DCI ENDOSCOPE
K991171 · Karl Storz Endoscopy · GCJ · Jun 23, 1999 · Gastroenterology, Urology
Device Facts
| Record ID | K991171 |
| Device Name | DCI ENDOSCOPE |
| Applicant | Karl Storz Endoscopy |
| Product Code | GCJ · Gastroenterology, Urology |
| Decision Date | Jun 23, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
The KSEA DCI Endoscope is indicated for use during endoscopic plastic, reconstructive and aesthetic procedures.
Device Story
The Karl Storz DCI Endoscope is a rigid, panoramic telescope utilizing rod lens technology. It is used by clinicians during endoscopic plastic, reconstructive, and aesthetic procedures to provide visualization of the surgical site. The device is constructed from surgical-grade stainless steel. It functions as a standard optical endoscope; it does not incorporate electronic sensors, software, or digital processing. The device is intended for prescription use.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Rigid, panoramic telescope; rod lens technology; surgical-grade stainless steel construction. No electronic components, software, or energy sources.
Indications for Use
Indicated for use during endoscopic plastic, reconstructive, and aesthetic procedures. Patient population not specified.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Related Devices
- K983363 — DCI ENDOSCOPE · KARL STORZ Endoscopy-America, Inc. · Dec 18, 1998
- K984369 — DCI ENDOSCOPE · KARL STORZ Endoscopy-America, Inc. · Jan 29, 1999
- K990004 — DCI ENDOSCOPE · KARL STORZ Endoscopy-America, Inc. · Apr 2, 1999
- K991228 — DCI ENDOSCOPE · Karl Storz Endoscopy · Jun 23, 1999
- K983384 — DCI ENDOSCOPE · KARL STORZ Endoscopy-America, Inc. · Nov 25, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
.
JUN 23 1999
K991171
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | Karl Storz Endoscopy - America, Inc.<br>600 Corporate Pointe Drive<br>Culver City, CA 90230<br>(310) 558-1500 |
|------------------------|---------------------------------------------------------------------------------------------------------------|
| Contact: | Kevin Kennan<br>Senior Regulatory Affairs Specialist |
| Device Identification: | Common Name:<br>Endoscope |
| | Trade Name: (optional)<br>Karl Storz DCI Endoscope |
Indication: The KSEA DCI Endoscope is indicated for use during endoscopic plastic, reconstructive and aesthetic procedures.
Device Description: The KSEA DCI Endoscope is comprised of a rigid, panoramic telescope which utilizes rod lens technology. The body contact portions of the KSEA DCI Endoscope are composed of surgical grade stainless steel, which is commonly used in medical devices for a wide range of applications and has a long history of biocompatibility for human use,
Substantial Equivalence: The KSEA DCI Endoscope is substantially equivalent to the predicate devices since the basic features, design and intended uses are the same. The minor differences between the KSEA DCI Endoscope and the predicate devices raise no new issues of safety and effectiveness, as these design differences have no affect on the performance, function or intended use of the devices.
Signed:
.
Kevin Kennah
Senior Regulatory Affairs Specialist
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle, there is a stylized graphic of three human profiles facing to the right, overlapping each other.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 23 1999
Mr. Kevin Kennan Senior Regulatory Affairs Specialist Karl Storz Endoscopy 600 Corporate Pointe Culver City, California 90230
K991171 Re: Trade Name: DCI Endoscope Regulatory Class: II Product Code: GCJ Dated: April 6, 1999 Received: April 7, 1999
Dear Mr. Kennan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
### Page 2 - Mr. Kevin Kennan
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## .
ن رن دل .
# ן ך וו ך וו ( 9 ( 1 ) ( 1 ( 1 ( 1
## Device Name: DCI Endoscope
Indications for Use: This instrument is indicated for use during endoscopic plastic, reconstructive and aesthetic procedures.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------|
| | |
| (Division Sign-Off) | |
| Division of General Restorative Devices<br>510(k) Number | K991171 |
| | |
| Prescription Use:<br>(Per 21 CFR 801.109) | <div style="display:inline-block; vertical-align:middle;">X</div> OR Over-The-Counter Use: |
| | (Optional Format 1-2-96) |