Secret MAX Multi-platform RF System
K241157 · Ilooda Co,., Ltd. · GEI · Nov 21, 2024 · General, Plastic Surgery
Device Facts
| Record ID | K241157 |
| Device Name | Secret MAX Multi-platform RF System |
| Applicant | Ilooda Co,., Ltd. |
| Product Code | GEI · General, Plastic Surgery |
| Decision Date | Nov 21, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4400 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Secret MAX DEEPLUS MENU and SMARTCURE MENU are intended use in dermatologic and general surgical procedures for electro-coagulation and hemostasis. The WAVPLUS MENU is intended for the treatment of the following medical condition; using the handpiece for delivery of non-thermal RF energy combined with massage; · Relief of minor muscle aches and pain, relief of muscle spasm · Temporary improvement of local blood circulation · Temporary reduction in the appearance of cellulite
Device Story
Secret MAX Multi-platform RF System is a multi-modal electrosurgical device. It utilizes RF energy for electro-coagulation and hemostasis (DEEPLUS/SMARTCURE menus) and non-thermal RF energy combined with massage (WAVPLUS menu) for musculoskeletal and aesthetic applications. Operated by clinicians in dermatologic and surgical settings. The device transforms electrical energy into RF output delivered via specialized handpieces. Clinical benefits include hemostasis during surgery and symptomatic relief of muscle pain/spasms and cellulite reduction. Healthcare providers use the system to perform procedures; output directly facilitates surgical control or therapeutic massage effects.
Clinical Evidence
No clinical data provided; device cleared based on technological characteristics and intended use equivalence.
Technological Characteristics
Multi-platform RF system; delivers thermal RF energy for coagulation/hemostasis and non-thermal RF energy for therapeutic massage. System includes multiple menus (DEEPLUS, SMARTCURE, WAVPLUS) and handpieces. Connectivity and specific material standards not disclosed in provided text.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electro-coagulation and hemostasis. Also indicated for relief of minor muscle aches/pain, muscle spasms, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Related Devices
- K201150 — InMode RF Multi-System · InMode , Ltd. · Jul 22, 2020
- K252367 — InbellaMulti System · Inbella Medical, Inc. · Oct 15, 2025
- K202780 — Sectum · Neauvia North America · Jul 30, 2021
- K182365 — TempSure System · Cynosure, Inc. · Oct 24, 2018
- K242996 — EndyMed PRO MAX · Endymed Medical, Ltd. · Jun 10, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2024
Ilooda Co.,Ltd. % Kim Do Hyun Regulatory Consultant CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul, Seoul 06210 Korea, South
Re: K241157
Trade/Device Name: Secret MAX Multi-platform RF System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, OUH, PBX Dated: April 26, 2024 Received: November 8, 2024
Dear Kim Do Hyun:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Long H. Chen
Long H. Chen -S-s
Date: 2024.11.21 12:48:53 -05'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241157
Device Name Secret MAX Multi-platform RF System
## Indications for Use (Describe)
The Secret MAX DEEPLUS MENU and SMARTCURE MENU are intended use in dermatologic and general surgical procedures for electro-coagulation and hemostasis.
The WAVPLUS MENU is intended for the treatment of the following medical condition; using the handpiece for delivery of non-thermal RF energy combined with massage;
· Relief of minor muscle aches and pain, relief of muscle spasm
· Temporary improvement of local blood circulation
· Temporary reduction in the appearance of cellulite
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."