EDGE Arthroscopic Energy System, Arthroscopic Energy System Probes, Arthroscopic Energy System Wireless Footswitch, Arthroscopic Energy System Corded Footswitch, Arthroscopic Energy System Foot Control, Extension

K153499 · Conmed Corporation · GEI · Dec 18, 2015 · General, Plastic Surgery

Device Facts

Record IDK153499
Device NameEDGE Arthroscopic Energy System, Arthroscopic Energy System Probes, Arthroscopic Energy System Wireless Footswitch, Arthroscopic Energy System Corded Footswitch, Arthroscopic Energy System Foot Control, Extension
ApplicantConmed Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateDec 18, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Arthroscopic Energy System is indicated for resection, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures.

Device Story

Arthroscopic Energy System (AES-1) is a radiofrequency (RF) generator used in arthroscopic and orthopedic surgery. System components include RF generator, single-use monopolar/bipolar probes, dispersive pad, and wired/wireless foot controls. Generator features LCD touch screen for power level adjustment. Surgeon operates device via hand-piece controls or foot switches to deliver RF energy for soft tissue resection, ablation, and hemostasis. Device functions in wet field environments. Output allows precise tissue management, potentially reducing surgical time and improving patient outcomes. System is used in clinical/OR settings by surgeons.

Clinical Evidence

Bench testing only. Performance data included verification and validation of the footswitch and extension cable, as well as electromagnetic compatibility and electrical safety testing to demonstrate substantial equivalence.

Technological Characteristics

Radiofrequency generator; bipolar and monopolar output; LCD touch screen interface; wired/wireless foot control connectivity; single-use disposable probes. Operates in cut and coagulation modes. Complies with electrical safety and electromagnetic compatibility standards.

Indications for Use

Indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in patients undergoing arthroscopic and orthopedic surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing right, with flowing lines that resemble hair or ribbons. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ConMed Corporation Ms. Dionne Sanders Manager, Regulatory Affairs 525 French Road Utica, New York 13502 December 18, 2015 Re: K153499 Trade/Device Name: Arthroscopic Energy System with Accessory Probes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 4, 2015 Received: December 7, 2015 Dear Ms. Sanders: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Joshua C. Nipper -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153499 Device Name Arthroscopic Energy System with Accessory Probes Indications for Use (Describe) The Arthroscopic Energy System is indicated for resection, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Conmed Corporation. The logo consists of a blue square with a white "C" shape inside, followed by the word "CONMED" in black, with the word "CORPORATION" in smaller black letters underneath. The logo is clean and professional, and the colors are simple and effective. ### 510(k) SUMMARY #### I. SUBMITTER ConMed Corporation 525 French Road Utica, NY 13502 Phone: 727-399-5564 Fax: 727-399-5264 Contact Person: Dionne Sanders, RAC Date Prepared: December 4, 2015 #### II. DEVICE | Name of Device: | Arthroscopic Energy System | |----------------------|-------------------------------------------------------------------| | Common Name: | Electrosurgical Device and Accessories | | Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories | | Regulatory Class: | Class II, per 21 CFR Part 878.4400 | | Review Panel: | General and Plastic Surgery Devices | | Product Codes: | GEI | #### III. PREDICATE DEVICE | Device Name: | Arthroscopic Energy System | |---------------|----------------------------| | Company Name: | ConMed Corporation | | 510(k) #: | K140578 | #### IV. DEVICE DESCRIPTION The ConMed Arthroscopic Energy System consists of a radiofrequency (RF) generator, probes or electrodes, a dispersive pad (as needed), and wired or wireless foot controls. The probes or electrodes are single-use devices used with the Arthroscopic Energy System for the purpose of wet field arthroscopic and orthopedic soft tissue resection (dissection), ablation (removal) and coagulation (hemostasis). The generator is compatible with ConMed monopolar and bipolar probes or electrodes. The generator outputs are delivered to the probes or electrode handpieces within the specification ranges currently cleared and marketed by generators that have equivalent indications. The generator has a liquid crystal display (LCD) touch screen user interface. Power levels are assigned numeric values and are adjustable via the {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a stylized blue square with a white curved shape inside on the left. To the right of the square is the company name, "CONMED," in a bold, sans-serif font, with a horizontal line underneath. Below the line, the word "CORPORATION" is printed in a smaller, sans-serif font. user interface. Control is accomplished via probe or electrode hand-piece controls, wired foot control or wireless foot control. #### INTENDED USE / INDICATIONS FOR USE V. The Arthroscopic Energy System is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures. ### VI. COMPARISION OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The following table represents a summary of the technological characteristics between the proposed and the predicate device. | | PROPOSED DEVICE | PREDICATE DEVICE | |--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | ConMed Corporation | ConMed Corporation | | | Arthroscopic Energy System | Arthroscopic Energy System | | | (AES-1) | (AES-1) | | Brief<br>Description | A radiofrequency generator that is<br>intended to be used with a selection<br>of bipolar or monopolar probes and<br>electrodes, a dispersive pad (as<br>needed), and wired or wireless foot<br>controls. | A radiofrequency generator that is<br>intended to be used with a selection<br>of bipolar or monopolar probes and<br>electrodes, a dispersive pad (as<br>needed), and wired or wireless foot<br>controls. | | Intended<br>Use/<br>Indications<br>for Use | Indicated for resection, ablation,<br>and coagulation of soft tissue and<br>hemostasis of blood vessels in<br>arthroscopic and orthopedic<br>surgical procedures. | Indicated for resection, ablation,<br>and coagulation of soft tissue and<br>hemostasis of blood vessels in<br>arthroscopic and orthopedic<br>surgical procedures. | | Functionality | Bipolar and Monopolar<br>radiofrequency | Bipolar and Monopolar<br>radiofrequency | | Operating<br>modes | Cut and Coagulation | Cut and Coagulation | | Probes | Single-use, disposable bipolar and<br>monopolar probes | Single-use, disposable bipolar and<br>monopolar probes | | Footswitch,<br>Extension<br>cable | Yes | N/A | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a blue square with a white abstract shape on the left. To the right of the square is the company name, "CONMED," in bold, black letters. Below the company name is the word "CORPORATION" in smaller, black letters. #### VII. PERFORMANCE DATA Testing has been completed to demonstrate that the device performs as intended given the modification, and is substantially equivalent to the predicate device. Completed test data includes the following: - > Footswitch, Extension Cable - Verification о - Validation o - > Standards evaluation - Electromagnetic compatibility O - Electrical safety 0 #### VIII. CONCLUSION The Arthroscopic Energy System is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the Arthroscopic Energy System (K140578) and raises no new issues of safety or effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...