Connected OR Hub with Device and Voice Control; SDC4K Information Management System with Device and Voice Control
Applicant
Stryker Endoscopy
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Aug 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
Connected OR Hub with Device and Voice Control: The use of the Connected OR Hub with Device and Voice Control is to allow for remote control and voice control of medical device settings by surgeons or operating room personnel, thereby eliminating the need to manually operate those devices compatible with the Connected OR Hub or to rely on verbal communication between the surgeon and other operating room personnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (non-medical device function), which is independent of the functions or parameters of any attached Stryker device. SDC4K Information Management System with Device and Voice Control: The use of the SDC4K Information Management System with Device and Voice Control is to allow for remote control and voice control of medical device settings by surgeons or operating room personnel, thereby eliminating the need to manually operate those devices compatible with the SDC4K Information Management System with Device and Voice Control or to rely on verbal communication between the surgeon and other operating room personnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (non-medical device function), which is independent of the functions or parameters of any attached Stryker device.
Device Story
Network-compatible hardware platform (Connected OR Hub/SDC4K) acting as Medical Device Data System (MDDS); enables remote/voice control of compatible surgical equipment (cameras, light sources, insufflators, pumps, RF probes, shavers, drills, ultrasonic aspirators). Inputs: touchscreen GUI, wireless headset (voice commands), IR remote, camera head keypad. System processes commands and communicates with connected devices via wired (USB/SORN/Ethernet) or wireless (Connected OR Spoke) links. VIP package automates image enhancement and smoke removal. Used in OR by surgeons/staff; centralizes foot pedal assignments. Benefits: reduces manual handling, improves OR workflow efficiency, eliminates reliance on verbal communication for equipment adjustment.
Clinical Evidence
Bench testing only. No clinical data. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless technology (AAMI TIR69, ANSI C63.27), reprocessing (ISO 17664-2), software verification/validation (IEC 62304), cybersecurity, usability (IEC 62366-1), and performance bench testing (video compatibility, voice recognition, system design validation). All tests passed.
Technological Characteristics
Hardware console with touchscreen GUI; wired (USB, Ethernet/SORN) and wireless (WiFi 802.11a/b/g/n/ac, DECT, IR) connectivity. Power: 100-240VAC. Standards: IEC 60601-1, IEC 60601-1-2. Software: Windows 10. Features: MDDS, device/voice control, video image processing. Non-patient contacting.
Indications for Use
Indicated for surgeons and operating room personnel to remotely or via voice control adjust settings of compatible medical devices, eliminating manual operation or verbal communication. Includes digital documentation for capturing, transferring, storing, and displaying medical device data.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
Connected OR Hub with Device and Voice Control (K232157)
SDC4K Information Management System with Device and Voice Control (K230886)
SDC4K Information Management System with Device and Voice Control (K222079)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 15, 2024
Stryker Endoscopy Sekar Divya Principal Regulatory Affairs Specialist 5900 Optical Ct San Jose, California 95138
### Re: K241401
Trade/Device Name: Connected OR Hub with Device and Voice Control; SDC4K Information Management System with Device and Voice Control Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, HRX Dated: May 16, 2024 Received: May 17, 2024
### Dear Sekar Divya:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long H. Chen -S Digitally signed by Long H.Chen -S
Date: 2024.08.15 14:20:11 -04'00'
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241401
#### Device Name
Connected OR Hub with Device and Voice Control; SDC4K Information Management System with Device and Voice Control
#### Indications for Use (Describe)
Connected OR Hub with Device and Voice Control:
The use of the Connected OR Hub with Device and Voice Control is to allow for voice control and remote control of medical device settings by surgeons or operating room personnel, thereby eliminating the need to manually operate those devices compatible with the Connected OR Hub with Device and Voice Control or to rely on verbal communication between the surgeon and other operating room personnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (non-medical device function), which is independent of the functions or parameters of any attached Stryker device.
