FUJIFILM Surgical System

K162836 · Fujifilm New Development U.S.A., Inc. · GCJ · Mar 13, 2017 · Gastroenterology, Urology

Device Facts

Record IDK162836
Device NameFUJIFILM Surgical System
ApplicantFujifilm New Development U.S.A., Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 13, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The FUJIFILM Surgical System includes a laparoscope, video processor and light source and is intended to be used with a monitor, hand instruments, electrosurgical unit, and other ancillary equipment for minimally invasive laparoscopic observation, diagnosis and treatment in general abdominal and gynecologic areas.

Device Story

System comprises EL-580FN Video Laparoscope, VP-4440FN Video Processor, and XL-4450FN Light Source. Laparoscope captures internal body cavity images via distal CCD sensor; transmits signals through cable to processor. Processor relays images to external monitor; controls light intensity. Light source uses Xenon lamp with emergency Halogen backup. Used in clinical settings by healthcare professionals for minimally invasive surgery. Provides real-time visualization of abdominal/gynecologic cavities to assist in diagnosis and treatment. Benefits include high-definition visualization for surgical guidance.

Clinical Evidence

Bench testing only. Compliance with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993 series (biocompatibility), and ISO 8600 series (optical performance) demonstrated. Performance criteria compared to predicate devices; subject device met all acceptance criteria.

Technological Characteristics

Rigid video laparoscope with CCD sensor; Xenon main lamp/Halogen emergency lamp light source. Connectivity via relay connector. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993-1/5/10/11, ISO 8600-1/3/4/5/6, IEC 62471, AAMI ANSI IEC 62304. Software lifecycle per IEC 62304.

