MODIFICATION TO ARTHROCARE ARTHROWANDS

K083306 · Arthrocare Corp. · GEI · Dec 10, 2008 · General, Plastic Surgery

Device Facts

Record IDK083306
Device NameMODIFICATION TO ARTHROCARE ARTHROWANDS
ApplicantArthrocare Corp.
Product CodeGEI · General, Plastic Surgery
Decision DateDec 10, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ArthroCare ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: [Table of procedures including ACL/PCL, acromioplasty, articular cartilage, bursectomy, chondroplasty, facia, ligament, notchplasty, scar tissue, soft tissue, subacromial decompression, synovectomy, tendon, acetabular labrum, articular labrum, capsule, capsular release, cartilage flaps, cysts, discoid meniscus, frozen shoulder release, glenoidale labrum, lateral release, loose bodies, meniscal cystectomy, meniscectomy, plica removal, synovial membrane, triangular fibrocartilage (TFCC), villusectomy, carpal ligaments, glenohumeral capsule, medial retinaculum, rotator cuff, and wrist tendons across various joints including ankle, elbow, hip, knee, shoulder, and wrist]. Prescription Use (Part 21 CFR 801 Subpart D).

Device Story

ArthroCare ArthroWands are bipolar, single-use, high-frequency electrosurgical devices; used in arthroscopic and orthopedic procedures. Device performs resection, ablation, and coagulation of soft tissue; provides hemostasis of blood vessels. Operated by surgeons in clinical settings. Output is thermal energy for tissue modification; assists in surgical debridement, excision, and release procedures. Benefits include precise tissue management during minimally invasive joint surgery.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Bipolar, high-frequency electrosurgical device; single-use; intended for arthroscopic/orthopedic use. Materials specification modified from predicate. Sterilization parameters remain unchanged.

