LAPMAN LAPAROSCOPE MANIPULATOR AND ACCESSORIES, MODEL 900-200 SERIES

K062968 · Medsys, S.A. · GCJ · Jan 3, 2007 · Gastroenterology, Urology

Device Facts

Record IDK062968
Device NameLAPMAN LAPAROSCOPE MANIPULATOR AND ACCESSORIES, MODEL 900-200 SERIES
ApplicantMedsys, S.A.
Product CodeGCJ · Gastroenterology, Urology
Decision DateJan 3, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The LapMan can be used during the laparoscopic procedures listed below and should not be used for any other surgical procedure or purpose. 1. General surgery: laparoscopic cholecystectomy, laparoscopic Nissen Fundoplication, laparoscopic inguinal hernia repair, laparoscopic gastric banding. 2. Gynaecology: laparoscopic adnexal surgery (ovarian cystectomy, adnexectomy, salpingectomy, salpingoneostomy, tubal reanastomosis), laparoscopic management of ectopic pregnancy, laparoscopic treatment of endometriosis. laparoscopic myomectomy, laparoscopic hysterectorny, laparoscopic colposuspension 3. Urology: laparoscopic prostatectomy.

Device Story

Robotic system for holding/positioning rigid laparoscope/endoscope; surgeon-controlled via radiofrequency hand-held device. Used in OR by surgeons/gynecologists/urologists. System receives RF commands to manipulate scope position; provides stable visualization during laparoscopic procedures. Reduced motion magnitude/speed compared to predicate enhances patient safety during unexpected movements. Benefits include stable, surgeon-controlled endoscopic visualization.

Clinical Evidence

Bench testing only. Compliance verified against EN 60529, IEC 601-1, EN 300 220-3, and various EN 60601 series standards for electrical safety, EMC, and radio spectrum performance. No clinical data presented.

Technological Characteristics

Robotic computer-driven manipulator for rigid laparoscopes. Radiofrequency-based hand control interface. Motion parameters (magnitude/speed) restricted for safety. Electrical/EMC compliance per IEC 601-1, EN 60601-1-2, EN 61000-4 series. CE marked per 93/42/EC.