SDC4K Information Management System with Device and Voice Control:
The use of the SDC4K Information Management System with Device and Voice Control is to allow for voice control and remote control of medical device settings by surgeons or operating room personnel, thereby eliminating the need to manually operate those devices compatible with the SDC4K Information Management System with Device and Voice Control or to rely on verbal communication between the surgeon and other operating room personnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (non-medical device function), which is independent of the functions or parameters of any attached Stryker device.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|-------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) summary is submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).
# 510(k) Number: K241401
#### Submitter:
| Applicant: | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 |
|-----------------|-----------------------------------------------------------------------------------------------------------------|
| Contact Person: | Divya Sekar<br>Principal Regulatory Affairs Specialist<br>Email: divya.sekar@stryker.com<br>Phone: 408-754-2473 |
| Date Prepared: | August 14, 2024 |
### Subject Device:
| Name of Device: | Connected OR Hub with Device and Voice Control;<br>SDC4K Information Management System with Device and Voice Control |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name | Endoscope and accessories |
| Classification Name: | Laparoscope, General & Plastic Surgery (21 C.F.R. §876.1500) |
| Subsequent Classification Name: | Arthroscope, Orthopedic (21 C.F.R. §888.1100) |
| Regulatory Class: | II |
| Product Code: | GCJ |
| Subsequent Product Code | HRX |
| 510(k) Review Panel: | General and Plastic Surgery |
# Predicate Devices:
| Connected OR Hub with Device and Voice Control | K232157 (primary) |
|-------------------------------------------------------------------|-------------------|
| SDC4K Information Management System with Device and Voice Control | K230886, K222079 |
NOTE: The predicate devices have not been subject to a design-related recall.
{4}------------------------------------------------
## Device Description:
The Connected OR Hub with Device and Voice Control and SDC4K Information Management System with Device and Voice Control are network compatible hardware platforms that carry out Medical Device Data System (MDDS) functionalities and allows the user to control the state, selection, and settings of compatible connected devices both wired and wirelessly.
The Connected OR Hub with Device and Voice Control and SDC4K Information Management System with Device and Voice Control consists of the following components:
- 1) Base Console which includes:
- a) Medical Device Data System (MDDS) functionalities
- b) Optional Device Control feature
- c) Optional Voice Control feature
- d) Optional Video Image Processing (VIP) feature
- 2) Device Control Package (software activation USB dongle and a handheld Infrared (IR) remote control)
- 3) Voice Control Package (software activation USB dongle and a wireless headset and base station)
- 4) Video Image Processing package (software activation USB dongle)
- 5) Connected OR Spoke (MDDS)
The Connected OR Hub and SDC4K consoles carry out the Medical Device Data System (MDDS) functionalities (i.e. Non-medical function) and can be marketed as standalone devices. When upgraded with the Device Control and/or Voice Control package, the consoles extend their functionalities to control compatible devices from their touchscreen graphical user interface (GUI), spoken commands via headset (voice control input), and an IR remote control or directional keypad from a camera head (device control input). The received user commands are then processed and communicated with the connected controllable devices, allowing the user to control the state, selection, and settings of those devices. The Connected OR Hub and SDC4K centrally manage foot pedal assignments for the Surgical Drill System and Ultrasonic Aspirator System handpieces, utilizing device or voice control inputs from the systems or directly from the connected foot pedal.
In addition, the Connected OR Hub with Device and Voice Control and SDC4K Information Management System with Device and Voice Control also provide compatibility with the Connected OR Spoke (also referred to as "Spoke") which is a standalone Class I Medical Device Data System. Once the Connected OR Hub with Device and Voice Control or SDC4K Information Management System with Device and Voice Control is connected to the Spoke, Device Control can be extended to compatible devices which are directly connected to the Spoke. Ethernet-compatible devices are connected to the Secondary Spoke via an ethernet cable/switch.
When upgraded with the Video Image Processing (VIP) package, the Connected OR Hub automates an image enhancing algorithm, and removal of surgical smoke through a compatible insufflator.
