Percutaneous Introducer Needle

K191295 · Progressive Medical, Inc. · GCJ · Aug 6, 2019 · Gastroenterology, Urology

Device Facts

Record IDK191295
Device NamePercutaneous Introducer Needle
ApplicantProgressive Medical, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateAug 6, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

Introducer for Endoscopic/Laparoscopic procedures.

Device Story

The PMIPIN15 Percutaneous Introducer Needle is a single-use, disposable device used by physicians during endoscopic and laparoscopic procedures. It consists of a 304 stainless steel needle with a polycarbonate luer hub and a 5 Fr low-density polyethylene (LDPE) introducer sheath. The device is manually inserted by the clinician to facilitate access. It functions as a mechanical introducer to assist in the performance of endoscopic/laparoscopic interventions. The device is provided sterile via ethylene oxide (EO) sterilization.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-5, ISO 10993-10), EO sterilization validation (ISO 11135), package validation (ASTM F2096, F1886, F88, D5276), shelf life validation (ASTM F1980), and mechanical performance testing (ISO 11070) including needle-to-hub union strength, force at break, and needle penetration testing.

Technological Characteristics

Materials: 304 stainless steel (needle), polycarbonate (luer hub), low-density polyethylene (sheath/sleeve). Dimensions: 15 gauge needle, 116 mm needle length, 104.4 mm introducer length, 5 Fr diameter. Sterilization: Ethylene Oxide (ISO 11135). Mechanical: Manual insertion. Standards: ISO 10993, ISO 11070, ASTM F1980, ASTM F2096, ASTM F1886, ASTM F88, ASTM D5276.