Indications for Use

Indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures involving the ankle, elbow, hip, knee, shoulder, and wrist.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083306 # 510(k) Summary # ArthroCare Corporation ArthroCare ArthroWands Page 1 of 3 DEC 1 D., 200 Y #### General Information Submitter Name/Address: Establishment Registration Number: Contact Person: Date Prepared: Device Description Trade Name: Generic/Common Name: Classification Name: ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, CA 94085-3523 2951580 Valerie Defiesta-Ng Director, Regulatory Affairs November 6, 2008 ArthroCare ArthroWands® Electrosurgical Device and Accessories Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400) Predicate Devices ArthroCare ArthroWands® K082980 #### Product Description The ArthroCare ArthroWands are bipolar, single use, high frequency electrosurgical devices designed for specific indications in arthroscopic and orthopedic procedures. {1}------------------------------------------------ K083306 ## Intended Uses , ← . (2) દિવેલું છે. જીવની સુરત દિવસ છે. જીવની સાથે છે. આ ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપા 2 The ArthroCare ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: | Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) | |----------------------------------------|--------------------------------------------------------------------------------------| | Ablation and Debridement | | | • ACL/PCL | Knee | | • Acromioplasty | Shoulder | | • Articular Cartilage | All Joints | | • Bursectomy | All Joints | | • Chondroplasty | All Joints | | • Facia | All Joints | | • Ligament | All Joints | | • Notchplasty | Knee | | • Scar Tissue | All Joints | | • Soft Tissue | All Joints | | • Subacromial Decompression | Shoulder | | • Synovectomy | All Joints | | • Tendon | All Joints | | Excision and Resection | | | • Acetabular Labrum | Hip | | • Articular Labrum | All Joints | | • Capsule | All Joints | | • Capsular Release | Knee | | • Cartilage Flaps | Knee | | • Cysts | All Joints | | • Discoid Meniscus | Knee | | • Frozen Shoulder Release | Shoulder | | • Glenoidale Labrum | Shoulder | | • Lateral Release | Knee | | • Ligament | All Joints | | • Loose Bodies | All Joints | | • Meniscal Cystectomy | Knee | | • Meniscectomy | Knee | {2}------------------------------------------------ K083306 Page 3 of 3 | Continued | | |----------------------------------------|--------------------------------------------------------------------------------------| | Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) | | • Plica Removal | All Joints | | • Scar Tissue | All Joints | | • Soft Tissue | All Joints | | • Synovial Membrane | All Joints | | • Tendon | All Joints | | • Triangular Fibrocartilage (TFCC) | Wrist | | • Villusectomy | Knee | | Coagulation | | | • ACL/PCL | Knee | | • Articular Cartilage | All Joints | | • Carpal Ligaments | Wrist | | • Glenohumeral Capsule | Shoulder | | • Ligament | All Joints | | • Medial Retinaculum | Knee | | • Rotator Cuff | Shoulder | | • Tendon | All Joints | | • Wrist Tendons | Wrist | #### Substantial Equivalence This Special 510(k) proposes modifications in the materials specification for the ArthroCare ArthroWands, which were previously cleared in K082980 (November 5, 2008). The indications for use, technology, principle of operation, and sterilization parameters of the ArthroCare ArthroWands remain the same as in the predicate cleared 510(k)s. ## Summary of Safety and Effectiveness The modified ArthroCare ArthroWands, as described in this Special 510(k), are substantially equivalent to the predicate device. The proposed modifications in the materials is not substantial changes or modifications, and do not significantly affect the safety or efficacy of the devices. {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. #### Public Health Service DEC 10 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Arthrocare Corporation % Ms. Valerie Defiesta-Ng Director, Regulatory Affairs 680 Vasqueros Avenue Sunnyvale, California 94085-3523 Re: K083306 Trade/Device Name: ArthroCare ArthroWands Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: November 6, 2008 Received: November 10, 2008 Dear Ms. Defiesta-Ng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set. {4}------------------------------------------------ ## Page 2 - Ms. Valerie Defiesta-Ng forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH) s) Office of Compliation at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OBB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark M. Millican Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement 510(k) Number: K_083306 Device Name ArthroCare ArthroWands® Indications for Use: The ArthroCare ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthonedic groadores: | | Arthroscopic and Orthopedic Procedures | Joint Specific or All Joints (ankle, elbow, hip, knee, shoulder, wrist) | |---|----------------------------------------|-------------------------------------------------------------------------| | | Ablation and Debridement | | | • | ACL/PCL | Knee | | • | Acromioplasty | Shoulder | | • | Articular Cartilage | All Joints | | • | Bursectomy | All Joints | | • | Chondroplasty | All Joints | | • | Facia | All Joints | | • | Ligament | All Joints | | • | Notchplasty | Knee | | • | Scar Tissue | All Joints | | • | Soft Tissue | All Joints | | • | Subacromial Decompression | Shoulder | | • | Synovectomy | All Joints | | • | Tendon | All Joints | | | Excision and Resection | | | • | Acetabular Labrum | Hip | | • | Articular Labrum | All Joints | | • | Capsule | All Joints | | • | Capsular Release | Knee | | • | Cartilage Flaps | Knee | | • | Cysts | All Joints | | • | Discoid Meniscus | Knee | | • | Frozen Shoulder-Release | Shoulder | | • | Glenoidale Labrum | Shoulder | | • | Lateral Release | Knee | | • | Ligament | All Joints | | • | Loose Bodies | All Joints | | • | Meniscal Cystectomy | Knee | (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number K083306 10 {6}------------------------------------------------ | Continued | | |----------------------------------------|--------------------------------------------------------------------------------------| | Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) | | • Plica Removal | All Joints | | • Scar Tissue | All Joints | | • Soft Tissue | All Joints | | • Synovial Membrane | All Joints | | • Tendon | All Joints | | • Triangular Fibrocartilage (TFCC) | Wrist | | • Villusectomy | Knee | | Coagulation | | | • ACL/PCL | Knee | | • Articular Cartilage | All Joints | | • Carpal Ligaments | Wrist | | • Glenohumeral Capsule | Shoulder | | • Ligament | All Joints | | • Medial Retinaculum | Knee | | • Rotator Cuff | Shoulder | | • Tendon | All Joints | | • Wrist Tendons | Wrist | · Prescription Use (Part 21 CFR 801 Subpart D) X AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Net Re dyke Forman (Division Sign-Off)" Division of General, Restorative, and Netrological Devices **510(k) Number** k083306
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...