Act, the device classification is Class II,<br>Product code GEI. |
| Predicate Device: | K052274 Guidant VasoView HemoPro Endoscopic Vessel<br>Harvesting System (SE: 21 September 2005) |
| Device Description: | The Harvesting Cannula has four lumens which house the<br>Endoscope, C-Ring, distal lens washer tube and<br>VASOVIEW Harvesting Tool for cutting and sealing of<br>vessel branches. The C-Ring with the built in distal lens<br>washer is independently controlled by a C-Ring Slider on<br>the handle of the Harvesting Cannula. The C-Ring retracts<br>the vessel and washes the distal tip of the Endoscope. The<br>Harvesting Tool can be inserted, removed, rotated,<br>extended, and retracted from the main Harvesting Cannula<br>through the Tool Adapter Port. |
DEC 0-3 2013
{1}------------------------------------------------
| Indications for Use: | The VASOVIEW System is indicated for use in minimally<br>invasive surgery allowing access for vessel harvesting, and<br>is primarily indicated for patients undergoing endoscopic<br>surgery for arterial bypass. It is indicated for cutting tissue<br>and controlling bleeding through coagulation, and for<br>patients requiring blunt dissection of tissue including<br>dissection of blood vessels, dissection of blood vessels of<br>the extremities, dissection of ducts and other structures in<br>the extraperitoneal or subcutaneous extremity and thoracic<br>space. Extremity procedures include tissue<br>dissection/vessel harvesting along the saphenous vein for<br>use in coronary artery bypass grafting and peripheral artery<br>bypass or the radial artery for use in coronary artery bypass<br>grafting. Thoracoscopic procedures include exposure and<br>dissection of structures external to the parietal pleura,<br>including nerves, blood vessels and other tissues of the<br>chest wall. |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | The Proposed Harvesting Cannula and the predicate<br>devices have the following similarities:<br>the same intended use, the same operating principles, same labeling, incorporates the same basic design, sterilized using the same materials and processes, same materials, has same packaging. The Proposed Harvesting Cannula and the predicate<br>devices have the following differences:<br>C-Ring Tube – Addition of a rib to the C-Ring Tube<br>as well as tube length change (to accommodate for<br>the C-Ring Tube Rib design). Blunt Tip – C-ring tube channel revision. Scope Wash Tube – Scope Wash Tube length<br>changed (to accommodate for new C-Ring Tube<br>Rib design). This difference is not considered a technological difference<br>and is substantially equivalent to the predicate devices. |
| Safety and<br>Performance: | MAQUET Cardiovascular's development process required<br>that the following activities be completed during the<br>development of the Harvesting Cannula: |
| | • Performance testing<br>• Shelf life testing |
| | The results of the in-vitro tests conducted demonstrate that<br>the functionality and performance characteristics of the<br>device are comparable to the currently marketed Harvesting<br>Cannula. |
| Conclusion: | Based upon the information submitted in this Traditional<br>510(k) premarket notification, MAQUET's Harvesting<br>Cannula is substantially equivalent to the currently<br>marketed Harvesting Cannula. The Harvesting Cannula is<br>similar to the predicate devices in the intended use and the<br>fundamental scientific technology of the device. The design<br>verification and validation testing established that the<br>Harvesting Cannula is substantially equivalent as the<br>predicate device. |
{2}------------------------------------------------
.
. . .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and is relatively small compared to the caduceus symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
MAQUET Cardiovascular LLC Mr. Mark Dinger Regulatory Affairs Specialist II 45 Barbour Pond Drive Wayne, New Jersey 07470
December 3, 2013
Re: K132743
Trade/Device Name: Harvesting Cannula Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 27. 2013 Received: September 30, 2013
Dear Mr. Dinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated on the May 28. 1976. the cnactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The rou may : merels provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has into a and regulations administered by other Federal agencies. You must or any rederal states and the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Mark Dinger
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Joshua C. Nipper -S
For
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
### 510(k) Number (if known) K132743
Device Name Harvesting Cannula
#### Indications for Use (Describe)
The VASOVIEW System is indicated for use in minimally invesive surgery allowing access for vessel harvesting, and is primerily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting bleeding through coagulation, and for patients requiring blunt dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extraperity and theracies pace. Extremity procedures include tissection/vessel harvesting along the saphencus vein for use in coronary artery bypass grafting and perioteral artery bypass or the radial artery for use in coronary artery bypass grafting. Thorecossipic procedures include exposure and dissection of structures external to the parietal pleases and other tissues of the chest wall.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Parl 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY | |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | |
| Long H. Chen A | for BSA |
| (Division Sign-off)<br>Division of Surgical Devices | 510(k) Number: K132743 |
| | <div style="font-size: 8px;">Digitally signed by Long H. Chen - A</div> <div style="font-size: 8px;">DN: c=US, o=U.S. Government, ou=HHS,</div> <div style="font-size: 8px;">ou=FDA, ou=People, cn=Long H. Chen - A</div> <div style="font-size: 8px;">0.9.2342.19200300.100.1.1=1300366056</div> <div style="font-size: 8px;">Date: 2013.12.02 08:44:22 -05'00'</div> |
| FORM FDA 3881 (9/13) | Page 1 of 2 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.