Indications for Use

Indicated for patients undergoing laparoscopic procedures in general surgery (cholecystectomy, Nissen Fundoplication, inguinal hernia repair, gastric banding), gynaecology (adnexal surgery, ectopic pregnancy management, endometriosis treatment, myomectomy, hysterectomy, colposuspension), and urology (prostatectomy).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ o 62968 # LapMan 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: #### Applicant Information: JAN - 3 2007 Xavier Moreels Managing Director Medsys S.A. Rue Chainisse, 39 B-5030 Gembloux. BELGIUM. ### Date Prepared: September 25, 2006 ### Proposed Device: Lapman Laparoscope Manipulator System and Accessories #### Classification: General and Plastic Surgery Class II 21 CFR PART 876.1500 #### Predicate Device: AESOP 3000 System and accessories. K972699 THE AESOP 3000 is a robotic computer-driven system whose basic function is to hold and position a rigid laparoscope/endoscope under the direct control of a surgeon in the field of endoscopy/laparoscopy surgery. Communication between surgeon and robotic system is ensured by a voice controlled or a wired system. #### Device Description: LapMan is a robotic computer-driven system used by surgeon to hold and position a rigid laparoscope/endoscope. Surgeon orders are transmitted to the robotic system by a radiofrequency Hand Controlled device. It is indicated for use in General surgery (laparoscopic cholecystectomy, laparoscopic Nissen Fundoplication, laparoscopic inguinal hernia repair, laparoscopic gastric banding), Gynaecology (laparoscopic adnexal surgery i.e. ovarian cystectomy, oophorectomy, salpingectomy, salpingoneostomy, tubal reanastomosis, laparoscopic management of ectopic pregnancy, laparoscopic treatment of endometriosis, laparoscopic myomectomy. laparoscopic hysterectomy, laparoscopic colposuspension) and Urology (laparoscopic prostatectomy). {1}------------------------------------------------ The LapMan should only be used by qualified medical personnel, such as trained surgeons, gynecologists, urologists, and similarly-trained medical personnel trained in the use of laparoscopic/endoscopic devices. #### Statement of intended use: The LapMan can be used during the laparoscopic procedures listed below and should not be used for any other surgical procedure or purpose. - 1. General surgery: laparoscopic cholecystectomy, laparoscopic Nissen Fundoplication, laparoscopic inguinal hernia repair, laparoscopic gastric banding. - 2. Gynaecology: laparoscopic adnexal surgery (ovarian cystectomy, adnexectomy, salpingectomy, salpingoneostomy, tubal reanastomosis), laparoscopic management of ectopic pregnancy, laparoscopic treatment of endometriosis. laparoscopic myomectomy, laparoscopic hysterectorny, laparoscopic colposuspension - 3. Urology: laparoscopic prostatectomy. ### Comparison of Technological Characteristics: The existing differences between the Predicate device and LapMan involve a communication system using radiofrequency instead of a voice controlled or wired system. Maqnitude and speed of the allowed motions for LapMan are significantly lower than for predicate. This reduces the risk for the patient in cases of unexpected motions. However the volume which can be explored with LapMan remains sufficient for the intended used and indications. The differences do not affect safety and effectiveness of the device. #### Testing: Performance testing was performed according to the EN 60529:1991 Amendment1:2000 and IEC 601-1 Amendment 1 (1993), IEC 601-1 Amendment 2 (1995). This electrical test has been conducted by PHOENIX TEST-LAB GmbH, Königswinkel, 10 D-32825 Blomberg, Germany. Performance testing was performed according to the EN 300 220-3 V1.1.1 (2000-09) Electromagnetic compatibility and Radio spectrum Matters (ERM): Short range devices (SRD); Radio equipment to be used in the 25 MHz to 1000 Mhz frequency range with power levels ranging up to 500 mW; Part 3 : Harmonized EN covering essential requirements of article 3.2 of the R&TTE Directive. This test has been conducted by PHOENIX TEST-LAB GmbH, Königswinkel, 10 D-32825 Blomberg Germany. Environmental Testing was conducted according to the following standards: - . IEC 601 (1988), EN 60601 - IEC 601-1 Amendment 1 (1991) . - IEC 601-1 Amendment 2 (1995) . - IEC 601-1-1 (1992) . - IEC 601-1-1 Amendment 1 (1995) . {2}------------------------------------------------ - IEC 601-1-2 (1993) . - t IEC 601-2-18 (1996) Electrical and EMC characteristics of the LapMan have been determined to be in compliance with the following standards by SNCH laboratory: - . EN 60601-1-2: 1993 Radiated disturbances (30 MHz to 1 GHz) - . EN 60601-1-2: 1993 Immunity requirements for medical equipment - EN 61000-4-2: 1995 Electrostatic discharge . - EN 61000-4-3: 1996 Radiofrequency electromagnetic field AM . - EN 61000-4-8: 1993 Power frequency magnetic field . The LapMan Laparoscope Manipulator System and Accessories intended to be introduced complies with the CE Declaration of Conformity European Directive 93/42/EC Annex VII. Testing was performed by SNCH / SEE-Certification (Testing & Calibration), Rue de la Gare, 11A, L-8325 CAPELLEN LUXEMBOURG without deviation or adaptation to the standards listed above. No differences were found to exist between the LapMan and the applicable standards. {3}------------------------------------------------ Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with three curved lines representing its wings or feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MedSys S.A. % JWM Associates, LLC Mr. Jeff Morgan Principal Consultant 13723 Dana Lane East Puyallup, Washington 98373 JAN - 3 2007 Re: K062968 Trade/Device Name: The LapMan Laparoscope Manipulator System and Accessories Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: December 1, 2006 Received: December 8, 2006 Dear Mr. Morgan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. Jeff Morgan forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. N elkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K062968 Device Name: The LapMan Laparoscope Manipulator System and Accessories Indications For Use: The LapMan can be used during the laparoscopic procedures listed below and should not be used for any other surgical procedure or purpose. - 1. General surgery: laparoscopic cholecystectomy, laparoscopic Nissen Fundoplication, laparoscopic inguinal hernia repair, laparoscopic gastric banding. - 2. Gynaecology: laparoscopic adnexal surgery (ovarian cystectomy, oophorectomy, adnexectomy, salpingectomy, salpingoneostomy, tubal reanastomosis), laparoscopic management of ectopic pregnancy. laparoscopic treatment of endometriosis, laparoscopic myomectomy, laparoscopic hysterectomy, laparoscopic colposuspension - 3. Urology: laparoscopic prostatectomy. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number 1206296f 14
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...