K123325 · Medafor, Inc. · GCJ · Jan 24, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K123325
Device Name
Medafor Direct Gas-Assisted Application System
Applicant
Medafor, Inc.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Jan 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The device is intended to assist the delivery of a powdered hemostatic agent to the treatment site in surgical procedures including endoscopic surgeries, using a 5 mm or larger trocar.
Device Story
System delivers powdered hemostatic agents to bleeding tissue during surgery, including laparoscopic/endoscopic procedures. Components include a reusable gas regulator and foot pedal, and a sterile, single-use applicator kit (tubing, adapter, rigid applicator tip). Operates by connecting to a CO2 gas source; user-controlled via foot pedal to create a gas stream that fluidizes and sprays the powder onto the treatment site. Used in OR settings by surgeons. Output is a controlled spray of hemostatic powder, assisting in blood clearing and hemostasis at the surgical site. Benefits include precise, localized delivery of hemostatic agents in minimally invasive environments.
Clinical Evidence
Bench and animal testing only. Mechanical testing verified trocar compatibility, delivery control, blood clearing, system leak testing, and pressure/flow monitoring. System verification testing performed in swine confirmed output meets input requirements under simulated conditions. No human clinical data presented.
Technological Characteristics
Materials: stainless steel and biosafe polymers. Energy source: external CO2 gas supply. Form factor: regulator, foot pedal, and disposable tubing/applicator kit. Connectivity: luer lock gas connections. Sterilization: ethylene oxide (disposable components).
Indications for Use
Indicated for patients undergoing surgical procedures, including endoscopic/laparoscopic surgeries, requiring the application of a powdered hemostatic agent via a 5 mm or larger trocar.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
FibriJet Air Assisted Endoscopic Applicator (K042834)
Related Devices
K162197 — EndoClot · Endoclot Plus Co., Ltd. · Jan 18, 2017
K123847 — MALLEABLE TIP ENDOSCOPIC APPLICATOR · Micromedics, Inc. · Jan 8, 2013
K172672 — Endoscopic Applicator, 41 cm · Micromedics, Inc. (D/B/A Nordson Medical) · Oct 6, 2017
K241739 — Endoscopic Applicator, 41 cm · Micromedics, Inc. · Jul 15, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
# JAN 2 4 2013
#### Date Prepared: January 9, 2013
#### Submitter Information:
| Submitter's Name/Address | Contact Person |
|------------------------------------|-----------------------------------------------------|
| Medafor | Melissa Young, Senior Regulatory Affairs Specialist |
| 2700 Freeway Boulevard, Suite 800. | |
| Minneapolis, MN 55430 | |
| Telephone: 763-231-8201 | |
| Fax: 763-571-1035 | |
## Device Information:
| Trade Name | MEDAFOR DIRECT Gas-Assisted Application System - MEDAFOR DIRECT™<br>Pressurized Gas Regulator and DIRECT Kit™ Gas-Assisted Applicator |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Gas Regulator and Gas-Assisted Spray Kit |
| Classification Name | Endoscope and Accessories |
| Product Code | GCJ |
| Regulation | 876.1500 |
| Panel | General and Plastic Surgery |
## Performance Standards:
No performance standards applicable to this product have been developed under Section 514 of the Act.
#### Predicate Device:
| Predicate Device | Manufacturer | 510(k) No. |
|---------------------------------------------|-------------------|------------|
| FibriJet Air Assisted Endoscopic Applicator | Micromedics, Inc. | K042834 |
#### Device Description:
The MEDAFOR DIRECT Gas-Assisted Application System is intended to be used on patients undergoing surgery including laparoscopic procedures. The gas-assisted applicator kit and regulator are provided to assist the user in the application of powdered hemostatic agents to bleeding tissue in surgical procedures including laparoscopic/endoscopic surgeries, using a 5 mm or larger trocar.
The following table describes the MEDAFOR DIRECT Gas-Assisted Application System that is intended to be commercially marketed. The device system consists of the following components in the following packaging configurations:
Page lot 3
{1}------------------------------------------------
| MEDAFOR DIRECT Configurations and Components | | |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| No. | Packaging Configuration | Components Included |
| 1. | Regulator Kit: Packaged in a corrugated box | Gas Regulator |
| | | Foot pedal |
| 2. | Gas Assisted Applicator Kit is supplied as a sterile, single use kit packaged in a preformed PETG tray and Tyvek®¹ lid | Tubing set |
| | | Adapter Handle |
| | | FlexiTip XL-R rigid applicator |
The gas regulator and foot pedal are supplied separately as non-sterile, reusable components. The Gas-Assisted Applicator kit is provided as a sterile, disposable unit, consisting of the tubing set, applicator adapter and a single unit of FlexiTip XL-R applicator. These disposable components are kitted together in a preformed tray and Tyvek® lid and terminally sterilized by ethylene oxide.
