Medical Monitor (32HS710S)

K251209 · Lg Electronics.Inc · GCJ · Jun 12, 2025 · Gastroenterology, Urology

Device Facts

Record IDK251209
Device NameMedical Monitor (32HS710S)
ApplicantLg Electronics.Inc
Product CodeGCJ · Gastroenterology, Urology
Decision DateJun 12, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

This medical monitor is intended to provide color video displays and images from medical equipment which include laparoscopy and endoscopy systems for surgery and various medical imaging systems.

Device Story

The 32HS710S is an LCD medical monitor designed to display color video and images from surgical endoscopes, laparoscopic camera systems, and other compatible medical imaging equipment. It features multiple video interface ports, including DisplayPort, HDMI, DVI, and SDI, and supports over 100% of the sRGB spectrum. The device is intended for use in clinical settings, such as operating rooms, where it is operated by healthcare professionals to visualize real-time surgical or diagnostic imagery. The monitor serves as a visual interface for the surgeon or clinician, facilitating clinical decision-making during procedures by providing high-fidelity imaging. It does not perform automated diagnostic analysis; it functions as a display peripheral for medical imaging systems.

Clinical Evidence

No clinical studies were conducted. Evidence is limited to bench testing, including electrical safety (IEC 60601-1 Ed 3.2) and electromagnetic compatibility (IEC 60601-1-2 Ed 4.1, IEC TR 60601-4-2 Ed 1.0), and software validation performed in accordance with IEC 62304.

Technological Characteristics

LCD medical monitor; video interfaces include DisplayPort, HDMI, DVI, and SDI; supports >100% sRGB spectrum. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC). Software is classified as Basic Documentation Level.

Indications for Use

Indicated for use by clinicians to display color video and images from surgical laparoscopy, endoscopy, and other medical imaging systems.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 12, 2025 LG Electronics Inc. Daseul An RA Specialist 168, Suchul-daero Gumi-si, Gyeongsangbuk-do 39368 Korea, South Re: K251209 Trade/Device Name: Medical Monitor (32HS710S) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: April 15, 2025 Received: April 18, 2025 Dear Daseul An: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251209 - Daseul An Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K251209 - Daseul An Page 3 See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2025.06.12 12:30:14 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251209 | | | Device Name Medical Monitor (32HS710S) | | | Indications for Use (Describe) This medical monitor is intended to provide color video displays and images from medical equipment which include laparoscopy and endoscopy systems for surgery and various medical imaging systems. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} | 510(k) #: K251209 | 510(k) Summary | Prepared on: 2025-06-09 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | LG Electronics Inc. | | | Applicant Address | 168, Suchul-daero Gumi-si Gyeongsangbuk-do 39368 Korea, South | | | Applicant Contact Telephone | +82-1089140116 | | | Applicant Contact | Mrs. Daseul An | | | Applicant Contact Email | DL-FDA@lge.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Medical Monitor (32HS710S) | | | Common Name | Endoscope and accessories | | | Classification Name | Laparoscope, General & Plastic Surgery | | | Regulation Number | 876.1500 | | | Product Code(s) | GCJ | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K241402 | 32HR734S | GCJ | | Device Description Summary | | 21 CFR 807.92(a)(4) | | 32HS710S LCD monitor is intended to provide color video displays of images from surgical endoscope, laparoscopic camera system and other compatible medical imaging systems. This Medical Monitor has multiple video interface ports such as Display port, HDMI, DVI and SDI. This monitor displays color video and images from medical equipment which include various medical imaging systems. This monitor can be covered over 100% of the sRGB spectrum. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | This medical monitor is intended to provide color video displays and images from medical equipment which include laparoscopy and endoscopy systems for surgery and various medical imaging systems. | | | | Indications for Use Comparison | | 21 CFR 807.92(a)(5) | | | | | | Technological Comparison | | 21 CFR 807.92(a)(6) | | The comparison tables show some differences in specifications such as resolution, connectors, maximum power consumption and physical specifications (weight). However, these differences do not affect the safe and effective use of the device. The performance of 32HS710S has been validated in the software validation report. In addition, the safety of 32HS710S has been confirmed through the test according to IEC 60601-1 and 60601-1-2. | | | {5} Those reports confirm that all the differences between the proposed and predicate devices do not affect safety and effectiveness of the device. Therefore, the proposed device is substantially equivalent to the predicate device in terms of technological characteristics. # Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The following data were provided in support of the substantial equivalence determination: 1) Electrical Safety and Electromagnetic Compatibility Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: - IEC60601-1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems 2) Software Validation 32HS710S contain software that belongs to Basic Documentation Level. The software was designed and developed according to a software development process and was verified and validated according to IEC 62304. Software information is provided in accordance with FDA guidance: - The content of premarket submissions for Device Software Functions (June 14, 2023) No clinical studies were considered necessary, and therefore, none were conducted. Based on the information provided in this premarket notification and in accordance with the Federal Food, Drug, and Cosmetic Act, 21 CFR Part 807, LG Electronics Inc. has determined that there are no significant differences between the proposed device and the predicate device that would adversely affect its effectiveness and safety. The proposed device has been proved to be as safe and effective as the predicate devices, which was previously cleared in K241402. They share the same indications for use and similar technological characteristics, with any differences supported by software validation reports that demonstrate the proposed device's safety and effectiveness. Therefore, the technological differences between the proposed device and its predicate device do not raise any new concerns regarding safety or effectiveness, and the proposed device can be considered substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...