HOPKINS Telescopes

K221004 · Karl Storz Endoscopy America, Inc. · GCJ · Dec 20, 2022 · Gastroenterology, Urology

Device Facts

Record IDK221004
Device NameHOPKINS Telescopes
ApplicantKarl Storz Endoscopy America, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateDec 20, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K221004 · Dec 20, 2022HOPKINS TelescopesKarl Storz Endoscopy America, Inc.Published clinical literaturePublished literature was used to support the safety and effectiveness of the HOPKINS Telescopes for use in pediatrics during laparoscopy, thoracoscopy, and general surgery.pediatric use; literature review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Published literature; Literature reviewPediatric patientsNot applicable for this studySafety and effectiveness in pediatrics

Indications for Use

The HOPKINS Telescopes are intended to provide visualization during laparoscopy, thoracoscopy and general surgery in adults and pediatrics.

Device Story

Rigid rod-lens telescope; provides visualization during laparoscopy, thoracoscopy, and general surgery. Input: optical image from surgical site via distal lens; transmission/magnification via internal optical glass rods to eyepiece. Output: visual image for surgeon. Used in OR/clinic by physicians. Enhances surgical site visibility to assist clinical decision-making and procedural guidance. Benefits: minimally invasive visualization.

Clinical Evidence

No new clinical trials conducted. Safety and effectiveness for pediatric use in laparoscopy, thoracoscopy, and general surgery supported by published literature.

Technological Characteristics

Rigid rod-lens endoscope; 5mm diameter; 29cm working length; 0°, 25°, 45° direction of view. External light source. Complies with ISO 8600-1, 8600-3, 8600-5, 8600-6 and IEC 60601-2-18:2009. Biocompatibility per ISO 10993-5, 10993-10, 10993-11. Reprocessing validated per AAMI TIR12, TIR30, ST8, ST77, ST79, ST81, and ISO 14937/17655-1. Sterilization: Steam, STERRAD (100S, NX, 100NX), STERIS V-PRO (1, 1 Plus).

Indications for Use

Indicated for visualization during laparoscopy, thoracoscopy, and general surgery in adult and pediatric patient populations.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. December 20, 2022 KARL STORZ Endoscopy America, Inc. Alita McElroy Senior Regulatory Affairs Specialist 2151 E. Grand Ave El Segundo, California 90245 Re: K221004 Trade/Device Name: HOPKINS Telescopes Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: November 19, 2022 Received: November 21, 2022 Dear Alita McElroy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jessica Carr -S for Long Chen, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K221004 Device Name HOPKINS Telescopes Indications for Use (Describe) The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults and pediatrics. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for STORZ. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. ## 510(k) Summary K221004 This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge. | Submitter: | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>EI Segundo, CA 90245 | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Alita McElroy<br>Senior Regulatory Affairs Specialist<br>Phone: (424) 218-8376<br>Email: Alita.McElroy@karlstorz.com | | Date of<br>Preparation: | December 19, 2022 | | Type of 510(k)<br>Submission: | Traditional | | Device<br>Identification: | Trade Name: HOPKINS Telescopes<br>Classification Name: Laparoscope, General and Plastic Surgery (21 CFR<br>876.1500) | | Regulatory Class: | II | | Product Code: | GCJ | | Classification<br>Panel: | General and Plastic Surgery | | Predicate<br>Device(s): | KARL STORZ HOPKINS I and II Rigid Autoclavable Telescopes (K935279) | | Device<br>Description: | The HOPKINS Telescopes are rigid telescopes that utilize the rod lens<br>technology. At the distal end of the telescope's shaft is the lens and the other<br>end of the shaft is attached to the eyepiece. Throughout the central lumen of<br>the HOPKINS Telescopes, optical glass rods are used to transmit and magnify<br>the image received from the lens. The HOPKINS Telescopes are available<br>with 0°, 25° and 45° direction of view, 5mm diameter and 29cm working<br>lengths. | | Intended Use and<br>Indications for<br>Use: | The HOPKINS Telescopes are intended to provide visualization during<br>laparoscopy, thoracoscopy and general surgery in adults and pediatrics. | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for STORZ. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. | Technological<br>Characteristics. | Technological<br>Characteristics | HOPKINS<br>Telescopes | KARL STORZ<br>HOPKINS I and II<br>Rigid Autoclavable<br>Telescopes | Comparison | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------------| | | | Subject<br>Device<br>K221004 | Predicate<br>Device K935279 | | | | Physical Characteristics | | | | | | Endoscope Type | Rigid, rod lens | Rigid, rod lens | same | | | Outer Diameter | 5mm | 4mm-10mm | similar | | | Working Length | 29cm | 29cm-35cm | similar | | | Optical Characteristics | | | | | | Direction of View | 0°, 25°, 45° | 0°, 30° | different | | | Field of View | 68° | 65°- 70° | similar | | | Depth of Field | 15.1mm-200mm | 11.4mm- 200mm | different | | | Light source | External | External | same | | | Reprocessing Methods | | | | | | Cleaning | Manual,<br>Automatic | Manual | different | | | Sterilization<br>modalities | Steam (pre-vacuum),<br>STERRAD<br>100S,<br>STERRAD<br>NX,<br>STERRAD<br>100NX,<br>STERIS V-<br>PRO 1, V-<br>PRO 1 Plus | Steam (pre-vaccum), EO,<br>Chemical<br>Disinfection | different | | Non-Clinical<br>Performance Data: | There are no performance standards or special controls developed under<br>Section 514 of the FD&C Act for endoscopes. However, the HOPKINS<br>Telescopes follows the FDA recognized consensus standards and is tested<br>according to the following standards and FDA Guidance:<br><br>ISO Endoscopic Standards<br>• ISO 8600-1<br>• ISO 8600-3<br>• ISO 8600-5<br>• ISO 8600-6 | | | | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. The logo is simple and professional, and it is likely used to represent the company's brand. | Biocompatibility Summary | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Cytotoxicity (ISO 10993-5) Acute Systemic Toxicity (ISO 10993-11) Intracutaneous Irritation (ISO 10993-10) Maximization Sensitization (ISO 10993-10) | | | Thermal Safety | | | IEC 60601-2-18:2009 (3RD Edition) | | | Reprocessing (Cleaning and Sterilization) | | | AAMI TIR12: 2010 AAMI TIR30: 2011 ANSI/AAMI ST8: 2013 ANSI/AAMI ST77:2013 ANSI/AAMI ST79:2017 ANSI/AAMI ST81:2004/(R)2010 AAMI/ISO 14937:2009 ANSI/AAMI/ISO 17655-1:2006/2013 Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling | | | Comparative bench testing between the subject and predicate device demonstrated that the HOPKINS Telescopes has met all its design specification and is substantially equivalent to its predicate device. | | | Clinical Performance Data: | Published literature was provided to support the safety and effectiveness of the HOPKINS Telescopes for use in pediatrics during laparoscopy, thoracoscopy and general surgery. | | Conclusion: | The conclusions drawn from the nonclinical test demonstrate that the subject device is as safe and effective as the predicate device. |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...