K160041 · Elliquence, LLC · GEI · Mar 7, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160041
Device Name
Cobbra RF Tissue Dissector
Applicant
Elliquence, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cobbra RF Tissue Dissector is intended for use by a physician familiar with resection, incision, and hemostasis in soft tissue surgical procedures. The types of surgery intended are: General surgery, Laparoscopic procedures, Endoscopic procedures, Open abdominal, Orthopedic coagulation, Thorascopic coagulation, Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization), Ear, Nose, Throat coagulation.
Device Story
Single-use, sterile, monopolar electrosurgical instrument; used exclusively with elliquence Surgi-Max Plus RF generator. Device consists of plastic handpiece, partially-insulated electrode with active edge, cable, and plug. RF electrical power from generator transfers to tissue via active electrode edge; time-varying voltage induces tissue cutting or coagulation. Used in OR settings by physicians for soft tissue surgical procedures. Output provides hemostasis and tissue resection/incision. Benefits include precise tissue management during surgery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only, including mechanical testing, thermal damage measurements, simulated use, biocompatibility (ISO 10993-1), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2).
Technological Characteristics
Monopolar electrosurgical instrument. Materials: 17-4PH stainless steel, ABS plastic, PFA, PVC, copper. Biocompatibility per ISO 10993-1. Energy source: RF generator (Surgi-Max Plus). Five models varying by electrode diameter (0.25 to 1.5 inches). EtO-sterilized, single-use. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2.
Indications for Use
Indicated for use by physicians for resection, dissection, incision, and hemostasis in soft tissue surgical procedures including general, laparoscopic, endoscopic, open abdominal, orthopedic, thorascopic, neurosurgical, gynecological (excluding female sterilization), and ENT procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
elliquence Surgi-Max Plus radiofrequency generator (K100390)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2016
Elliquence LLC. Mr. Paul Buhrke IV QA/RA Manager 2455 Grand Avenue Baldwin, New York 11510
Re: K160041
Trade/Device Name: Cobbra RF Tissue Dissector Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 6, 2016 Received: January 8, 2016
Dear Mr. Buhrke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160041
Device Name Cobbra RF Tissue Dissector
#### Indications for Use (Describe)
The Cobbra RF Tissue Dissector is intended for use by a physician familiar with resection, incision, and hemostasis in soft tissue surgical procedures. The types of surgery intended are: General surgery, Laparoscopic procedures, Endoscopic procedures, Open abdominal, Orthopedic coagulation, Thorascopic coagulation, Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization), Ear, Nose, Throat coagulation.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo that features a stylized white outline of a person's back and shoulder. The person is facing away from the viewer, and the outline emphasizes the spine and shoulder muscles. The background is a dark blue circle with a glossy effect, giving it a three-dimensional appearance. A white shadow is visible at the bottom right of the circle, adding depth to the image.
Page 1 of 3
510 (k) Summary
(As required by 21 CFR 807.92(a))
Date Prepared
January 6, 2016
# Submitter's Information (807.92(a)(1))
Company Name and Address:
Elliquence, LLC. 2455 Grand Avenue Baldwin, NY 11510 Phone: (516) 277-9000 Fax: (516) 277-9001 www.elliquence.com
Establishment Registration for Elliquence, LLC is 3007024186.
Contact Information:
Daniel Connell - Regulatory Affairs Associate Phone: (516) 277-9036 Fax: (516) 277-9011 Email: dconnell@eliquence.com
Paul Buhrke – QA/RA Manager Phone: (516) 277-9010 Fax: (516) 277-9011 Email: pbuhrke@elliquence.com
## Device Information (807.92(a)(2))
Trade Name Cobbra™ RF Tissue Dissector
Common/Usual Name Monopolar electrosurgical device
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "elliquence Innovative Medical Solutions". The word "elliquence" is written in a bold, sans-serif font, with the "e" at the beginning stylized as a circle with a dot inside. Below "elliquence" is the phrase "Innovative Medical Solutions" in a smaller, sans-serif font. The logo is in black and white.
Page 2 of 3
Classification Name and Regulation Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400.
Class FDA Classification: Class II FDA Product Code: GEI
## Predicate Devices (807.92(a)(3))
- Elliquence Electrodes (510(k) K142410)
- PEAK Surgery System (510(k) K082786) .
#### Device Description (807.92(a)(4))
The Cobbra™ RF Tissue Dissector is a single-use, sterile, monopolar electrosurgical instrument intended for use exclusively with the elliquence Surgi-Max® Plus radiofrequency generator (510(k) K100390). The basic design of the device is a plastic handpiece, a partiallyinsulated electrode with an uninsulated, active edge, and a cable and plug. The electrical power operating at radio frequency (RF) is transferred to tissue at the surgical site via the active edge of the Cobbra™ RF Tissue Dissector's electrode. The time-varying voltage produced by RF electrical power source vields a predetermined electrosurgical effect, such as tissue cutting or coagulation.
The Cobbra™ RF Tissue Dissector is classified as a sterile, surgically-invasive, active device with an intended patient contact period of <1hr (transient). It is composed of 17-4PH medical-grade stainless steel, ABS plastic, PFA, PVC, and copper. The biocompatibility of patient contacting materials has been assured through appropriate testing in accordance with ISO 10993-1.
The Cobbra™ RF Tissue Dissector is individually-packaged and supplied EtO-sterilized for single use. It is available in five models which differ only by the diameter of the electrode: RF-COBB/XS (0.25 in.), RF-COBB/S (0.375 in.), RF-COBB/M (0.5 in.), RF-COBB/L (1.0 in.), and RF-COBB/XL (1.5 in.).
#### Intended Use (807.92(a)(5))
The Cobbra™ RF Tissue Dissector is intended for use by a physician familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures. The types of surgery intended are: General surgery, Laparoscopic procedures, Endoscopic procedures, Open
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a company logo for "elliquence Innovative Medical Solutions". The logo is in black font and the words "Innovative Medical Solutions" are smaller than the company name. Below the logo is the text "510(k) K160041 Page 3 of 3".
abdominal, Orthopedic coagulation, Thorascopic coagulation, Neurosurgical coagulation, Gynecological coagulation, (except for use in female sterilization), and Ear, Nose, Throat coaqulation.
## Substantial Equivalence Comparison (807.92(a)(6))
The Cobbra™ RF Tissue Dissector is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the two chosen predicate devices:
- (1) elliquence Electrodes (510(k) K142410)
- (2) PEAK Surgery System (510(k) K082786)
An extensive comparison chart and discussion is provided within the 510(k) submission.
## Non-Clinical Testing (807.92(b)(1))
Bench testing performed to support substantial equivalence included:
- Various performance tests including mechanical testing, thermal damage measurements, . and simulated use tests
- Applicable biocompatibility tests conducted in accordance with ISO 10993-1 ●
- . Electrical safety and electromagnetic compatibility testing in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2
- . Testing to support the shelf life determination and sterility of the product
# Clinical Testing (807.92(b)(2))
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
## Conclusion (807.92(b)(3))
Based upon a comparison of the intended uses, technological characteristics, and performance testing, we have concluded that the subject device is as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.