GELPORT BLUNT TIP TROCAR SYSTEM

K060629 · Applied Medical Resources Corp. · GCJ · Apr 7, 2006 · Gastroenterology, Urology

Device Facts

Record IDK060629
Device NameGELPORT BLUNT TIP TROCAR SYSTEM
ApplicantApplied Medical Resources Corp.
Product CodeGCJ · Gastroenterology, Urology
Decision DateApr 7, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Intended Use

The GelPort Blunt Tip Trocar System is indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes, extraperitoneal spaces and/or potential spaces for endoscopic instruments.

Device Story

GelPort Blunt Tip Trocar System is a sterile, single-use surgical instrument; designed to establish entry paths for endoscopic instruments during minimally invasive procedures (general, abdominal, gynecological, thoracic). Device consists of an obturator, seal, and cannula system; used in conjunction with Applied Medical's existing trocar products. Operated by surgeons in an OR setting to gain access through tissue planes or extraperitoneal spaces. Provides a stable access port for endoscopic tools; facilitates minimally invasive surgical techniques; benefits patients by enabling access with reduced trauma compared to open surgery.

Clinical Evidence

Bench testing only; no clinical data provided. Performance and functional testing compared the device to the predicate to demonstrate substantial equivalence.

Technological Characteristics

Sterile, single-use trocar assembly consisting of an obturator, seal, and cannula system. Mechanical device for surgical access. No electronic, software, or energy-based components.

Indications for Use

Indicated for patients undergoing general, abdominal, gynecological, or thoracic minimally invasive surgical procedures requiring endoscopic instrument access through tissue planes or extraperitoneal spaces.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## APR 7 2006 ## 510(K) SUMMARY 510(k) NUMBER: PENDING - SUBMITTED BY: Applied Medical Resources Corporation 22872 Avenida Empresa Rancho Santa Margarita, CA-92688 (949) 713-8000 - CONTACT PERSON: Cheryl Blake Vice President, Regulatory Affairs and Quality Systems DATE OF PREPARATION: February 7, 2006 NAME OF DEVICE: Blunt Tip Trocar System CLASSIFICATION NAME: Laparoscope, General & Plastic Surgery (21CFR 876.1500) TRADE NAME: GelPort® Blunt Tip Trocar System PREDICATE DEVICE: United States Surgical Blunt Tip Trocar (K924011) Originally filed under Origin Medsystems, Incorporated INTENDED USE: The GelPort Blunt Tip Trocar System is indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes, extraperitoneal spaces and/or potential spaces for endoscopic instruments. DEVICE DESCRIPTION: The GelPort Blunt Tip Trocar System is a sterile single use device, intended for use in conjunction with Applied's currently marketed Trocar products. A standard trocar assembly consists of an obturator, a seal and a cannula system. PERFORMANCE DATA SUMMARY: The performance and functional testing of the GelPort Blunt Tip Trocar System compared to its predicate device demonstrated that the GelPort Blunt Tip Trocar System is substantially equivalent to its predicate device and it introduces no new safety and effectiveness issues when used as instructed. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a logo with a circular border containing text, surrounding a stylized graphic. The graphic consists of three curved lines that resemble a series of waves or abstract human figures. The text around the border is small and difficult to read, but it appears to be arranged in a circular fashion, following the curve of the border. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APK 7 2006 Applied Medical Resources Corp. c/o Underwriters Laboratories, Inc. Mr. Morten Simon Christensen Staff Engineer & FDA Office Coordinator 455 E. Trimble Road San Jose, California 95131-1230 Re: K060629 Trade/Device Name: GellPort® Blunt Tip Trocar System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: March 28, 2006 Received: March 29, 2006 Dear Mr. Christensen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Revister. Please be advised that FDA s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {2}------------------------------------------------ ## Page 2 – Mr. Morten Simon Christensen forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. email ~ Mark N. Melk Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use 510(k) Number (if known): __________ Device Name: ____GelPort® Blunt Tip Trocar System_ Indications for Use: The GelPort Blunt Tip Trocar System is indicated for use in general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes, extraperitoneal spaces and/or re potin spaces for endoscopic instruments. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 'Malli Page / of / Division of General, Restorative. and Neurological Devices 510(k) Number_Koco629
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...