Indications for Use

Indicated for use as an introducer during endoscopic and laparoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, and on the right is the FDA logo. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION". August 6, 2019 Progressive Medical, Inc. % David Mullis Regulatory Consultant Mullis & Associates, Inc. 367 Pleasant Valley Rd., P.O. Box 39 Good Hope, Georgia 30641 Re: K191295 Trade/Device Name: Percutaneous Introducer Needle Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: April 30, 2019 Received: May 14, 2019 Dear David Mullis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K191295 Device Name PMIPIN15 Percutaneous Introducer Needle (5Fr) Indications for Use (Describe) Introducer for Endoscopic/Laparoscopic procedures. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Progressive Medical, Inc. PMIPIN15 Percutaneous Introducer Needle K191295 | 6.<br>510(k) Summary | | | | | |--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--| | Submitter: | Progressive Medical, Inc.<br>997 Horan Drive<br>Fenton, MO 63026<br>Establishment Registration Number: 300236238 | | | | | Contact: | David W. Mullis, Jr., Ph.D., RAC, FRAPS<br>Regulatory Consultant<br>Mullis & Associates, Inc.<br>367 Pleasant Valley Rd., P.O. Box 39<br>Good Hope, GA 30641<br>Phone: 770-722-6413<br>Fax:<br>770-207-7682 | | | | | Submission Date: | April 30, 2019 | | | | | Trade Name: | PMIPIN15 Percutaneous Introducer Needle (5 Fr) | | | | | Common Name: | Introducer | | | | | Regulation: | 876.1500 | | | | | Classification: | Class II | | | | | Product Code: | GCI | | | | | Panel: | General & Plastic Surgery | | | | | Medical Specialty: | Gastroenterology/Urology | | | | | Predicate Devices: | Ranfac Percutaneous Introducer PIN-15, K951090 | | | | | Device Description: | The PMIPIN15 Percutaneous Introducer Needle is a single-<br>use, disposable EO sterilized device made of 304 stainless steel<br>with a luer hub made of polycarbonate. The introducer sheath<br>is 5 Fr and manufactured from low density polyethylene. The<br>device is manually inserted by the physician during endoscopic<br>and/or laparoscopic procedures. | | | | | Indication for Use: | Introducer for Endoscopic/Laparoscopic procedures. | | | | | Substantial Equivalence: | | | | | The PMIPIN15 Percutaneous Introducer Needle (5 Fr) is substantially equivalent to the Ranfac PIN-15 Introducer that is {4}------------------------------------------------ legally marketed in the United States. Each device has the same indication use and is manually inserted by the physician. Both devices have common design specifications and materials. The PMIPIN15 Percutaneous Introducer Needle is a single use, disposable, EO sterilized device made of 304 stainless steel with a luer hub made of polycarbonate. Its 5 Fr introducer sheath is manufactured from low density polyethylene which is the same material as its predicate. Both the PMIPIN15 and the PIN-15 have a protective sleeve made of low density polyethylene. | Comparison | |------------| | | New Device | Predicate Device | Comment | |--------------------|----------------------|--------------------|---------| | Product Name | PMIPIN15 (5 Fr) | Ranfac PIN-15 | | | | Percutaneous | Percutaneous | | | | Introducer Needle | Introducer Needle | | | 510(k) Number | K191295 | K951090 | | | Company | Progressive Medical, | Ranfac Corp. | | | | Inc. | | | | Indication for Use | Introducer for | Introducer for | Same | | | Endoscopic/ | Endoscopic/ | | | | Laparoscopic | Laparoscopic | | | | procedures | procedures | | | Classification | Laparoscope, | Laparoscope, | Same | | Name | General & Plastic | General & Plastic | | | | Surgery | Surgery | | | Common Name | Introducer Needle | Introducer | | | Product Code | GCJ | GCJ | Same | | Regulatory Class | Class II | Class II | Same | | Regulation # | 876.1500 | 876.1500 | Same | | Sterility | Yes | Yes | Same | | Sterilization | EO | EO | Same | | Method | | | | | Single Use | Yes | Yes | Same | | Lengths/Sizes | Needle Gauge: 15 | Needle Gauge: 15 | Same | | | Needle Length: 116 | Needle Length: 116 | | | | mm | mm | | | | Introducer length: | Introducer length: | | | | 104.4 mm | 104.4 mm | | | | Introducer | Introducer | | | | Diameter: 5 Fr | Diameter: 5Fr | | {5}------------------------------------------------ | Materials | Needle: 304 | Needle: 304 | Same | |-------------------|---------------------|--------------------|------| | | Stainless Steel | Stainless Steel | | | | Handle: Low | Handle: Low | | | | Density | Density | | | | Polyethylene | Polyethylene | | | | Female Luer Hub: | Female Luer Hub: | | | | Polycarbonate | Polycarbonate | | | | Introducer: Low | Introducer: Low | | | | Density | Density | | | | Polyethylene | Polyethylene | | | | Protective Sleeve: | Protective Sleeve: | | | | Low Density | Low Density | | | | Polyethylene | Polyethylene | | | Mode of Operation | Manual Insertion | Manual Insertion | Same | | Package | Tyvek/Plastic Pouch | Tyvek/Plastic | Same | | Biocompatible | Yes | Yes | Same | | Shelf Life | 2 yr. | 5 Yr. | | | Bioburden | Yes | Unknown | | Non-clinical Testing: A series of preclinical tests were performed to assess the safety and performance of the PMIPIN15 Percutaneous Introducer Needle (5Fr) including biocompatibility physical and mechanical performance. > The following biocompatibility and safety tests were conducted in compliance with ISO standards: - Cytotoxicity Test: ISO10993-5 ● - Dermal Sensitization Guinea Pig Maximization Tests: 1SO10993-10 - Intracutaneous Test: ISO10993-10 - Ethylene Oxide Sterilization Validation: ISO 11135:2014 ● - Package Validation: ASTM 2096-11, ASTM ● F1186/1886M-16, ASTM F88/F88M-15, ASTM D5276-98 - Shelf Life Validation: ASTM F-1980-16 Physical measurements and visual inspection of the PMIPIN15 and Ranfac PIN-15 were performed to confirm similarity in design and materials. Performance evaluation, in compliance with ISO 11070:2014(E), was performed for strength of union of needle and hub as well as force at break. Needle penetration testing was conducted by an independent lab in accordance with ISO 11070:2014(E). Accelerated aging testing was conducted per ASTM F1980-16. Package integrity testing to assess seal strength and leak detection were performed in conformance with ASTM standards, as referenced above. The PMIPIN15 Percutaneous Introducer Needle (5 Fr) met its design specifications. {6}------------------------------------------------ Progressive Medical, Inc. PMIPIN15 Percutaneous Introducer Needle K191295 > Based on preclinical testing results, indications for use, Conclusion: performance evaluation, design specifications, and comparison with the predicate device, the PMIPIN15 Percutaneous Introducer Needle (5Fr) is substantially equivalent to its referenced predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...