SOCORRO SMI TUNNEL RELEASE DEVICE

K102219 · Socorro Medical, Inc. · GCJ · Sep 24, 2010 · Gastroenterology, Urology

Device Facts

Record IDK102219
Device NameSOCORRO SMI TUNNEL RELEASE DEVICE
ApplicantSocorro Medical, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateSep 24, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Socorro SMI Tunnel Release Device is intended for the minimally invasive examination, isolation and recession/incision of ligament, tendons, or fascia as in: -Plantar Fasciotomy, -Carpal Tunnel Release, -Elbow Cubital Tunnel Release and -Gastrocnemius Tenotomy.

Device Story

The Socorro SMI Tunnel Release Device is a single-use, sterile surgical instrument designed for minimally invasive procedures. It consists of a handpiece with a tissue-cutting probe, used alongside an endoscope camera and fiber optic illumination system. The handpiece allows the surgeon to extend, retract, and angularly rotate surgical blades at the distal end. During surgery, the endoscope is inserted through small incisions to provide nearly 360-degree visualization of the operative site on an overhead monitor. The surgeon uses the device to directly visualize and excise targeted tissue, such as thickened tendon sheaths or soft-tissue masses causing entrapment neuropathy. The device is intended for use by physicians in a clinical or surgical setting to relieve nerve or tendon compression in the wrist, elbow, knee, ankle, or heel. By enabling direct visualization during the cutting process, the device assists the surgeon in verifying complete tissue excision, potentially improving procedural accuracy and patient outcomes.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Handpiece with tissue-cutting probe; stainless steel patient-contacting components; mechanical blade actuation (extend, retract, rotate); endoscopic visualization via fiber optic illumination; single-use, sterile; manual operation.

Indications for Use

Indicated for patients requiring minimally invasive examination, isolation, and recession/incision of ligaments, tendons, or fascia for conditions including plantar fasciotomy, carpal tunnel release, elbow cubital tunnel release, and gastrocnemius tenotomy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Section 5: 510(k) Summary # K102219 SEP 2 4 2010 | Device Information: | | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Category | Comments | | Sponsor: | Socorro Medical, Inc.<br>450 Sutter Street, Suite 500<br>San Francisco, CA 94108<br>Contact: Ed Diao, M.D.<br>ediao@sf-sc.com<br>415-362-8880 | | Correspondent Contact<br>Information: | Craig Coombs<br>Coombs Medical Device Consulting<br>1193 Sherman Street<br>Alameda, CA 94501<br>Tel: 510-337-0140<br>Fax: 510-337-0416 | | Device Common Name: | Surgical Knife | | Device Classification & Code: | Class II (due to specific claims, otherwise Class I),<br>EMF | | Device Classification Name: | 21 CFR 878.4800 | Socorro SMI Tunnel Release Device #### Predicate Device Information: Device Proprietary Name: | Predicate Device: | A.M. Surgical Mountable Endoscopic Blade | |-----------------------------------------|------------------------------------------| | Predicate Device Manufacturer: | A.M. Surgical, Inc. | | Predicate Device Common Name: | Surgical Knife | | Predicate Device Classification: | 21 CFR 878.4800 | | Predicate Device Classification & Code: | Class II, EMF | | Premarket Notification Number: | K080133 | ## b. Date Summary Prepared 18 June 2010 # c. Description of Device The SMI Tunnel Release Device is comprised of a hand piece with a tissue cutting probe which is used in conjunction with an endoscope camera and fiber optic illumination system. The device is supplied sterile and is for single use only. The hand piece provides an actuation member for extending, retracting, and angularly rotating surgical blades which are attached to the distal end of the hand piece. Basic features of these probes are a retraction region and visualization that is nearly threehundred and sixty degrees (360°). The visualization scope connects with the camera head and optic cables which provide the surgeon a clear and complete view of the surgical site on the overhead monitor. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with the word "SOCORRO" written in a sans-serif font. Above the word is a circular graphic that appears to be a stylized representation of a celestial body, possibly a moon or planet, with craters or surface details visible. The text below "SOCORRO" is illegible due to the image's resolution. The endoscope is used to illuminate and visualize the anatomy in or around the operative site. It allows the physician to directly visualize the actions of the probe during the procedure and to verify complete excision of the targeted tissue. # d. Intended Use The Socorro SMI Tunnel Release Device is intended for the minimally invasive examination, isolation and recession/incision of ligament, tendons, or fascia as in: -Plantar Fasciotomy, -Carpal Tunnel Release, -Elbow Cubital Tunnel Release and -Gastrocnemius Tenotomy. # e. Comparison to Predicate Device The Socorro SMI Tunnel Release Device is substantially equivalent in intended use, technology, design and materials to the A.M. Surgical Mountable Endoscopic Blade (K080133). The Socorro SMI Tunnel Release Device and its predicate all provide technology to access and relieve an entrapment neuropathy that occurs when a nerve or tendon at the wrist, elbow, knee, ankle or heel is compressed by a thickened tendon sheath, edema or soft-tissue mass. An endoscope is guided into the joint through a small incision(s) above and/or below the affected joint. Small cutting tools (mechanical blades) are then used to excise the tendon sheath or mass under direct observation. The patient contacting components that make up the application device and its predicate is stainless steel. The blades are provided sterile and are for single use only. Other components of the predicate device are sterilized between uses while the entire application device is provided sterile and is for single use only. The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy. Socorro Medical concludes that the devices are substantially equivalent. ## f. Summary of Supporting Data Bench testing has demonstrated that the device is in compliance with international standards, the expectations of the medical community and the product labeling. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle faces to the left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Socorro Medical, Inc. % Coombs Medical Device Consulting Mr. Craig Coombs 1193 Sherman Street Alameda, California 94501 SEP 2 4 2010 Re: K102219 Trade/Device Name: Socorro SMI Tunnel Release Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ, EMF Dated: September 20, 2010 Received: September 22, 2010 #### Dear Mr. Coombs: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Mr. Craig Coombs or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Section 4: Indications for Use Statement 510(k) Number (if known): K102277 SEP 24 2010 Device Name: Socorro SMI Tunnel Release Device Indications For Use: The Socorro SMI Tunnel Release Device is intended for the minimally invasive examination, isolation and recession/incision of ligament, tendons, or fascia as in: -Plantar Fasciotomy, -Carpal Tunnel Release, -Elbow Cubital Tunnel Release and - -Gastrocnemius Tenotomy. Prescription Use AND/OR X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Neil R. Doyle for nxn (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K\|02219
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...