GUIDANT VASOVIEW ENDOSCOPIC VESSEL HARVESTING SYSTEM
Applicant
Guidant Corp.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
May 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Guidant VasoView® Endoscopic Vessel harvesting system is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for patients requiring blunt dissection of tissue including dissection of blood vessels, separation of the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection along the saphenous vein or radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.
Device Story
System comprises blunt tip trocar, 5mm extended length endoscope, dissection cannulas, and bipolar instruments (BiSector/scissors). Used in minimally invasive surgery for vessel harvesting (saphenous vein/radial artery) and thoracic dissection. Surgeon inserts trocar to create access port; endoscope provides visualization of subcutaneous/thoracic cavities. Cannulas tunnel and separate tissue planes; bipolar instruments perform coagulation and mechanical transection of vessel branches. Output is visual feedback for surgeon; clinical decision-making relies on endoscopic view to guide dissection and ligation. Benefits include reduced surgical trauma compared to open harvesting.
Clinical Evidence
Clinical data from 84 cases presented to support the expanded indication of endoscopic radial artery harvesting. No new pre-clinical testing performed as the device design and materials are identical to previously cleared versions. Evidence demonstrates safety and effectiveness for the expanded clinical application.
Technological Characteristics
System includes stainless steel endoscope, single-use sterile cannulas, and bipolar electrosurgical instruments. Operates via endoscopic visualization and bipolar coagulation energy from compatible generators. Mechanical actuation used for vessel transection. Trocar includes balloon for insufflation/gas seal. No software or electronic processing components described.
Indications for Use
Indicated for patients undergoing minimally invasive surgery for vessel harvesting, specifically for arterial bypass (coronary artery bypass grafting). Applicable for blunt tissue dissection, including blood vessels, extraperitoneal/subcutaneous extremity spaces, and thoracic spaces (e.g., chest wall structures external to the parietal pleura).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
# 510(k) Premarket Notification Supplement Guidant VasoView Endoscopic Vessel Harvesting System
510(k) SUMMARY
K030512
MAY 1 4 2003
This summary of 510(k) safety and effectiveness information is submitted in accordance3 with the requirements of SMDA 1990 and 21 CFR 8078.92.
#### A. Name, Address, Phone and Fax number of the Applicant
Guidant Corporation Cardiac Surgery 3200 Lakeside Drive Santa Clara, CA 95054
Telephone: (408) 845-1910 (408) 845-1800 Fax:
#### B. Contact Person
Nancy Lincé Regulatory Affairs Consultant
#### C. Date Prepared
February 14, 2003 (revised May 1, 2003)
#### D. Device Name
Trade Name: Guidant VasoView System Classification Name: Surgical Trocar, Cannula, Electrosurgical Devices and Endoscopes
#### E. Device Description
The VasoView System is made up of 6 components including:
- 1) Blunt Tip Trocar (BTT) with Syringe,
- 2) 5 mm Extended Length Endoscope,
- 3) Uniport Plus Dissection Cannula,
- 4) VasoView Dissection Cannula,
- 5) Uniport BiSector and
- 6) Uniport Bipolar Scissors.
These are the exact same devices used for saphenous vein harvesting and have been previously cleared by FDA. Each device is described below.
# BTT:
The BTT is a single use device provided sterile. The BTT is used to provide a port of access for insertion of endoscopic instruments into an incision site. The device consists of a valve body assembly with a balloon on the sleeve. The valve body assembly contains an internal flapper valve and seal to prevent gas leakage when instruments are
{2}------------------------------------------------
# Guidant Corporation Cardiac Surgery
# 510(k) Premarket Notification Supplement Guidant VasoView Endoscopic Vessel Harvesting System
inserted or withdrawn. It also includes built-in converters to allow insertion of instruments of different diameters than the main seal and an external one-way valve for gas insufflation. To minimize leakage and secure the port, the distal end of the sleeve has a balloon. A 30cc syringe is provided for inflation/deflation of the balloon.
### 5mm Extended Length Endoscope:
The 5mm Extended Length Endoscope is a multi-use product supplied non-sterile and must be cleaned and sterilized prior to use. The device is used to visualize endoscopic cavities. It consists of a stainless steel shaft containing optics. The proximal end has an eyepiece, for camera attachment, and a light cable post for light cable connection.
