4K UHD Laparoscope

K203255 · Scivita Medical Technology Co., Ltd. · GCJ · Feb 24, 2021 · Gastroenterology, Urology

Device Facts

Record IDK203255
Device Name4K UHD Laparoscope
ApplicantScivita Medical Technology Co., Ltd.
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 24, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The 4K UHD Laparoscope is intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indicated for visualization of transanal and transvaginal applications.

Device Story

Rigid endoscope; utilizes external light source transmitted via internal lighting system to illuminate surgical site; optical imaging system captures images via objective lens; transmits images to eyepiece for direct viewing or to external camera system for monitor display. Used in OR/clinical settings by physicians for endoscopic surgery. Provides high-resolution visualization to assist in surgical navigation and clinical decision-making; benefits patient through minimally invasive access to thoracic/peritoneal cavities.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), IEC 60601-2-18 (safety/performance), and ISO 8600-1/5 (optical resolution). Image quality performance tests compared depth of field, geometric distortion, color reproduction, SNR, intensity uniformity, frame frequency, and system delay against predicate.

Technological Characteristics

Rigid endoscope; materials include AISI 304 stainless steel, sapphire objective cover, and epoxy resin fiber optic adhesive. Features 5.5mm/10mm diameters, 290-330mm lengths, and 0°/30°/45° view angles. Light source transmitted via fiber optics. Sterilization via steam or low-temperature plasma. No internal software/algorithm; passive optical device.

