STONE GRASPING BASKET, MODEL PD-T1 2115
K082735 · Sialo Technology, Ltd. · GCJ · Mar 13, 2009 · Gastroenterology, Urology
Device Facts
| Record ID | K082735 |
| Device Name | STONE GRASPING BASKET, MODEL PD-T1 2115 |
| Applicant | Sialo Technology, Ltd. |
| Product Code | GCJ · Gastroenterology, Urology |
| Decision Date | Mar 13, 2009 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Stone Grasping Basket is a medical device indicated for use by qualified surgeons in the treatment of salivary gland diseases. It is used endoscopically to entrap and remove mobile stones and stone fragments smaller than 5mm from the salivary gland.
Device Story
Sialo Stone Grasping Basket; 30mm long, 4-wire, 10mm diameter Nitinol basket. Designed for endoscopic passage through narrow working channels of endoscopes. Used by surgeons to mechanically entrap and retrieve mobile salivary stones/fragments <5mm. Device provides physical retrieval mechanism; no electronic or automated components. Used in clinical settings during sialoendoscopy procedures. Benefits include minimally invasive removal of salivary calculi.
Clinical Evidence
Bench testing only. Device subjected to safety and performance testing to ensure compliance with functional specifications.
Technological Characteristics
Nitinol construction; 30mm length, 4-wire, 10mm diameter basket configuration. Mechanical stone retrieval device; manual operation via endoscope working channel. No energy source, software, or connectivity.
Indications for Use
Indicated for qualified surgeons treating salivary gland diseases in patients with mobile stones or stone fragments smaller than 5mm in the salivary gland.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- Stone basket of KSEA Sialoendoscope (K012527)
- Stone Basket for Midiview series (K051073)
- Wittich Nitinol Stone Basket (K902944)
Related Devices
- K120468 — NCIRCLE(R) NITINOL TIPLESS STONE EXTRACTOR NGAGE(R) NITINOL STONE EXTRACTOR · Cook Incorporated · Jul 6, 2012
- K974300 — TELEMED SYSTEMS RETRIEVAL BASKET · Telemed Systems, Inc. · Dec 19, 1997
- K970121 — STONE DISLODGER, BASKET · Boston Scientific Corp · Apr 2, 1997
- K170811 — Single Use Retrieval Nitinol Basket V · Olympus Medical Systems Corp. · Nov 27, 2017
- K012527 — KSEA SIALOENDOSCOPES AND ACCESSORIES · KARL STORZ Endoscopy-America, Inc. · Nov 1, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
MAR 1 3 2009
#### 510(K) SUMMARY
# SIALOTECH MODULAR STONE GRASPING BASKET
#### 510(k) Number K082735
Applicant's Name:
- Sialo Technology Ltd. Suite 220 11 Ben Gurion Boulevard Ashkelon 78182 Israel Tel: +972-8-6710795 Fax: +972-9-6782524 e-mail: reuven@sialotechnology.com
Contact Person:
Ahava Stein/ Eyal Ozeri A. Stein - Regulatory Affairs Consulting 20 Hata'as St. -Kfar Saba 44425 Israel Tel. + 972-9-7670002 Fax. +972-9-7668534 e-mail: ahava@asteinrac.com or eyal(@asteinrac.com
Date Prepared: Trade Name:
September 14, 2008 Sialo Stone Grasping Basket
Classification Name: CFR Classification section 876.1500 (Product code GCJ)
Class II medical Device Classification:
The Sialo Modular Stone Grasping Basket device is comparable to Predicate Device: the following predicate devices:
- Stone basket of KSEA Sialoendoscope (K012527) manufactured by Karl Storz Endoscopy.
- Stone Basket for Midiview series (K051073) manufactured by Millennium Devices Inc.
- Wittich Nitinol Stone Basket (K902944) manufactured by Cook Inc
{1}------------------------------------------------
- The stone grasping basket is a 30mm long, 4 wire, 10mm Nitinol Device Description: Stone removal Basket, which is designed to pass through low diameter working channels of endoscopes for the purpose of retrieval of stones from the salivary gland.
#### Intended Use / Indication for Use:
The Stone Grasping Basket is a medical device indicated for use by qualified surgeons in the treatment of salivary gland diseases. It is used endoscopically to entrap and remove mobile stones and stone fragments smaller than 5mm from the salivary gland.
### Performance Standards: None.
- The Stone Grasping Basket device has been subjected to safety and Test Data: performance testing before release. Testing of the device included various performance tests designed to ensure that the device meets all its functional specifications and complies with safety standards.
## Substantial Equivalence:
The Sialo Stone Grasping Basket device is similar to the currently distributed endoscopic accessory devices intended for stone removal, such as the basket accessory of the Karl Storz KSEA Sialoendoscope (K012527), the stone basket of Midiview series (K051073) manufactured by by Millennium Devices Inc. and the Wittich Nitinol Stone Basket (K902944).
The device has the same intended use and uses the same basic technology as the predicate devices.
The results of the performance tests demonstrate that the stone grasping basket meets its specifications for efficacy and exhibits similar physical properties as the predicate devices.
- The conclusions drawn from the above Performance Testing and Conclusions: comparison to predicate devices, is that the Sialo Stone Grasping Basket device is substantially equivalent to the predicate devices listed above.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring an abstract design of an eagle or bird-like figure.
MAR 1 3 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sialo Technologies Ltd. c/o Ms. Ahava Stein General Manager, A. Stein - Regulatory Affairs Consulting Ltd. Beit Hapa'amon (Box 124) 20 Hata'as St. 44425 Kfar Saba, Israel
Re: K082735
Trade/Device Name: Stone Grasping Basket Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Il Product Code: GCJ Dated: February 25, 2009 Received: March 3, 2009
Dear Ms. Stein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic, product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Malvina B. Egleston, and
Malvina B. Eydelman, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K082.735
Sialo Modular Stone Grasping Basket 510(k)
#### INDICATIONS FOR USE
510(k) Number (if known): K082735
Sialo Stone Grasping Basket Device Name:
Intended Use Statement:
The Stone Grasping Basket is a medical device indicated for use by qualified surgeons in the treatment of salivary gland diseases. It is used endoscopically to entrap and remove mobile stones and stone fragments smaller than 5mm from the salivary gland.
Prescription Use (Per 21 C.F.R. 801 Subpart D) C)
OR
Over-The-Counter Use (Optional Format Subpart
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel C. Olson
(Division Sign-Off) Division of Ophthalmic and Ear. Nose and Throat Devices
082735 510(k) Number