HPR45i

K220725 · F Care Systems USA, LLC · GEI · Mar 20, 2023 · General, Plastic Surgery

Device Facts

Record IDK220725
Device NameHPR45i
ApplicantF Care Systems USA, LLC
Product CodeGEI · General, Plastic Surgery
Decision DateMar 20, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

HPR45i electrosurgical electrode is intended to conduct radio frequency (RF) current for coagulation from the RF electrosurgical generator to target soft tissue in a broad range of surgical procedures. The device is a prescription use (Rx) device.

Device Story

Monopolar electrosurgical electrode; conducts RF current from generator to soft tissue for hemostasis/coagulation. Components: handle, output cable, stainless-steel (AISI316L) tip, PTFE insulation. Used in hospital/clinical settings by licensed physicians/specialists. Device connects to compatible RF generator (MedRF4000); RF signal converts to heat upon tissue contact to stop bleeding. No power regulation, diagnostic, or monitoring functions. Disposable, single-use, sterile (EO). Benefits patient by providing controlled coagulation during surgery.

Clinical Evidence

No clinical data submitted. Substantial equivalence supported by non-clinical bench testing, including ESU testing, electrical safety, electromagnetic compatibility, and thermal effect on tissue, in accordance with FDA guidance for electrosurgical devices.

Technological Characteristics

Monopolar electrosurgical electrode. Materials: AISI316L stainless steel tip, PTFE insulation. Energy: RF (4MHz). Max power: 25W. Max voltage: 800V. Form factor: disposable handle/cable/tip assembly. Sterilization: Ethylene Oxide (ISO 11135). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 10993 series.