Indications for Use

Indicated for minimally invasive laparoscopic observation, diagnosis, and treatment in general abdominal and gynecologic areas in patients requiring such procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 13, 2017 Fujifilm New Development U.S.A., Inc. % Ms. Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, Massachusetts 01864 Re: K162836 Trade/Device Name: Fujifilm Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: February 23, 2017 Received: February 24, 2017 Dear Ms. O'Connell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Image /page/1/Picture/8 description: The image shows the name "Jennifer R. Stevenson -S" in a large, sans-serif font. The text is arranged vertically, with "Jennifer R." on the top line and "Stevenson -S" on the bottom line. The background is a light blue color, and the letters are black. For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162836 Device Name FUJIFILM Surgical System Indications for Use (Describe) The FUJIFILM Surgical System includes a laparoscope, video processor and is intended to be used with a monitor, hand instruments, electrosurgical unit, and other ancillary equipment for minimally invasive laparoscopic observation, diagnosis and treatment in general abdominal and gynecologic areas. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="text-decoration: underline;"></span> </span> | | | Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block; width:20px; height:20px; border:1px solid black; text-align:center;"><b>X</b></div> | | Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block; width:20px; height:20px; border:1px solid black; text-align:center;"></div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) SUMMARY ## FUJIFILM New Development U.S.A. Inc. FUJIFILM Surgical System ### 510(k) Owner FUJIFILM New Development U.S.A, Inc. 318 Bear Hill Road Waltham, MA 02451 Phone: 781-860-2463 Contact Person: Stephen Mariano #### Submission Correspondent: Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 978-207-1245 Phone: Date Prepared:October 6, 2016 ## Trade Name of Device FUJIFILM Surgical System #### Common or Usual Name Laparoscope and Video Processor #### Classification Name Endoscope and Accessories: 21 C.F.R. 876.1500 Class II Product Code: GCJ #### Predicate Devices Fujinon, Inc. EL2-TF410 Video Laparoscope (K983561) FUJIFILM Corp. EPX-4440 HD Digital Video Processor with FICE (K140149) ### Device Description The FUJIFILM Surgical System is comprised of: - . EL-580FN Video Laproscope with RC-4440FN Relay Connector - . EPX-4440FN Digital Video Processor with light source - o VP-4440FN Video Processor - o XL-4450FN Light Source - o DK-4440E Data Keyboard {4}------------------------------------------------ The EL-580FN Video Laparoscope is comprised of a rigid insertion portion, cable portion. video connector and light guide connector. An optical system, CCD image sensor and electrical circuits are located within the distal end portion of the laparoscope. The video signal lines from the CCD sensor and the light guide fiber bundles are connected to the video connector and light guide connector respectively through the laparoscope and video cables. The EL-580FN Video Laparoscope is used with the EPX-4440FN Digital Video processor which consists of the VP-4440FN Processor and the XL-4450FN Light Source. The video connector is connected to the video processor via the RC-4440FN Relay Connector and the light connector is connected to the XL-4450FN Light Source. The VP-4440FN Video Processor relays the image from a laparoscope to a video monitor. The processor also controls the light projected to the body cavity. The XL-4450FN Light Source employs a Xenon lamp with an emergency back-up Halogen lamp. Brightness control is performed by the user. ## Intended Use / Indications for Use The FUJIFILM Surgical System includes a laparoscope, video processor and light source and is intended to be used with a monitor, hand instruments, electrosurgical unit, and other ancillary equipment for minimally invasive laparoscopic observation, diagnosis and treatment in general abdominal and gynecologic areas. ## Substantial Equivalence The FUJIFILM Surgical System is substantially equivalent to Fujinon. Inc.'s EL2-TF410 Video Laparoscope cleared in K983561 and FUJIFILM's EPX-4440 HD Digital Video Processor with FICE cleared in K140149 ("the predicate devices"). The FUJIFILM Surgical System has the same intended use, principles of operation, and similar technological characteristics as the previously cleared predicate devices. The FUJIFILM Surgical System includes a laparoscope, video processor and light source while the predicate devices were cleared in two separate 510(k)s. Fujinon, Inc.'s EL2-TF410 Video Laparoscope is also intended for use during minimally invasive laparoscopic observation, diagnosis and treatment. FUJIFILM's EPX-4440 HD Digital Video Processor with FICE is intended for observation, diagnosis, treatment, and image recording by processing electronic signals transmitted from a video endoscope. All of these devices are prescription devices used by healthcare professionals. Although the subject devices and predicate devices are very similar, there are minor differences in the technological characteristics of the devices. The FUJIFILM Surgical System is comprised of a video laparoscope and a video processor with light source. The video laparoscope has similar technological characteristics to the EL2-TF410 Video Laparoscope in that both consist of a rigid portion, cable portion, image connector and light connector. Identical characteristics include direction of view, image sensor type, resolution, and sterility requirements. Each difference between the subject device and predicate device has been evaluated and determined not to raise new questions of safety or effectiveness. {5}------------------------------------------------ The VP-4440FN Video Processor and XL-4450FN Light Source have the same technological characteristics as the EPX-4440 HD Digital Processor with FICE also made by FUJIFILM. Both systems have a Xenon lamp as the main lamp with a Halogen lamp as the emergency lamp. All technical specifications relevant to the indications for use of the subject device and predicate device are identical. Performance testing was performed which compared the subject and predicate devices and demonstrates substantial equivalence. Therefore, the FUJIFILM Surgical System is substantially equivalent to the predicate devices. # Performance Data Bench testing was conducted in order to demonstrate compliance with recognized consensus standards: - . ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2007: Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility -Requirements and tests - . IEC 60601-2-18 Edition 3.0 2009-08 Medical electrical equipment-Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equinment - ISO 10993-1:2009 Biological evaluation of medical devices. Evaluation and testing within a risk management process. - AAMI ANSI ISO 10993-5: Biological evaluation of medical devices Part 5: Tests . for in vitro cytotoxicity - . ISO 10993-10: Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization - . ISO 10993-11: Biological evaluation of medical devices-Part 11: Tests for systemic toxicity - ISO 8600-1 Fourth edition Optics and photonics-Medical endoscopes and endotherapy devices-Part 1: General requirements - ISO 8600-3 First edition 1997-07-01 Optics and optical instruments - Medical endoscopes and endoscope accessories - Part 3: Determination of field of view and direction of view of endoscopes with optics (Including: Amendment 1 (2003) - . ISO 8600-4 Second edition 2014-03-15 Ontics and optical instruments-Medical endoscopes and certain accessories-Part 4: Determination of maximum width of insertion portion - ISO 8600-5 First Edition 2005-03-15 Optics and photonics - Medical endoscopes and endotherapy devices - Part 5: Determination of optical resolution of rigid endoscopes with optics - ISO 8600-6 First edition 2005-03-15 Optics and photonics - Medical endoscopes and endotherapy devices - Part 6: Vocabulary {6}------------------------------------------------ - . ISO 14971:2007, Medical Device Risk Management-Part 1: Application of risk management to medical devices - . IEC 62471: 2006 Photobiological safety of lamps and lamp systems - AAMI ANSI IEC 62304:2006 Medical device software. Software lifecycle processes. ● Additionally, performance criteria were compared between the subject device and the predicate device and the predicate device was determined to be substantially equivalent to the predicate. The EL-580FN Video Laparoscope and the predicate device, the Fujinon EL2-TF410 Laparoscope (K983561) met all the acceptance criteria. Substantial equivalence of the two devices was demonstrated in terms of the technological characteristics. The conclusions drawn from the non-clinical testing demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...