FloNavi Endoscopic Fluorescence Imaging System

K221861 · Guangdong Optomedic Technologies, Inc. · GCJ · Feb 18, 2023 · Gastroenterology, Urology

Device Facts

Record IDK221861
Device NameFloNavi Endoscopic Fluorescence Imaging System
ApplicantGuangdong Optomedic Technologies, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 18, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging. Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

Device Story

FloNavi Endoscopic Fluorescence Imaging System provides real-time visible and near-infrared (NIR) fluorescence imaging during minimally invasive surgery. System components include an image processing unit, camera head, and light source. Input consists of visible light illumination and NIR excitation light; the system captures reflected visible light and ICG-induced fluorescence. The device processes these signals to display real-time video on a monitor. Operated by physicians in hospital settings, the system assists surgeons in visualizing blood flow, tissue perfusion, bile ducts, and lymphatic structures. Output enables intraoperative visual assessment, aiding surgical decision-making and potentially improving patient outcomes by identifying critical anatomical structures and perfusion status. The system is used with ICG imaging agents.

Clinical Evidence

No clinical data. Bench testing only. Performance testing included imaging and light source verification, and compliance with electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), endoscopic equipment safety (IEC 60601-2-18), laser safety (IEC 60825-1), and photobiological safety (IEC 62471).

Technological Characteristics

System includes image processing unit, camera head (4K or HD CMOS), and LED light source with NIR laser diode. Connectivity via DVI, HDMI, SDI. Reusable, non-sterile. Laser Class 3R. Complies with IEC 60601-1, 60601-1-2, 60601-2-18, 60825-1, and 62471.