{5}------------------------------------------------
# Intended Use/Indications for Use:
| Subject Devices | Predicate Devices |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Connected OR Hub with Device and Voice Control;<br>SDC4K Information Management System with Device and Voice Control | Connected OR Hub with Device and<br>Voice Control;<br>SDC4K Information Management<br>System with Device and Voice Control |
| NOTE: Connected OR Hub with Device and Voice Control and SDC4K<br>Information Management System with Device and Voice Control have the<br>same Intended Use/Indications for Use; only the names of the device differ. | |
| Connected OR Hub with Device and Voice Control:<br>The use of the Connected OR Hub with Device and Voice Control is to allow<br>for remote control and voice control of medical device settings by surgeons or<br>operating room personnel, thereby eliminating the need to manually operate<br>those devices compatible with the Connected OR Hub or to rely on verbal<br>communication between the surgeon and other operating room personnel in<br>order to adjust the surgical equipment. It also has additional digital<br>documentation functionality to electronically capture, transfer, store and<br>display medical device data (non-medical device function), which is<br>independent of the functions or parameters of any attached Stryker device. | K232157 (primary), K230886, K222079<br>Same as subject device (Connected OR<br>Hub with Device and Voice Control) |
| SDC4K Information Management System with Device and Voice<br>Control:<br>The use of the SDC4K Information Management System with Device and<br>Voice Control is to allow for remote control and voice control of medical<br>device settings by surgeons or operating room personnel, thereby eliminating<br>the need to manually operate those devices compatible with the SDC4K<br>Information Management System with Device and Voice Control or to rely on<br>verbal communication between the surgeon and other operating room<br>personnel in order to adjust the surgical equipment. It also has additional<br>digital documentation functionality to electronically capture, transfer, store<br>and display medical device data (non-medical device function), which is<br>independent of the functions or parameters of any attached Stryker device. | |
# Comparison of Technological Characteristics with the Predicate Device:
| Item | Subject Devices | Predicate Devices | |
|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Connected OR Hub with Device and Voice<br>Control; SDC4K Information Management<br>System with Device and Voice Control | Connected OR Hub with Device and Voice<br>Control; SDC4K Information<br>Management System with Device and<br>Voice Control | |
| Manufacturer | Stryker | Same as subject devices | |
| Submission Reference | Current Submission | K232157 (primary), K230886, K222079 | |
| Principles of Operation | Use of IR remote control for device control<br>and RF communication for voice control of<br>connected devices. | Same as subject devices | |
| Device Components | Console (Connected OR Hub & SDC4K) Device Control Package Voice Control Package Connected OR Hub VIP Package Connected OR Spoke | Same as subject devices | |
| Item | | Subject Devices<br>Connected OR Hub with Device and Voice Control; SDC4K Information Management<br>System with Device and Voice Control | Predicate Devices<br>Connected OR Hub with Device and Voice Control; SDC4K Information<br>Management System with Device and<br>Voice Control |
| Feature(s) | Documentation<br>Functionalities<br>(Non-Medical<br>Device<br>functionalities) | Gathering patient demographic data, Capture,<br>Record, Transfer, Display image/video of<br>various formats, Archiving information | Same as subject devices |
| | Device Control | Remote control of compatible medical device<br>settings | Same as subject devices |
| | Voice Control | Voice control of compatible medical device<br>settings | Same as subject devices |
| | Video Image<br>Processing (VIP) | Smoke Detection Enhanced Imaging Smoke Evacuation NOTE: The VIP feature is not available on the<br>SDC4K. | Same as subject devices |
