LOGIC LAPAROSCOPIC INSTRUMENT SYSTEM

K111986 · Surgical Innovations Plc · GEI · Jan 26, 2012 · General, Plastic Surgery

Device Facts

Record IDK111986
Device NameLOGIC LAPAROSCOPIC INSTRUMENT SYSTEM
ApplicantSurgical Innovations Plc
Product CodeGEI · General, Plastic Surgery
Decision DateJan 26, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Logic Laparoscopic Instrument System, Electrosurgical, Cutting & Coagulation & Accessories is designed to dissect, grasp, manipulate and/or cauterise various tissues during manual operations in minimally invasive surgery and general endoscopic/laparoscopic procedures.

Device Story

Logic Laparoscopic Instrument System is a manual surgical instrument set for minimally invasive surgery. Device functions include tissue dissection, grasping, manipulation, and cauterization. Operated by surgeons in clinical/OR settings. System provides mechanical tissue interaction and electrosurgical coagulation capabilities. Benefits include facilitating surgical access and tissue management during endoscopic procedures.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Electrosurgical cutting and coagulation device. Class II, Product Code GEI. Manual laparoscopic instrument system.

Indications for Use

Indicated for use in patients undergoing minimally invasive surgery and general endoscopic/laparoscopic procedures requiring tissue dissection, grasping, manipulation, and cauterization.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Surgical Innovations PLC % Mr. Paul Birtles Clayton Wood House 6 Clayton Wood Bank Leeds. West Yorkshire United Kingdom LS16 6ZQ JAN 2 6 2012 Re: K111986 Trade/Device Name: Logic Laparoscopic Instrument System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 6, 2012 Received: January 10, 2012 Dear Mr. Birtles: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ ## Page 2 – Mr. Paul Birtles or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address · http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Attachment 2 ## Indications for Use Statement K11986 510(k) Number Logic Laparoscopic Instrument System, Electrosurgical, Cutting & Device Name Coagulation & Accessories Indications for Use The Logic Laparoscopic Instrument System, Electrosurgical, Cutting & Coagulation & Accessories is designed to dissect, grasp, manipulate and/or cauterise various tissues during manual operations in minimally invasive surgery and general endoscopic/laparoscopic procedures. Prescription Use _X (Per 21 CFR 801.109) OR Over-The-Counter Use_ PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of DFRH, Office of Device Evaluation (ODE) signature (Division ign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K111986
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...