# Uniport Plus Dissection Cannula
The Uniport Plus Dissection Cannula is a single use device provided sterile. The Uniport Plus Dissection Cannula is designed to be used in conjunction with the 5mm Extended Length Endoscope. The cannula has a removable conical tip and four lumens to house the endoscopes, vessel cradle/distal lens washer and bipolar instrument for ligation and division of vessel branches. The vessel cradle/distal lens washer is independently controlled by a slider button on the handle of the device for retraction and blunt dissection of tissue as well as washing of the distal tip of the scope. The bipolar instrument is inserted independently into the cannula and can be controlled at the proximal end for blunt dissection.
### Dissection System:
The Dissection System is a single-use device provided sterile. It is designed to be used in conjunction with a 5mm Extended Length Endoscope. The cannula has a tapered tip for tunneling, which separates tissue planes, forming a cavity. The cannula tip may also be used for dissection and isolation of structures in the cavity. This tip is clear for endoscopic visualization during tunneling. An elliptical bulb aid tissue separation. A handle is provided to assist with advancement of the cannula.
### Uniport BiSector:
The Uniport BiSector is a single use device provided sterile. The Uniport BiSector is a 5mm flexible endoscopic ligating forcep and is intended to be used with the Uniport Dissection Cannula. Bipolar coagulation is achieved using electro surgical energy under endoscopic visualization. Transection is achieved through mechanical actuation of the slide button. This device is intended to be used with the bipolar outputs of compatible generators.
### Bipolar Scissors:
The Bipolar Scissors are a single use device provided sterile. The Bipolar Scissors are 5mm cutting and bipolar coagulation scissors and are intended to be used with the Dissection Cannulas. Mechanical cutting and bipolar coagulation is achieved using electrosurgical energy under endoscopic visualization. This device is intended to be used with the bipolar outputs of compatible generators.
{3}------------------------------------------------
### 510(k) Premarket Notification Supplement Guidant Corporation Cardiac Surgery
### F. Intended Use
The Guidant VasoView® Endoscopic Vessel harvesting system is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for patients requiring blunt dissection of tissue including dissection of blood vessels, separation of the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection along the saphenous vein or radial artery for use in coronary artery bypass grafting. Thoracoscopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.
### Substantial Equivalence G.
The Guidant VasoView Endoscopic Vessel Harvesting System submitted under this 510K application is substantially equivalent to the predicated devices in intended use, indications, technological characteristics, labeling and packaging materials, and manufacturing processes. The technological characteristics, packaging materials and manufacturing processes are unchanged from 510K numbers K904993, K945975, K960637, K974608, K981700 and K992353, their amendments and supplements. All devices are cleared through FDA or exempt. The present submission is limited to an indication expansion. The clinical data presented in the submission demonstrates that the Guidant VasoView Endoscopic Vessel Harvesting System does no raise new questions of safety and effectiveness when used to perform the expanded indication of radial artery harvesting.
### H. Device Testing Results and Conclusion
The pre-clinical testing information for the Vaso View System devices is unchanged from 510K numbers K904993, K945975, K960637, K974608, K981700 and K992353, their amendments and supplements. All devices are cleared through FDA or exempt. The present submission is limited to an indication expansion, which does not affect the preclinical testing.
The results of over 84 cases demonstrate that the Guidant VasoView Vessel Harvesting System is a safe and effective approach to endoscopic radial artery harvesting. Furthermore, the safety and efficacy of performing radial artery harvesting using the endoscopic approach has been well demonstrated and presented.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of four stylized human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 1 4 2003
Ms. Nancy Lincé Regulatory Affairs Consultant Guidant Corporation Cardiac Surgery 3200 Lakeside Drive Santa Clara, California 95054
Re: K030512
Trade/Device Name: Guidant VasoView Endoscopic Harvesting System Regulation Number: 21 CFR 876.1500 Regulation Names: Endoscope and accessories Regulatory Class: II Product Codes: GCJ Dated: February 17, 2003 Received: February 19, 2003
Dear Ms. Lincé:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Nancy Lincé
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### 510(k) Premarket Notification Supplement Guidant Corporation Guidant VasoView Endoscopic Vessel Harvesting System Cardiac Surgery
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K030512
Guidant VasoView Endoscopic Harvesting System Device Name:
Indications For Use:
The Guidant Vaso View® Endoscopic Vessel harvesting system is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for patients requiring blunt dissection of tissue including dissection of blood vessels, separation of the extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissection along the saphenous vein or radial artery for use in coronary artery bypass grafting. Thorascopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR Over-The-Counter Use (Per 21 CFR 801.109)
(Optional Format 1-2-96)
Muriam C. Provost
Division of General, Restorative and Neurological Devices
510(k) Number k030512
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.