Indications for Use

Indicated for coagulation of soft tissue in surgical procedures. Prescription use only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 20, 2023 F Care Systems USA LLC % Shilpa Gampa Delivery Manager and US Agent-MDV Freyr, Inc. 150 College Rd W, #102 Princeton, New Jersey 08540 Re: K220725 Trade/Device Name: HPR45i Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 18, 2023 Received: January 20, 2023 Dear Shilpa Gampa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, {1}------------------------------------------------ including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH and Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark Digitally signed by Mark Trumbore -S Trumbore -S Date: 2023.03.20 12:13:51 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K220725 Device Name HPR45i Indications for Use (Describe) HPR45i electrosurgical electrode is intended to conduct radio frequency (RF) current for coagulation from the RF electrosurgical generator to target soft issue in a broad range of surgical procedures. The device is a prescription use (Rx) device. Type of Use (Select one or both, as applicable) | <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |------------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary This 510(K) summary of safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: March 14, 2023 ## 1. Submitter Information: | Official<br>Correspondent (US<br>Agent): | Shilpa Gampa<br>Freyr Inc.<br>150 College Rd W #102, Princeton, NJ 08540 | |------------------------------------------|------------------------------------------------------------------------------| | Phone: | +1 908-483-7958 Ext 1780<br>+1 760-583-4401 | | E-mail: | usagent@freyrsolutions.com | | Legal<br>Manufacturer: | F Care Systems USA LLC<br>11098 Biscayne Blvd<br>Suite 301<br>Miami FL 33161 | | Phone: | 786 288 0740 | | Contact Person: | Patrick Danciu | | E-mail: | pdanciu@fcaresystems.us | | Fax Number: | +32 3 451 51 39 | | Date Prepared: | 03-March-2022 | # 2. Device Identification: | Device Trade Name: | HPR45i | |----------------------|-------------------------------------------------------------------| | Common Name: | Electrosurgical electrode | | Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories | ## Table 1 Device Identification | Regulation Number | Product Code | Device Class | |-------------------|--------------|--------------| | 878.4400 | GEI | Class II | {4}------------------------------------------------ ### 3. Legally Marketed Equivalent Device: | 510(k) Number | Device Name | Manufacturer | |---------------|----------------------------------------|-----------------------------------| | K081791 | E-Z Clean Electrosurgical<br>Electrode | Megadyne Medical<br>Products, Inc | #### Table 2 Predicate Device ## 4. Device Description HPR45i is a monopolar electrosurgical electrode intended to remove tissue and control bleeding using radio frequency electrical current. It is a prescription use device which is intended to be used by a licensed doctors or specialist in the healthcare facility/ hospital. The HPR45i is supplied sterile (Sterilized by EO Sterilization) and is not intended to be reused. It consists of 3 different parts such as HPR45i handle, output cable and stainless-steel tube. The stainless-steel tube is insulated over most of its exposed length, and it is connected to the stainless steel (AISI316L) tip at the end. The material of insulation is PTFE. The stainless steel (AISI316L) tip and PTFE insulation are the direct patient contacting parts of the device. It is an external communicating device with limited duration (<24hrs) of contact with tissue and blood. The HPR45i is designed to be used in combination with a radiofrequency generator (MedRF4000 (K210077) manufactured by F Care Systems USA LLC) for the application of haemostasis of soft tissue. The RF signal, which is generated by the generator, passes through the HPR45i, and is applied on the soft tissue by means of the HPR45i tip. When contact is made with tissue, the RF signal is converted into heat resulting in the bleeding to stop and haemostasis to appear. The HPR45i has no power regulating function, and it does not contain a diagnostic or monitoring function. Compatible device: - MEDRF 4000 ● No accessories are required to achieve the intended purpose of the HPR45i. ### 5. Intended Use/Indications for Use HPR45i electrosurgical electrode is intended to conduct radio frequency (RF) current for coagulation from the RF electrosurgical generator to target soft tissue in a broad range of surgical procedures. The device is a prescription use (Rx) device. ### 6. Substantial Equivalence Comparison The substantial equivalence comparison table is provided in Table 3. {5}------------------------------------------------ | Characteristic / Feature | E-Z<br>CLEAN<br>Electrosurgical Electrode<br>(Predicate Device) | HPR45i (Subject device) | Comments | | |-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | Device Name | E-Z CLEAN Electrosurgical<br>Electrode | HPR45i Electrode | N/A | | | Manufacturer Name | Megadyne Medical<br>Products, Inc. | F Care Systems | N/A | | | 510(K) Number | K081791 | K220725 | N/A | | | Classification Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Electrosurgical cutting and<br>coagulation device and<br>accessories | Same | | | Product code | GEI | GEI | Same | | | Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 | Same | | | Panel | General & Plastic Surgery | General & Plastic Surgery | Same | | | Class | Class II | Class II | Same | | | Indications for Use | E-Z Clean electrosurgical<br>electrodes are intended to<br>conduct radio frequency<br>(RF) current for cutting and<br>coagulation from the RF<br>electrosurgical generator to<br>target soft tissue in a broad<br>range of surgical procedures<br>requiring the use of electro<br>surgery for cutting and<br>cauterization.