Indications for Use

Indicated for patients undergoing endoscopic surgery within thoracic and peritoneal cavities, including female reproductive organs, as well as transanal and transvaginal visualization applications.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 24, 2021 Scivita Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.Box 120-119 Shanghai, 200120 China Re: K203255 Trade/Device Name: 4K UHD Laparoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: December 25, 2020 Received: December 30, 2020 Dear Diana Hong: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203255 Device Name 4K UHD Laparoscope Indications for Use (Describe) The 4K UHD Laparoscope is intended to be used for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indication of transanal and transvaginal applications. | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff *PRAStaff@fda.hhs.gov* "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Type of Use (Select one or both, as applicable) Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ # Tab #6 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: _K203255 - 1. Date of Preparation: 12/24/2020 - 2. Sponsor Identification # Scivita Medical Technology Co., Ltd. No. 8, Zhong Tian Xiang, Suzhou Industrial Park, Suzhou, Jiangsu Prov., China. Establishment Registration Number: Not yet registered. Contact Person: Ruqin Wu Position: Quality Manager Tel: +86-512-81877788 Fax: +86-512-85187285 Email: wuruqin@scivitamedical.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person) ### Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: 4K UHD Laparoscope Common Name: Laparoscopes Model: 4K5500, 4K5500R, 4K5530, 4K5530R, 4K5545, 4K5545R, 4K1000, 4K1000R, 4K1030, 4K1030R, 4K1045 and 4K1045R Regulatory Information Classification Name: Laparoscope, General & Plastic Surgery Classification: II; Product Code: GCJ / NMH / HET; Regulation Number: 21 CFR 876.1500 & 21 CFR 884.1720 Review Panel: General & Plastic Surgery Indication for use: The 4K UHD Laparoscope is intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indicated for visualization of transanal and transvaginal applications. Device Description: The 4K UHD Laparoscope is a rigid endoscope, which consists of lighting system and optical imaging system. The 4K UHD Laparoscope has 12 models which are available in two insertion portion widths (5.5 mm and 10 mm), four working lengths (290mm, 300mm, 320mm) and three different directions of view (0°, 30°, 45°). The 12 models are all connected with adaptor 2 for different brand source light. The specifications of the proposed device are listed in Table 1. The 4K UHD Laparoscope can be connected to Scivita 4K UHD Camera System, Scivita LED Light Source or Wolf light source and SONY Monitor. The associated compatible device information is as follows: | Endoscope camera system | 4K UHD Camera System, Model: ES-CS4K100/100C and ES-CS4K200/100C (K200216) | |-------------------------|------------------------------------------------------------------------------------------------------| | Scivita light source | Led Light Source manufactured by Scivita, Model: ES-LS110D (510K Exempt, product code: NTN) | | Wolf light source | Visera 4k UHD Xenon Light Source manufactured by Richard Wolf Medical Instruments Corp., Model: 5123 | | Monitor | Monitor manufactured by SONY, Model: LMD-X2705MD, | {5}------------------------------------------------ ### LMD-X550MD, LMD-X310MD and LMD-X310MT (K150377) The 4K UHD Laparoscope is a reusable device that is cleaned and steam sterilized before first use and each subsequent use. | No. | Model | Diameter | Working Length | Directions of View | |-----|---------|----------|----------------|--------------------| | 1 | 4K5500 | 5.5 mm | 300 mm | 0° | | 2 | 4K5500R | 5.5 mm | 290 mm | 0° | | 3 | 4K5530 | 5.5 mm | 300 mm | 30° | | 4 | 4K5530R | 5.5 mm | 290 mm | 30° | | 5 | 4K5545 | 5.5 mm | 300 mm | 45° | | 6 | 4K5545R | 5.5 mm | 290 mm | 45° | | 7 | 4K1000 | 10 mm | 330 mm | 0° | | 8 | 4K1000R | 10 mm | 320 mm | 0° | | 9 | 4K1030 | 10 mm | 330 mm | 30° | | 10 | 4K1030R | 10 mm | 320 mm | 30° | | 11 | 4K1045 | 10 mm | 330 mm | 45° | | 12 | 4K1045R | 10 mm | 320 mm | 45° | Table 1 Specifications - 5. Identification of Predicate Devices Predicate Device 510(k) Number: K150633 Product Name: "ULTRA" Telescopes Manufacturer: Olympus Winter & Ibe GmbH Models: WA4KL100, WA4KL130, WA4KL145, WA4KL500, WA4KL530 and WA4KL545 - 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - > ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity; - > ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization; - > IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment - > ISO 8600-1:2015 Endoscopes - Medical endoscopes and endotherapy devices -- Part 1: General {6}------------------------------------------------ requirements - > ISO 8600-5:2005 Optics and photonics - Medical endoscopes and endotherapy devices - Part 5: Determination of optical resolution of rigid endoscopes with optics In addition, the image quality performance tests were conducted to quantitatively compare the proposed device and predicate device in terms of depth of field, geometric distortion, color reproduction and SNR, intensity uniformity, image frame frequency and system delay. - 7. Clinical Test Conclusion No clinical study is included in this submission. - 8. Substantially Equivalent (SE) Comparison | ITEM | Proposed Device | Predicate Device K150633 | Remark | |------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Product Code | GCJ / NMH / HET | GCJ / NMH / HET | Same | | Regulation | 21 CFR 876.1500 | 21 CFR 876.1500 | Same | | No. | 21 CFR 884.1720 | 21 CFR 884.1720 | Same | | Class | II | II | Same | | Indication for<br>Use | The 4K UHD Laparoscope is intended