HEAD MOUNTED DISPLAY

K961800 · Vista Medical Technologies, Inc. · GCJ · Sep 11, 1996 · Gastroenterology, Urology

Device Facts

Record IDK961800
Device NameHEAD MOUNTED DISPLAY
ApplicantVista Medical Technologies, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Vista Head Mounted Display System is designed to display video images while mounted on the users head.

Device Story

Head-mounted display (HMD) system for endoscopic surgery; provides binocular field display to allow surgeon to view surgical site and operating field simultaneously. Inputs: standard NTSC or PAL video signals (CVBS or S-Video) from endoscopic cameras. Processing: controller unit drives active matrix liquid crystal displays (LCDs) mounted on headgear. Output: visual display of endoscopic images directly in surgeon's field of view. Used in OR by surgeons; allows viewing without turning away from surgical site. Benefits: maintains surgeon's focus on patient; provides natural 'look down' visualization; supports simultaneous use of up to four HMDs and two standard monitors. Serves as backup to direct visualization or standard 2D monitors.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Active matrix liquid crystal display (LCD) technology; head-mounted form factor; controller station for signal processing; supports NTSC/PAL, CVBS, and S-Video (Y,C) inputs; wired connection between HMD and controller; supports monocular and stereo endoscope inputs; allows use of corrective eyewear.

Indications for Use

Indicated for use by surgeons as a component of a visualization system during endoscopic surgery to display video images.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} SEP 11 1996 K961800 510(K) SUMMARY July 31, 1996 COMPANY NAME AND ADDRESS Vista Medical Technologies 5451 Avenida Encinas, Suite A Carlsbad, CA. 92008 CONTACT PERSON Debbie Iampietro Director of Regulatory Affairs and Quality Assurance Telephone (508) 366-8882 Fax: (508) 366-8858 DEVICE TRADE NAME Head Mounted Display COMMON NAME Stereo Viewing Monitor Display for Endoscopic Visualization PREDICATE DEVICE 1. Device Name: 3DSCOPE™ Three Dimensional Medical Video System Classification: Camera, Television, Endoscopic, without audio Manufacturer: American Surgical Technologies 300 Billerica Rd Chelmsford, MA 01824 510(k) #: K923160 2. Device Name: Crystal Eyes View/Record and Playback Controllers with Eyewear and Emitters Classification: Endoscope and Accessories, Class II per 21 CFR 876.1500 Manufacturer: Oktas 134 Flanders Rd. Westborough, MA 01581 510(k) #: K944254 {2} # DEVICE DESCRIPTION The Head Mounted Display (HMD) System is a binocular field display, occupying a large portion of the normal straight ahead field of view. This display is intended to be used as a component of a visualization system for use in endoscopic surgery. Traditional video-endoscopic surgery images are displayed on a TV monitor. This requires the surgeon to maintain contact with the traditional monitor and remain in position with it at all times. Having to turn away from the monitor causes the surgeon to lose sight of what is being accessed. The HMD system enables the surgeon to see the surgical site as well as the general operating field with the lateral and 45 degree look down vision, without having to look away, thus providing the means to present the endoscopic picture in a more natural and acceptable look down location with respect to the patient. This also provides a safe back up, the physician can resort to either direct visualization or utilize the back up 2D video monitor. The HMD takes a standard signal and displays it on two small electronic image display devices called liquid crystal display (LCD). It can be connected to any standard video generating equipment. The HMD consists of an image generating array, eyepiece, and electronics. The image generating array is an active matrix liquid crystal display (LCD). These elements are mounted on a lightweight and compact headgear which fits comfortably on the user's head. The remaining electronics are housed separately in a controller station. The HMD communicates with the controller by a lightweight, durable, flexible cable. The HMD can be driven by both monocular and stereo endoscopes. The HMD Controller will accept NTSC, PAL with CVBS, or S Video (Y,C) video input. Interfaces to external and remote data sets, power supplies and drivers for the LCD and backlight are included in the controller. A maximum of four (4) HMDs and two (2) standard video monitors can be driven simultaneously with the single controller. # INTENDED USE The Vista Head Mounted Display System is designed to display video images while mounted on the users head. {3} # DEVICE COMPARISON The HMD system is similar to the display component of the 3DSCOPE™, Three Dimensional Medical Video System which received clearance from FDA in December 1992, (K923160). The 3DSCOPE™ Stereo Display is a high resolution, high speed, multi-synch RGB full color video display monitor. The liquid crystal screen switches its circular polarization from left to right in precise synchronization (120 Hz) with the left and right images that appear on the Stereo Display. When the left image appears on the Stereo Display, the liquid crystal shutter switches to the left polarized state. The HMD system accepts mono or binocular input from the endoscope and passes the input to the LCD matrix. The small number of elements, compactness and low weight enable this technology to be incorporated into a head mounted display. Most endoscopes are optical and attached to the video system which can feed the video electronic imager rather than an optical image from the endoscope itself to the HMD. Corrective eyewear may be worn with this device as the HMD provides large eye relief.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...