VersaOne Bladed Trocar, VersaOne Bladeless Trocar

K151548 · Covidien · GCJ · Jul 2, 2015 · Gastroenterology, Urology

Device Facts

Record IDK151548
Device NameVersaOne Bladed Trocar, VersaOne Bladeless Trocar
ApplicantCovidien
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 2, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

Use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry

Device Story

VersaOne™ trocars (bladed and bladeless) facilitate endoscopic access; create/maintain port of entry for surgical instruments. Bladed version: obturator with dolphin-nose conical tip, sharp linear blade, spring-loaded locking shield; blade cuts tissue during insertion; shield covers blade upon entry into free space. Bladeless version: obturator with dolphin-nose conical tip; spreads tissue layers for insertion. Both include cannula assembly with seal system, housing, transparent sleeve, stopcock. Seal system maintains pneumoperitoneum; accommodates various instrument diameters. Housing removable for specimen retrieval. Used in OR by surgeons; sterile, single-use. Benefits: smooth insertion, reduced penetration forces, visualization, instrument access.

Clinical Evidence

Bench testing only. Performance testing demonstrated that the VersaOne™ Bladed and Bladeless Trocars meet design specifications and are comparable to predicate devices. Biocompatibility evaluation conducted in accordance with ISO 10993-1.

Technological Characteristics

Sterile, single-use surgical trocars. Bladed: conical tip with linear blade and spring-loaded shield. Bladeless: conical tip for tissue spreading. Cannula assembly includes seal system, housing, transparent sleeve, and stopcock. Materials comply with ISO 10993-1. Mechanical operation; no energy source or software.

Indications for Use

Indicated for use in gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized depiction of a human figure in profile, with three overlapping faces representing the department's focus on health and human well-being. The logo is in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 2, 2015 Covidien Ms. Trang Huynh Senior Specialist, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473 Re: K151548 Trade/Device Name: VersaOne" Bladed Trocar and VersaOne™ Bladeless Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: June 2, 2015 Received: June 9, 2015 Dear Ms. Huynh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use 510(k) Number (if known) K151548 Device Name VersaOne™ Bladed Trocar and VersaOne™ Bladeless Trocar Indications for Use (Describe) indicated for use in a variety of gynecologic, general, thoracic, and urologic endoscopic procedures to create and maintain a port of entry. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (1/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ ## 510(k) Summary This 510(k) summary information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92. | SUBMITTER: | Covidien | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------| | | 60 Middletown Avenue | | | North Haven, CT 06473 USA | | CONTACT PERSON: | Trang Huynh | | | Senior Specialist, Regulatory Affairs | | | Phone: (203) 492-7473 | | | Fax: (203) 492-5029 | | | Email: trang.huynh27@medtronic.com | | DATE PREPARED: | June 2, 2015 | | TRADE/PROPRIETARY<br>NAME: | VersaOne™ Bladed Trocar<br>VersaOne™ Bladeless Trocar | | COMMON/USUAL NAME: | Surgical Trocar | | CLASSIFICATION NAME: | Endoscope and Accessories | | PRODUCT CODE: | GCJ | | FDA PANEL NUMBER: | 78 | | CLASS CODE: | Pursuant to 21 CFR 876.1500, Class II device | | PREDICATE DEVICES: | Autosuture™ Modified Versaport™ Trocar with Fixation Sleeve<br>(K062326) | | | Autosuture™ Versaport™ Plus Bladeless Trocar (K081169) | | | Versaport™ V2 Bladeless Optical (K130435) | | PURPOSE OF<br>SUBMISSION: | Obtain market clearance for VersaOne™ Bladed Trocar 11mm &<br>12mm and VersaOne™ Bladeless Trocar 5mm, 11mm, & 12mm | | INTENDED USE: | Use in a variety of gynecologic, general, thoracic, and urologic<br>endoscopic procedures to create and maintain a port of entry | {4}------------------------------------------------ DEVICE DESCRIPTION: VersaOne™ Bladed Trocar is available in 11mm fixation and smooth standard. 12mm fixation and smooth standard, and 12mm fixation and smooth long. VersaOne™ Bladed Trocar consists of an obturator component with a bladed tip and a cannula assembly. VersaOne™ Bladeless Trocar is available in 5mm fixation short. standard, and long, 11mm fixation standard, 12mm fixation short, standard, and long, and 12mm smooth standard. VersaOne™ Bladeless Trocar consists of an obturator component with a bladeless tip and a cannula assembly. The devices are supplied sterile single-use. TECHNOLOGICAL / VersaOne™ Bladed Trocar obturator has a dolphin nose (conical) DESIGN shaped tip with a sharp linear blade and a spring-loaded locking CHARACTERISTICS: shield. The blade is used to cut the tissues as the trocar is inserted into the body cavity. The shield advances to cover the blade upon entry into a free space. The cannula assembly is composed of a seal system, trocar housing, transparent cannula sleeve, and a stopcock. The seal system prevents loss of Pneumoperitoneum when instruments are inserted or withdrawn. The seal system accommodates instruments ranging from 5mm to 11mm in diameter for 11mm trocars and 5mm to 12mm in diameter for 12mm trocars. The 11mm and 12mm trocar housing is removable for specimen retrieval. The cannula sleeve has a bevel shape at the distal end to reduce penetration forces during insertion and transparent for better visualization. The stopcock valve is for insufflation and desufflation. VersaOne™ Bladeless Trocar obturator has a dolphin nose (conical) shaped bladeless tip. The conical shape is for smooth insertion into the body cavity by spreading the tissue layers during insertion. The cannula assembly is composed of a seal system. trocar housing, transparent cannula sleeve, and a stopcock. The seal system prevents loss of Pneumoperitoneum when instruments are inserted or withdrawn. The seal system accommodates instruments ranging from 5mm to 11mm in diameter for 11mm trocars. 5mm to 12mm in diameter for 12mm trocars, and 5mm in diameter for 5mm trocars. The 11mm and 12mm trocar housing is removable for specimen retrieval. The cannula sleeve has a bevel shape at the distal end to reduce penetration forces during insertion and transparent for better visualization. The stopcock valve is for insufflation and desufflation. The biocompatibility evaluation was conducted in accordance with MATERIALS: ISO 10993-1. Acceptable results demonstrated VersaOne™ Bladed Trocar and VersaOne™ Bladeless Trocar comply with ISO 10993-1 standard. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "COVIDIEN" in blue font. To the left of the word is a blue square with a white plus sign in the center. The plus sign is made up of smaller squares. The image is simple and clean. | PERFORMANCE DATA: | Results from performance testing demonstrated the proposed<br>VersaOne™ Bladed Trocar perform to specifications and<br>comparable to predicate devices K062326 and K130435.<br><br>Results from performance testing demonstrated the proposed<br>VersaOne™ Bladeless Trocar perform to specifications and<br>comparable to predicate devices K081169 and K130435. | |-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CONCLUSION: | VersaOne™ Bladed Trocar and VersaOne™ Bladeless Trocar are<br>substantially equivalent to predicates devices in terms of<br>indications for use, contraindications, technological characteristics,<br>labeling, sterilization, shelf life, biocompatibility, and performance. |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...