| | Device Control User<br>Interface | Capacitive Graphical User Interface on<br>LCD touchscreen Voice recognition and control via<br>wireless headset Device Control via IR remote control and<br>camera head directional keypad | Same as subject devices |
| | Connection to Controllable<br>Devices | Wired connection:<br>Using USB port: The controllable devices<br>are connected to the console's USB<br>port(s) via USB cable. Using SORN port: The Connected OR Spoke is<br>connected to the console's SORN<br>port via ethernet cable, while the<br>controllable devices are connected to<br>the Connected OR Spoke via USB<br>and/or firewire cables. The controllable devices are<br>connected to the console's SORN<br>port via ethernet cable/switch. | Wired connection:<br>Using USB port: The controllable<br>devices are connected to the console's<br>USB port(s) via USB cable. Using SORN port: The Connected OR Spoke is<br>connected to the console's<br>SORN port via ethernet cable,<br>while the controllable devices<br>are connected to the Connected<br>OR Spoke via USB and/or<br>firewire cables. |
| | | Wireless connection:<br>The primary Connected OR spoke is<br>connected to the console's SORN port via<br>an ethernet cable, while controllable<br>devices at remote locations within the<br>same OR are connected to the secondary<br>Connected OR Spoke via USB and/or<br>firewire cables. Ethernet- compatible<br>devices are connected to the Secondary<br>Spoke via an ethernet cable/switch. The<br>primary and secondary Spoke act as the<br>wireless transfer medium to transfer<br>device control data to / from the console. | Wireless connection:<br>The primary Connected OR spoke is<br>connected to the console's SORN port<br>via an ethernet cable, while controllable<br>devices at remote locations within the<br>same OR are connected to the secondary<br>Connected OR Spoke via device control<br>cables. The primary and secondary<br>Spoke act as the wireless transfer<br>medium to transfer device control data to<br>/ from the console. |
| Item | Subject Devices | Predicate Devices | |
| | Connected OR Hub with Device and Voice<br>Control; SDC4K Information Management<br>System with Device and Voice Control | Connected OR Hub with Device and Voice<br>Control; SDC4K Information<br>Management System with Device and<br>Voice Control | |
| Controllable Devices | Class II Devices:<br>● Surgical Cameras (K132785, K211202,<br>K212511, K214046, K231854, K233327,<br>K240174)<br>● Light Sources (K142310, K151243,<br>K192292, K214046, K221611, K231854)<br>● Insufflators (K063367, K170784,<br>K201361)<br>● Irrigation Pumps (K123441, K191259)<br>● RF Probes and Shaver System (K071859,<br>K160050, K171391)<br>● Wireless Monitor (K081995)<br>● Surgical Drill System (K171840) | Class II Devices:<br>● Surgical Cameras (K132785, K211202,<br>K212511, K214046, K231854,<br>K233327, K240174)<br>● Light Sources (K142310, K151243,<br>K192292, K214046, K221611,<br>K231854)<br>● Insufflators (K063367, K170784,<br>K201361)<br>● Irrigation Pumps (K123441, K191259)<br>● RF Probes and Shaver System<br>(K071859, K160050, K171391)<br>● Wireless Monitor (K081995) | |
| | Unclassified Devices:<br>● Ultrasonic Aspirator System (K213824,<br>K190070) | Class I/ II 510(k) exempt devices:<br>● Ceiling Mounted Room Lights (Class II,<br>Product Code: FSY) | |
| | Class I/ II 510(k) exempt devices:<br>● Ceiling Mounted Room Lights (Class II,<br>Product Code: FSY)<br>● Wired Monitor (Class I device) | ● Wired Monitor (Class I device) | |
| Embedded Software Design | Microsoft Windows 10 | Same as subject device | |
| Electronic Circuit Design | ● Custom designed chipset, storage solution<br>and Capture Card.<br>● CD/DVD drive: Not included in chassis<br>● On-board storage: Hard Disk Drive<br>(HDD) and Solid-State Drive (SSD) | Same as subject device | |
| Video Input and Output | (Connected OR Hub with Device and Voice<br>Control)<br>Input: DVI, RGBHV and HDMI<br>Output: DVI, HDMI<br><br>(SDC4K Information Management System<br>with Device and Voice Control)<br>Input: HDMI<br>Output: HDMI | Same as subject device | |