Indications for Use

Indicated for use in patients undergoing minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence imaging for assessment of vessels, blood flow, tissue perfusion, and major extra-hepatic bile ducts, or visualization of the lymphatic system (vessels and nodes) following ICG administration.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 18, 2023 Guangdong OptoMedic Technologies, Inc. % Joyce Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518100 China Re: K221861 Trade/Device Name: FloNavi Endoscopic Fluorescence Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, IZI Dated: January 13, 2023 Received: January 17, 2023 Dear Joyce Yang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jianting Wang -S Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221861 ### Device Name FloNavi Endoscopic Fluorescence Imaging System ## Indications for Use (Describe) Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging. Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary - K221861 Date of Summary prepare: February 7, 2023 # 1. Submission Sponsor | Applicant Name: | Guangdong OptoMedic Technologies, Inc. | |-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | Suite 503, Building A, Golden Valley Intellicreation<br>Community, No.2 Yonganbei Street, Daxu,<br>Guicheng, Nanhai, 528200, Foshan, Guangdong,<br>P.R.China | | Contact person: | Jane Guo | | Phone: | +86 (757) 8670 2920 | | Email: | guoweijuan@optomedic.com | #### 2. Submission correspondent | Name: | Shenzhen Joyantech Consulting Co., Ltd | |-----------------|---------------------------------------------------------------------------------------------| | Address: | 1713A, 17th Floor, Block A, Zhongguan Times<br>Square, Nanshan District, Shenzhen<br>518000 | | Contact person: | Joyce Yang | | Phone: | +86-755-86069197 | | Email: | joyce@cefda.com | #### 3. Device Identification | Trade Name: | FloNavi Endoscopic Fluorescence Imaging System | |-----------------------|--------------------------------------------------------------------------------------------------------| | Common or Usual Name: | Endoscopic Imaging System | | Model: | OPTO-CAM214K, OPTO-CHD214KE, OPTO-CHD214KH, OPTO-LED214K;<br>OPTO-CAM2100, OPTO-CHD2100, OPTO-LED2100; | | Classification name: | Endoscope and accessories | | Review Panel: | Gastroenterology/Urology | | Product Code: | GCJ | | Device Class: | Class II | | Regulation Number: | 21 CFR § 876.1500 | #### Legally Marketed Predicate Device 4. | Predicate Device: | | |--------------------|----------------------------------------------------| | Trade Name: | PINPOINT Endoscopic Fluorescence Imaging<br>System | | Regulation number: | 21 CFR § 876.1500 | | Regulation class: | Class II | {5}------------------------------------------------ | Regulation name: | Endoscope and accessories | |--------------------|------------------------------------| | 510(k) Number: | K182606 | | Product Code: | GCJ; IZI | | Manufacturer: | Novadaq Technologies ULC. | | | | | Reference Device: | | | Trade Name: | Image 1 SPIES System | | Regulation number: | 21 CFR § 876.1500 | | Regulation class: | Class II | | Regulation name: | Endoscope and accessories | | 510(k) Number: | K160044 | | Product Code: | FET | | Manufacturer: | Karl Storz Endoscopy America, Inc. | | | | | Trade Name: | Power LED 175 | | Regulation number: | 21 CFR § 876.1500 | | Regulation class: | Class II | | Regulation name: | Endoscope and accessories | | 510(k) Number: | K123956 | | Product Code: | FCW, NTN | | Manufacturer: | Karl Storz Endoscopy America, Inc. | #### Device Description 5. The proposed system, FloNavi Endoscopic Fluorescence Imaging System (FloNavi System) is comprised of an image processing unit, a camera head, and a light source (including a flexible light guide cable). There are two models of the proposed system. The primary components of each model are provided in Table 1. | FloNavi Endoscopic<br>Fluorescence<br>Imaging System | Primary component | |------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | Model 1 | Image Processing Unit: OPTO-CAM214K<br>Camera Head: OPTO-CHD214KE/OPTO-CHD214KH<br>Light Source: OPTO-LED214K | | Model 2 | Image Processing Unit: OPTO-CAM2100<br>Camera Head: OPTO-CHD2100<br>Light Source: OPTO-LED2100 | | Table 1 System Model and primary component | | | | | | |--------------------------------------------|--|--|--|--|--| |--------------------------------------------|--|--|--|--|--| {6}------------------------------------------------ Except slight differences in appearance, other differences between components of either model are shown in Table 2. | Image Processing Unit | | | | |-------------------------|--------------------------------------------------------------------|-----------------------------------------------|-------------------------| | Item | OPTO-CAM214K | OPTO-CAM2100 | | | Video outputs | DVI, HDMI, 4×3G-SDI, 12G-SDI | SDI, DVI, CVBS, S-VIDEO | | | Video output resolution | $4096 \times 2160p, 3840 \times 2160p, 1920 \times 1080p, 50/60Hz$ | $1920 \times 1080p, 720 \times 576i, 50/60Hz$ | | | Size (mm) | 403*370*150 | 403*370*150 | | | Weight (kg) | 9±15% | 9±15% | | | Camera Head | | | | | Item | OPTO-CHD214KH | OPTO-CHD214KE | OPTO-CAM2100 | | Image sensor | 4K CMOS sensor assembly | 4K CMOS sensor assembly | HD CMOS sensor assembly | | Focusing mechanism | Manual | Motor-driven | Manual | | Weight (kg) | | 0.6±15% | 