<br><br>Some tip configurations<br>have a specific geometry that<br>minimizes blanching and<br>thermal damage in skin<br>incisions when used in<br>conjunction with the<br>generator's Advanced<br>Cutting Effect (ACE) mode. | HPR45i electrosurgical<br>electrode is intended to conduct<br>radio frequency (RF) current for<br>coagulation from the RF<br>electrosurgical generator to<br>target soft tissue in a broad range<br>of surgical procedures. The<br>device is a prescription use (Rx)<br>device. | Similar. Refer<br>Justification 1 | | | Mode of Action | Thermocoagulation (RF) of<br>tissue by administration of<br>high frequency energy | Thermocoagulation (RF) of<br>tissue by administration of high<br>frequency energy | Same | | | Characteristic / Feature | E-Z<br>Electrosurgical<br>(Predicate Device) | CLEAN<br>Electrode | HPR45i (Subject device) | Comments | | Components | | Electrosurgical Generator, active electrode, return electrode, foot pedal. | Electrosurgical Generator was approved along with the neutral electrode, active electrode, foot pedal, | Same | | Output Energy type | | Radio Frequency | Radio Frequency | Same | | Mode of delivery | | Disposable Electrode | Disposable Electrode | Same | | Modality | | Monopolar | Monopolar | Same | | Rx or OTC | | Prescription Use | Prescription Use | Same | | Maximum power | | 300 watts | 25 Watts | Different. Refer<br>justification 2. | | Maximum Voltage | | ≤10.8 KV | 800 v | Different. Refer<br>justification 3. | | Maximum Frequency | | 510 kHz | 4MHZ | Different. Refer<br>justification 4. | | Tip Head<br>Dimensions | Tip<br>length of<br>Blade<br>Electrode | 21 mm | N/A | This<br>configuration is<br>not available in<br>subject device. | | | Tip<br>length of<br>Ball<br>Electrode | 5.3 mm | N/A | This<br>configuration is<br>not available in<br>subject device. | | | Tip<br>length of<br>Needle<br>Electrode | 4.2 mm | 10 mm | Different. Refer<br>justification 5. | | Tissue/<br>Tip<br>head surface-<br>contacting<br>area | Blade<br>electrode | 53mm² | 44mm² | Similar | | | Ball<br>electrode | 56mm² | | | | | Needle<br>electrode | 5.8mm² | | | | Electrode Material | | 300 series stainless steel | Stainless steel, AISI316L | Same | | Patient<br>Contacting<br>Material | | SS, Polyolefin, PTFE, vinyl, and Silicone | AISI316L Stainless steel and PTFE | Same | | Insulation Material | | Polyolefin and PTFE | PTFE | Similar. Refer<br>Justification 6. | | Characteristic / Feature | E-Z Electrosurgical Electrode (Predicate Device) | CLEAN | HPR45i (Subject device) | Comments | | Sterilization | Radiation Gamma<br>EO | EO | | Same | | Single use | Yes | Yes | | Same | | Coating Material | PTFE | PTFE | | Same | # Table 3 Comparison of Technological Characteristics between subject and predicate device {6}------------------------------------------------ {7}------------------------------------------------ Justification 1: The minor difference in the intended use of the device is that HPR45i is intended only for coagulation not for cutting but the predicate device is intended to be used for both cutting and coagulation. However, this difference will not raise any concern on the safety and effectiveness of the subject device since coagulation is covered under the intended use of predicate device. Justification 2: The predicate device, E-Z CLEAN Electrosurgical Electrode is used for both cutting and coagulation whereas HPR45i is not used for cutting, only control of hemostasis, therefore the maximum power of HPR45i (25 W) is much lower than the predicate (300 W). Justification 3: The predicate device, E-Z CLEAN Electrosurgical Electrode is used for both cutting and coagulation whereas HPR45i is not used for cutting, only control of hemostasis, therefore the maximum voltage of HPR45i (800V) is much lower than the predicate (≤10.8 KV). Justification 4: The predicate device, E-Z Clean Electrosurgical uses a lower frequency (510 kHz), the HPR45i uses a higher frequency (4MHZ). The higher frequency in HPR45i has no effect on efficiency as evidenced by the tissue testing. Justification 5: The tip length of HPR45i and E-Z Clean Electrosurgical Electrode different. However, tip length of HPR45i (10mm) tip length falls within the tip length (21mm, 5.3mm and 4.2mm) of the different versions of the predicate device, E-Z Clean Electrosurgical Electrode. Justification 6: The PTFE material was selected as it was more durable material than polyolefin. Therefore, the above difference does not raise any safety or effectives tissue for the subject device {8}------------------------------------------------ ## 7. Summary of Non- Clinical Data The non-clinical performance tests have been executed in line with recommendations of the FDA guidance: "Premarket Notification [510(k)] Submissions for Electrosurgical Devices for General Surgery" - Guidance for Industry and Food and Drug Administration Staff, August 15, 2016. The following performance tests were carried out for HPR45i: | S. No | Test Performed | Results | |-------|-----------------------------------------------------|---------| | 1 | ESU Testing | Pass | | 2 | Active Components/ Accessories Testing | Pass | | 3 | Neutral Electrode Testing | Pass | | 4 | Miscellaneous Components/ Accessories Testing | Pass | | 5 | Capacitive Coupling Testing | Pass | | 6 | Electrical safety and Electromagnetic Compatibility | Pass | | 7 | Thermal Effect on Tissue | Pass | #### Table 4 Non-Clinical Performance Tests HPR45i comply with the following international and FDA-recognized consensus standards: - . IEC 60601-2-2: 2018 Medical electrical equipment - Part 2-2: Requirements for the safety of high frequency surgical equipment. - IEC 60601-1:2006/A11:2011/A1:2013 Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance. - IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment- Part 1-2: General . Requirements for Basic safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests. - . ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. - . 10993-3 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity. - 10993-4 Third edition 2017-04 Biological Evaluation of Medical Devices Part 4: ● Selection of Tests for interactions with Blood. - ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. - ISO 10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. - . ISO 10993-11: Biological evaluation of medical devices - Part 11: Tests for systemic toxicity. - ISO 10993-18 Second edition 2020-01 Biological Evaluation of Medical Devices Part . 18: Chemical characterization of medical device materials within a risk management process. {9}------------------------------------------------ Traditional 510(K) HPR45i HPR45i Monopolar Electrodes are sterilized by using a validated ethylene oxide cycle. The sterilization cycle has been validated to ensure a sterility level of (SAL) 10-6 in accordance with ISO 11135. Furthermore, the packaging integrity and accelerated aging test were performed on the subject device to support the proposed shelf life. All non-clinical tests met the acceptance criteria specified in the standards. #### 8. Summary of Clinical Data As per FDA's guidance document, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" Clinical data are generally not necessary to support 510(k) submissions for electrosurgical devices that are intended for general surgery indications. Moreover, indications for use of HPR45i or device technology or mechamism of action is not significantly different when compared to the predicate device. Therefore, no clinical data was submitted for HPR45i. #### 9. Conclusion Based on the comparison and analysis above, the proposed device "HPR45i" is determined to be Substantially Equivalent (SE) to the predicate device "E-Z Clean Electrosurgical Electrode".
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...