| Communication Protocol(s) | Wired: SIDNE, DCM<br>Wireless: DCM, SIDNE, SFB | Same as subject device | |
| Data Transfer,<br>Documentation and Storage<br>(Non-Medical<br>functionality) | ● Wireless Standard: WLAN<br>802.11a/b/g/n/ac<br>● Frequency: 2.4GHz and 5GHz | Same as subject device | |
| Wireless technology for<br>Device and Voice Control | ● Voice Control headset (DECT<br>technology)<br>● IR Remote (Infrared)<br>● Connected OR Spoke (WiFi) | Same as subject device | |
| Power rating | ● 100-240VAC ~50/60 Hz, 4A/2A maximum | Same as subject device | |
| Electrical Safety | ● ANSI/AAMI ES60601-1 | Same as subject device | |
| EMC | ● IEC 60601-1-2 | Same as subject device | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Summary of technological characteristics:
The subject and the predicate devices (Connected OR Hub with Device and Voice Control and SDC4K information Management System with Device and Voice Control) have the same following technological characteristics (i.e., Principles of operation, Components, User interface, Software design, Circuit design, Video input and output, Communication protocol(s), Wireless technology and Electrical safety / EMC) except for the inter device communication with additional Stryker devices via ethernet-based communication.
#### Description of nonclinical and clinical tests:
The subject devices (Connected OR Hub with Device and Voice Control and SDC4K information Management System with Device and Voice Control) have been evaluated and conform with recognized standards, device input specifications, user needs and intended uses. A summary of testing is provided in the table below:
| Test | Method | Results |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electrical Safety | • In accordance with FDA-recognized voluntary consensus standard IEC<br>60601-1:2020 (19-49) | Pass |
| EMC | • In accordance with FDA-recognized voluntary consensus standard IEC<br>60601-1-2:2014+A1:2020 (19-36) | Pass |
| Wireless Technology | • In accordance with FDA-recognized voluntary consensus standard<br>AAMI TIR69:2017/(R2020) (19-22)<br>• In accordance with FDA-recognized voluntary consensus standard<br>ANSI C63.27:2017 (19-29) | Pass |
| Reprocessing | • In accordance with FDA-recognized voluntary consensus standard ISO<br>17664-2:2021 (14-579)<br>• In accordance with AAMI TIR12:2020<br>• In accordance with AAMI TIR 30: 2011 | Pass |
| Software Verification<br>and Validation | • In accordance with FDA-recognized voluntary consensus standard IEC<br>62304:2006+A1:2015 (13-79) | Pass |
| Cybersecurity | • In accordance with FDA Guidance for Industry and Staff -<br>Cybersecurity in Medical Devices: Quality System Considerations and<br>Content of Premarket Submissions (September 2023) | Pass |
| Usability | • In accordance with FDA-recognized voluntary consensus standard IEC<br>62366-1:2015+AMD1:2020 (5-129) | Pass |
| Performance - Bench | • In accordance with device input specifications, user needs and intended<br>uses:<br>• Video Compatibility<br>• Environmental Compatibility<br>• Voice Recognition Performance<br>• System Design Validation | Pass |
NOTE: The Connected OR Hub with Device and Voice Control and SDC4K Information Management System with Device and Voice Control is not patient contacting; therefore, biocompatibility testing is not required to support the determination of substantial equivalence.
{9}------------------------------------------------
### Conclusion:
The modifications to the subject devices do not raise new types of questions regarding safety and effectiveness compared to the predicate devices, and performance testing demonstrates that differences between the subject and predicate devices do not negatively impact the safety and effectiveness of the subject devices compared to the predicates when they are used for the proposed indications for use. The Connected OR Hub with Device and Voice Control and SDC4K Information Management System with Device and Voice Control subject devices in this 510(k) are substantially equivalent to the predicate devices K232157 (primary), K230886, K222079 for the proposed indications for use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.