NOMAD ND1000M AUGMENTED VISION SYSTEM

K030940 · Microvision Incorporated · GCJ · Apr 9, 2003 · Gastroenterology, Urology

Device Facts

Record IDK030940
Device NameNOMAD ND1000M AUGMENTED VISION SYSTEM
ApplicantMicrovision Incorporated
Product CodeGCJ · Gastroenterology, Urology
Decision DateApr 9, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The intended use of the Nomad™ ND1000M system is to display video data or images while worn on the user's head.

Device Story

Nomad™ ND1000M is a monochromatic, head-worn, monocular augmented vision system. It accepts SVGA video input from external sources, processing and projecting images directly into the user's field of vision. The system consists of a head-mounted display module, a video control electronics module, and a power source (Li-Ion battery or AC/DC supply). It functions as a see-through display, superimposing images over the ambient environment, or as an occluded display using an optional ocular cover. Used by clinicians to view video data or images during procedures. The device provides visual feedback to the operator, potentially enhancing situational awareness or data accessibility during clinical tasks.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Monochromatic head-worn monocular display. Accepts SVGA video input. Powered by rechargeable Li-Ion battery or medical-grade AC/DC power supply. Includes headgear and video control electronics module. See-through or occluded (with cover) operation. No specific material standards or software algorithm classes are detailed.

Indications for Use

Indicated for displaying video data or images to a user via a head-worn monocular display. No specific patient population, age, or disease state contraindications are defined.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR 0 9 2003 Microvision, Inc. P.O. Box 3008 19910 North Creek Parkway Bothell, Washington 98011 K030940 Nomad™ ND1000M Augmented Vision System 510(k) Submission Image /page/0/Picture/3 description: The image shows a trefoil knot, which is the simplest non-trivial knot. It is a closed curve that is embedded in three-dimensional space and cannot be untangled to form a simple loop without cutting the knot. The trefoil knot has three crossings, which are the points where the curve crosses over itself. MICROVISION 6) 510(k) Summary Date: # Company Name and Address Microvision Incorporated 19910 North Creek Parkway Bothell, WA 98011 ## Contact Person Karl Bylund Regulatory and Compliance Engineer Telephone: 425-415-6634 ### Device Trade Name Nomad™ ND1000M Augmented Vision System Common Name Video Monitor ### Classification Name Endoscope and Accessories ## Predicate Devices | 1) Device name: | Head Mounted Display | |---------------------|------------------------------------------------------------------------------------------------| | Manufacturer: | Vista Medical Technologies<br>5451 Avenida Encinas, Suite A<br>Carlsbad, CA 92008 | | Classification : | Class II | | 510K Number : | K961800 | | Regulation Number : | 876.1500 | | Product code : | GCJ | | 2) Device name: | i-View Personal Video Display | | Manufacturer: | MediVision Endoscopy Incorporated<br>1440 S. State College Boulevard, #1D<br>Anaheim, CA 92806 | | Classification : | Class II | | 510K Number : | K000669 | | Regulation Number : | 876.1500 | | Product code : | GCJ | 425-415-6847 phone: facsimile: 425-415-6600 website: www.mvis.com {1}------------------------------------------------ ### Nomad™ ND1000M Augmented Vision System 510(k) Submission #### Description of the Device The Nomad™ ND1000M Augmented Vision System is a monochromatic head-worn monocular display. The intended use of the Nomad™ ND1000M system is to display video data or images while worn on the user's head. It can be connected to any SVGA video source. The display combines the ambient environment with the displayed image (the image is superimposed over the normal field of vision). The display may be adjusted over either eve. Optionally, the display may be used in a non see-through fashion (occluded) with addition of the provided ocular cover. The system consists of a display module, attached to headgear, that is connected via a .non-detachable interconnecting cable to a video control electronics module. The video input is connected to the video control electronics module. The device accepts power from a battery or an optional medical grade AC to DC power supply, both supplied with the device. A beit to hold the video control module and battery is provided. A rechargeable Li-Ion battery is also supplied along with a battery charger. #### Intended Use The Nomad™ ND1000M Augmented Vision System is designed to display video data or images while worn on the user's head. ## Technological Characteristics Comparison with Predicate Devices The Nomad™ ND1000M system is similar to the Vista HMD (Head Mounted Display) which received FDA clearance on September 11, 1996 (K961800), and the MediVision i-View Personal Video Display which received FDA clearance on May 26, 2000 (K000669). An exception is that the Nomad™ ND1000M system is monochromatic while the other systems are full color. All of these devices accept video signals and display video data or images while worn on the user's head. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 0 9 2003 Microvision Incorporated c/o Mr. Charles Mack Engineering Team Leader Underwriters Laboratories, Inc. 2600 NW Lake Road Camas, Washington 98607-8642 Re: K030940 Trade/Device Name: Nomad™ ND1000M Augmented Vision System Regulation Number: 21 CFR 876.1500 Regulation Name: Laparoscope and accessories Regulatory Class: II Product Code: GCJ Dated: March 21, 2003 Received: March 25, 2003 Dear Mr. Mack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Charles Mack This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Attachment 1 b Page __ of _ 510(k) Number (if known): Device Name: Indications For Use: K030940 Video Monitor The intended use of the Nomad™ ND1000M system is to display video data or images while worn on the user's head. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Meriam C. Provost (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number KC30940 (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...