VersaOne Bladeless Trocar
K162584 · Covidien · GCJ · Oct 18, 2016 · Gastroenterology, Urology
Device Facts
| Record ID | K162584 |
| Device Name | VersaOne Bladeless Trocar |
| Applicant | Covidien |
| Product Code | GCJ · Gastroenterology, Urology |
| Decision Date | Oct 18, 2016 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
The VersaOne™ bladeless trocars are intended for use in a variety of gynecologic, general, thoracic and urologic endoscopic procedures to create and maintain a port of entry.
Device Story
VersaOne™ Bladeless Trocar is a surgical instrument used to create and maintain a port of entry for endoscopic procedures; used in gynecologic, general, thoracic, and urologic surgery. Device consists of a cannula and obturator; modified to accommodate 8mm instruments. Operated by surgeons in clinical settings. Functions by penetrating tissue to establish access; maintains port for instrument passage. Benefits patient by facilitating minimally invasive surgical access. Substantial equivalence established via bench and animal testing comparing leak rates, insertion/removal forces, and penetration/fixation forces against predicate devices.
Clinical Evidence
No human clinical data provided. Evidence consists of bench top testing (leak rate, insertion/removal force, penetration/fixation force) and animal testing (penetration/fixation force) to demonstrate functional performance. Biocompatibility confirmed per ISO 10993-1; sterilization validated per ISO 11135 and ISO 10993-7.
Technological Characteristics
Surgical trocar consisting of cannula and obturator. Materials comply with ISO 10993-1. Sterilization via ethylene oxide (EO) per ISO 11135; EO residuals per ISO 10993-7. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients undergoing gynecologic, general, thoracic, or urologic endoscopic procedures requiring a port of entry.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- VersaOne™ Bladeless Trocar (K151548)
- Autosuture™ Versaport™ Plus Bladeless Trocar (K081169)
Related Devices
- K152149 — VersaOne Bladed Trocar 5mm · Covidien · Jan 22, 2016
- K151548 — VersaOne Bladed Trocar, VersaOne Bladeless Trocar · Covidien · Jul 2, 2015
- K202931 — Disposable Trocar · Wuxi Shukang Medical Appliance Co., Ltd. · Nov 17, 2020
- K130435 — VERSAPORT V2 BLADELESS OPTICAL TROCAR · Covidien, Formerly US Surgical A Divison of Tyco H · Mar 14, 2013
- K160230 — VersaOne Optical Trocar 15mm, VersaOne Bladeless Trocar 15mm · Covidien · Mar 1, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 18, 2016
Covidien LLC Ms. Trang Huynh Principal Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K162584
Trade/Device Name: VersaOne™ Bladeless Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: October 11, 2016 Received: October 12, 2016
Dear Ms. Huynh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162584
Device Name VersaOne™ Bladeless Trocar
Indications for Use (Describe)
The VersaOne™ bladeless trocars are intended for use in a variety of gynecologic, general, thoracic and urologic endoscopic procedures to create and maintain a port of entry.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logos of Medtronic and Covidien. The Medtronic logo is on the left, and the Covidien logo is on the right. The Medtronic logo includes a circular design to the left of the company name, which is written in blue. The Covidien logo includes a blue square with a white cross in the center, and the company name is written in blue to the right of the square.
# 510(k) Summary
This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92.
| SUBMITTER: | Covidien llc<br>60 Middletown Avenue<br>North Haven, CT 06473 USA |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Trang Huynh<br>Principal Regulatory Affairs Specialist<br>Phone: (203) 492-7473<br>Fax: (203) 492-5029<br>Email: trang.huynh27@medtronic.com |
| DATE PREPARED: | September 16, 2016 |
| PRODUCT CODE: | GCJ |
| REGULATION NUMBER: | 78 |
| CLASS CODE: | Class II, Pursuant to 21 CFR 876.1500 |
| CLASSIFICATION NAME: | Endoscope and Accessories |
| TRADE /PROPRIETARY NAME: | VersaOne™ Bladeless Trocar |
| COMMON/USUAL NAME: | Surgical Trocar |
| PREDICATE DEVICES: | VersaOne™ Bladeless Trocar (K151548)<br>Autosuture™ Versaport™ Plus Bladeless Trocar (K081169) |
| INTENDED USE: | The VersaOne™ bladeless trocars are intended for use in a variety of gynecologic, general, thoracic and urologic endoscopic procedures to create and maintain a port of entry. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logos of Medtronic and Covidien. The Medtronic logo is on the left, and the Covidien logo is on the right. The Medtronic logo is a blue wordmark with a stylized atom symbol to the left. The Covidien logo is a blue square with a white cross in the center, followed by the company name in black text.
Modifications to the design of the current VersaOne™ Bladeless Trocar (K151548) include diameter change to the cannula and obturator components to accommodate an 8mm instrument.
Sterilization assessment was performed to confirm EO sterilization equivalency to the predicate device with respect to sterilization processes per ISO 11135 requirements and EO residuals per ISO 10993-7 requirements.
Shelf life assessment was performed per Covidien standard operating procedures to confirm product performance over time.
SUMMARY COMPARING THE TECHNOLOGICAL CHARACTERISTIC S OF THE SUBJECT AND PREDICATE DEVICES:
Biocompatibility testing was performed to confirm that all components are comprised of materials that are in accordance with ISO 10993-1 for their intended patient contact profile.
Performance testing including bench top and animal was performed to compare the functional performance of the subject device and the predicate device.
- o Bench top testing that supports the intended use of this device includes:
- . Leak rate
- . Insertion force & Removal force
- Penetration force & Fixation force
- Animal testing that supports the intended use of the device o includes:
- Penetration force & Fixation force
The results of testing demonstrated the modified VersaOne™ Bladeless CONCLUSION: Trocar 8mm is substantially equivalent to the legally marketed device.