0.5±15% | | Light Source | | | | | Item | OPTO-LED214K | OPTO-LED2100 | | | Size (mm) | 403*370*150 | 403*370*150 | | | Weight (kg) | 11±15% | 11±15% | | | Light guide cable | Diameter: 4.5mm | Diameter: 4.8 mm | | FloNavi Endoscopic Fluorescence Imaging System is capable of providing real-time endoscopic visible and near-infrared fluorescence imaging. During surgical procedures, FloNavi may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflecting light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein. {7}------------------------------------------------ To provide NIR fluorescence imaging, FloNavi is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the light source, and the fluorescence response is then imaged with the camera, processed and displayed on a video monitor. The proposed system is designed to be used with rigid endoscopes, monitors and other ancillary equipment. The compatible rigid endoscope is O-Mec laparoscopes 690 Series (Model: 690-331030H), which was cleared under K201151. ## Intended Use/ Indications for Use Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide realtime endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging. Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes. | Comparison item | Subject Device (K221861) | Predicate Device(K182606) | | |-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------| | Product Name | FloNavi<br>Fluorescence<br>Imaging System | PINPOINT<br>Fluorescence Imaging System | | | Product Code | GCJ | GCJ, IZI | | | Regulation Number | 21 CFR § 876.1500 | 21 CFR § 876.1500 | | | Classification | Class II | Class II | | | Type of use | Prescription Use | Prescription Use | | | Comparison item | Subject Device (K221861) | Predicate Device(K182606) | | | Intended use &<br>Indication s for Use | Upon intravenous<br>administration and use of<br>an ICG consistent with its<br>approved label, the FloNavi<br>Endoscopic Fluorescence<br>Imaging System is intended<br>to provide real-time<br>endoscopic visible and<br>near-infrared fluorescence<br>imaging. During minimally<br>invasive surgery, the<br>FloNavi Endoscopic<br>Fluorescence Imaging<br>System enables surgeons<br>to perform minimally<br>invasive surgery using<br>standard endoscopic visible<br>light as well as visual<br>assessment of vessels,<br>blood flow and related<br>tissue perfusion, and at<br>least one of the major extra-<br>hepatic bile ducts, using<br>near-infrared imaging.<br>Upon interstitial<br>administration and use of<br>an ICG consistent with its<br>approved label, the FloNavi<br>Endoscopic Fluorescence<br>Imaging System is used to<br>perform intraoperative<br>fluorescence imaging and<br>visualization of the<br>lymphatic system, including<br>lymphatic vessels and<br>lymph nodes. | Upon intravenous administration of<br>TRADENAME (ICG drug product),<br>the PINPOINT Endoscopic<br>Fluorescence Imaging System is<br>used with TRADENAME to perform<br>intraoperative fluorescence<br>angiography, and it is also indicated<br>for use in fluorescence imaging of<br>biliary ducts, and when indicated,<br>during intraoperative<br>cholangiography.<br>The PINPOINT Endoscopic<br>Fluorescence Imaging System is<br>indicated for use to provide real time<br>endoscopic visible and near-infrared<br>fluorescence imaging. The<br>PINPOINT System enables<br>surgeons to perform minimally<br>invasive surgery using standard<br>endoscope visible light as well as<br>visual assessment of vessels, blood<br>flow and related tissue perfusion,<br>and at least one of the major extra-<br>hepatic bile ducts (cystic duct,<br>common bile duct or common<br>hepatic duct), using near infrared<br>imaging.<br>Fluorescence imaging of biliary<br>ducts with the PINPOINT System is<br>intended for use with standard<br>of care white light, and when<br>indicated, intraoperative<br>cholangiography. The device is not<br>intended for standalone use for<br>biliary duct visualization.<br>Upon interstitial administration of<br>TRADENAME (ICG drug product),<br>the PINPOINT System is<br>used to perform intraoperative<br>fluorescence imaging and<br>visualization of the lymphatic<br>system,<br>including lymphatic vessels and<br>lymph nodes. | | | Applicable user | Physicians | Physicians | | | Environment of use | Healthcare facility/hospital | Healthcare facility /hospital | | | Single use / Reusable | Reusable | Reusable | | | Sterile /non-sterile | Marketed as non-sterile | Marketed as non-sterile | | | Device System<br>components | - Image Processing Unit<br>- Camera Head<br>- Light Source | - Endoscopic video processor /<br>illuminator (VPI)<br>- Laparoscope | | | Comparison item | Subject Device (K221861) | Predicate Device(K182606) | | | | - Light Guide Cable | - Camera head | | | | | - Light guide cable | | | Video output signals | OPTO-CAM214K:<br>DVI, HDMI, 4×3G-SDI,<br>12G-SDI | HD-SDI, DVI | | | Video output signals | OPTO-CAM2100:<br>SDI, DVI, CVBS, S-VIDEO | | | | Video output resolution | OPTO-CAM214K:<br>4096×2160p; 3840×2160p;<br>1920×1080p | 1920×1080 | | | Video output resolution | OPTO-CAM2100:<br>1920×1080p; 720×576i | | | | Voltage | 110-240V~ | 100-240V~ | | | Frequency | 50/60 Hz | 50/60 Hz | | | Power consumption | Image Processing Unit:<br>150VA<br>Light Source: 200VA | 300VA | | | Image sensors | OPTO-CHD214KE/OPTO-<br>CHD214KH :<br>4K CMOS sensor assembly | CMOS HD sensor assembly…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...