to<br>be used for endoscopy and endoscopic<br>surgery within the thoracic and<br>peritoneal cavities including the female<br>reproductive organs. The device is also<br>indicated for visualization of transanal<br>and transvaginal applications. | These endoscopes are intended to<br>be used for endoscopy and<br>endoscopic surgery within the<br>thoracic and peritoneal cavities<br>including the female reproductive<br>organs. The device is also<br>indicated for visualization of<br>transanal and transvaginal<br>applications. | Same | | Single use /<br>Reusable | Reusable | Reusable | Same | | Feature | Rigid endoscope | Rigid endoscope | Same | | Mechanism | Viewing Optics | Viewing Optics | Same | | Principle of<br>operation | During operation, the external light<br>source is transmitted to the front end of<br>the laparoscope through the lighting<br>system of the laparoscope to illuminate<br>the area to be observed. The object is<br>imaged on the imaging system through<br>the objective lens, and then transmitted | During operation, the external<br>light source is transmitted to the<br>front end of the laparoscope<br>through the lighting system of<br>the laparoscope to illuminate the<br>area to be observed. The object<br>is imaged on the imaging system | Same | | | | | | | to the eyepiece for direct observation<br>by the human eye, or magnified by an<br>external camera system and observed<br>on the monitor. | | through the objective lens, and<br>then transmitted to the eyepiece<br>for direct observation by the<br>human eye, or magnified by an<br>external camera system and<br>observed on the monitor. | | | Outer<br>Diameter | 5.5 mm and 10 mm | 5.4 mm and 10 mm | Different | | Working<br>Length | 4K5500, 4K5530, 4K5545: 300mm<br>4K5500R, 4K5530R, 4K5545R: 290mm<br>4K1000, 4K1030, 4K1045: 330mm<br>4K1000R, 4K1030R, 4K1045R: 320mm | WA4KL500: 315.3±0.5mm<br>WA4KL530: 315.5±1.5mm<br>WA4KL545: 315.5±1.5mm<br>WA4KL100: 315.9±0.5mm<br>WA4KL130: 317.1±1.5mm<br>WA4KL145: 317.7±1.5mm | Different | | Direction of<br>View | 0°, 30°, 45° | 0°, 30°, 45° | Same | | Field of View | 75° | 88° | Different | | Depth of Field | 3-200 mm | 6-200 mm | Different | | Patient Contact Material | | | | | Objective<br>glass cover | Sapphire | Sapphire | | | Optical fiber<br>emission<br>surface | Light Guide fibre | Light Guide fibre | Same | | Metal filler | AISI 304 Medical stainless steel | Medical stainless steel | | | Adhesives | Epoxy resin fiber optic adhesive (Epotek<br>353ND) | Epoxy resin fiber optic adhesive | | | Main part | AISI 304 Medical stainless steel | Medical stainless steel | | | Thermal<br>Safety | IEC 60601-2-18 | IEC 60601-2-18 | Same | | Biocompatibility | | | | | Cytotoxicity | Comply with ISO 10993-5 | Comply with ISO 10993-5 | Same | | Sensitization | Comply with ISO 10993-10 | Unknown | Different | | Intracutaneous<br>Reactivity | Comply with ISO 10993-10 | Unknown | Different | | Sterilization | | | | | Method | Steam sterilization / Low-temperature<br>plasma sterilization | Autoclaved sterilization | Different | | Labeling | Complied with 21 CFR Part 801 | Complied with 21 CFR Part 801 | Same | ## Table 2 General Comparison {7}------------------------------------------------ K203255 {8}------------------------------------------------ #### Different - Outer Diameter Both the proposed device and the predicate device provide two options for outer diameter. Although the declared size of the outer diameter of the proposed device is not exactly the same as that of the predicate device, the difference is too slight to cause safety and performance issues. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device. #### Different - Working Length Although the working length of the proposed device is not the same as that of the predicate device, the proposed device offers a wider range of options, allowing physicians to make choices based on the patient's condition. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device. #### Different - Field of View Field of view, namely the view of an endoscope with optics, refers to the maximum range that the endoscope can observe at the same time, that is, the cone angle with the vertex at the end of the endoscope head. Although the field of view of the proposed device is different from that of predicate device, it can meet the needs of clinical diagnosis and treatment. In addition, the field of view of the proposed device meets the requirements of ISO 8600-1 with a deviation of not greater than 15%. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device. #### Different - Depth of Field The proposed device has a wider range of depth of field than the predicate device to give physicians more options for diagnosis and treatment based on the patient's condition. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device. #### Different - Sensitization and Intracutaneous Reactivity Although the results of sensitization and intracutaneous reactivity of the predicate device are unknown, the proposed device did both biocompatibility tests. The results showed that the proposed device posed no risk to humans. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device. #### Different - Sterilization {9}------------------------------------------------ Although the sterilization method of the proposed device is different from that of the predicate device, the sterilization verification is performed for the proposed device. The results showed that the Sterility Assurance Level (SAL) reached 10 after steam sterilization and low-temperature plasma sterilization. Thus, this difference does not affect substantially equivalence between the proposed device and predicate device. - 9. Substantially Equivalent (SE) Conclusion Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...