# Intended Use/Indications for Use:
The device is intended to assist the delivery of a powdered hemostatic agent to the treatment site in surgical procedures including endoscopic surgeries, using a 5 mm or larger trocar.
# Summary of Non-Clinical Testing:
The MEDAFOR DIRECT Gas-Assisted Application System underwent mechanical, performance, and biocompatibility assessments to verify that the device functions in a safe and effective manner.
The mechanical tests performed on the MEDAFOR DIRECT Gas-Assisted Application System include:
| Critical Parameter | Mechanical Test |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Interface<br>Compatibility | Trocar luer compatibility |
| | Trocar 5 mm cannula compatibility |
| | CO2 tank connector compatibility |
| Device Delivery | Container content delivery of hemostatic agent in powdered form |
| | Delivery Control (location and area) |
| | Blood clearing from site |
| | System Leak Testing (Maintaining Pneumoperitinum) |
| Device Functionality | Pressure Monitoring |
| | Flow Control & Monitoring |
| | System Performance Testing (Internal device component compatibility (gas tank<br>tubing to regulator/tubing to adapter/adapter to FlexiTip XL-R ) |
1 Tyvek® is a registered trademark of DuPont
{2}------------------------------------------------
| Critical Parameter | Mechanical Test |
|--------------------|----------------------------------------------------------------------------------------------------------------------|
| | System Verification Test in Swine: Performed as intended under simulated conditions<br>to confirm output meets input |
All tests met acceptance criteria. These results provide assurance that the device has been designed and evaluated to assure conformance to the requirements for its intended use.
# Substantial Equivalence Comparison
The MEDAFOR DIRECT Gas Assisted Application System is substantially equivalent to the predicate device based on a comparison of the indications for use and the technological The technological characteristics include design configuration, materials, characteristics. method of sterilization, compatibility with the same ancillary devices (such as a standard 5 mm trocar) and method of operation.
While the subject and predicate device are intended to deliver materials to the patient during endoscopic surgery, they differ in the types of material to be delivered. The predicate device is intended to apply two nonhomogeneous liquids, while the subject device delivers a powder hemostat to the treatment site. However, when both devices are activated, the compressed gas fluidizes the liquid and powdered hemostat in the same manner, forcing them into an air stream in the form of a spray when exiting the applicator system.
Both subject and predicate devices are made of stainless steel and biosafe polymer materials and are terminally sterilized by ethylene oxide.
Both systems include components consistent with that used in endoscopic gas-assisted systems (regulator, footswitch, applicator/tubing set), however the applicator tubing set design and connections differ. The predicate applicator is designed as a dual path system that allows connection of two syringes by luer lock at the distal end. At the proximal end, the applicator is attached by threaded connection to the Replaceable tips. In contrast, the subject applicator has a simpler single path design that connects to a hemostatic powder container by friction fit at the distal end and does not require any attachment at the proximal end. Both device tubing sets attach to the gas supply by luer lock connections. The testing performed confirms that the MEDAFOR DIRECT Gas- Assisted Application System does will perform as intended, safely and effectively.
#### Conclusion
Based on the results from the bench and animal testing performed and the substantial equivalence comparison, the MEDAFOR DIRECT Gas-Assisted Application System does not raise new questions of safety and effectiveness when compared to the legally marketed predicate device and is, therefore, substantially equivalent.
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of an eagle or bird-like figure. The logo is in black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medafor, Incorporated % Ms. Melissa Young Senior Regulatory Affairs Specialist 2700 Freeway Boulevard, Suite 800 Minneapolis, Minnesota 55430
January 24, 2013
Re: K123325
Trade/Device Name: MEDAFOR DIRECT Gas-Assisted Application System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: January 09, 2013 Received: January 10, 2013
Dear Ms. Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Ms. Melissa Young
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): ____K123325_________________________________________________________________________________________________________________________________________
MEDAFOR DIRECT Gas-Assisted Application System Device Name:
#### Indications for Use:
The device is intended to assist the delivery of a powdered hemostatic agent to the treatment site in surgical procedures including endoscopic surgeries, using a 5 mm or larger trocar.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR :
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Neil R Ogden 2013.01.18 14:06:00 -05'00'
(Division Sign-Off) Division of Surgical Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.