APPLIED GELPORT LAPAROSCOPIC HAND ACCESS DEVICE

K014047 · Applied Medical Resources Corp. · GCJ · Feb 28, 2002 · Gastroenterology, Urology

Device Facts

Record IDK014047
Device NameAPPLIED GELPORT LAPAROSCOPIC HAND ACCESS DEVICE
ApplicantApplied Medical Resources Corp.
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 28, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Applied GelPort Laparoscopic Hand Access Device is indicated for use in patients undergoing laparoscopic surgical procedures to obtain abdominal access by the surgeon's hand and instrumentation. Additional slits may be made using a trocar at the periphery of the Gel Seal Cap to allow instrument access to the surgical site. The Applied GelPort Laparoscopic Hand Access Device may be used in procedures such as nephrectomy, colectomy and splenectomy in colorectal, urological and general surgery to access the surgical site.

Device Story

Disposable, single-use laparoscopic hand access device; facilitates hand-assisted laparoscopic surgery (HALS). Components: Gel Seal Cap, base ring assembly, retracting sheath. Surgeon inserts hand through central slit in Gel Seal Cap; gel material conforms to hand/instrument to maintain pneumoperitoneum. Additional instrument/trocar access provided via central slit or peripheral slits created by trocar. Used in OR by surgeons; restores tactile feedback and hand-eye coordination of open surgery while maintaining minimally invasive benefits. Reduces need for additional incisions. Packaged in PETG tray with Mylar cover.

Clinical Evidence

No clinical data provided; substantial equivalence based on design and technological characteristics.

Technological Characteristics

Disposable, single-use device. Components: Gel Seal Cap (gel material), base ring assembly, retracting sheath. Packaging: PETG tray with Mylar peel cover. Mechanical seal principle via conformable gel material. No electronic components or software.

Indications for Use

Indicated for patients undergoing laparoscopic surgical procedures (e.g., nephrectomy, colectomy, splenectomy) in colorectal, urological, and general surgery requiring abdominal access for surgeon's hand and instrumentation.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## K014047 510(k) SUMMARY ## FEB 2 8 2002 PENDING 510(k) NUMBER: Applied Medical Resources Corporation SUBMITTED BY: 22872 Avenida Empresa Rancho Santa Margarita, CA-92688 (949) 713-8000 CONTACT PERSON: Anil Bhalani Vice President of Regulatory Affairs and Clinical Programs December 6, 2001 DATE OF PREPARATION: Laparoscopic Hand Access Port NAME OF DEVICE: Laparoscope, General & Plastic Surgery. CLASSIFICATION NAME: (Regulation Number 21CFR 876.1500, Endoscope and accessories). Applied GelPort Laparoscopic Hand Access Device TRADE NAME: 1. Smith & Nephew Handport System (K990414), Smith & PREDICATE DEVICE: Nephew, Inc., Andover, MA. 2. Applied Intromit Laparoscopic Hand Access Port (K990663), Medtech Ltd., Ireland. Marketed by Applied Medical Resources, Rancho Santa Margarita, CA. SUMMARY STATEMENT: The Applied GelPort Laparoscopic Hand Access Device is indicated for use in patients undergoing laparoscopic surgical procedures to obtain abdominal access by the surgeon's hand and instrumentation. Additional slits may be made using a trocar at the periphery of the Gel Seal Cap to allow instrument access to the surgical site. The Applied GelPort Laparoscopic Hand Access Device may be used in procedures such as nephrectomy, colectomy and splenectomy in colorectal, urological and general surgery to access the surgical site.. With the use of the Applied GelPort Laparoscopic Hand Access Device, the surgeon regains tactile sense and feedback along with the increased hand-eye and instrumentation manipulation capacity of open surgery. The device allows the surgeon multiple passages of the hand to access the surgical site without losing pneumoperitoneum. Instruments and Trocars may also be placed through the port in the center of the device and at the periphery of the device to allow instrument access to the surgical site, thereby reducing the need for additional incisions to the patient. {2}------------------------------------------------ The device is designed with a slit for the hand at the center of the Gel Seal Cap. In addition to providing hand access, an instrument or trocar can also be inserted through this port during surgery to access the surgical site. The gel material conforms to the shape of the hand, instrument or trocar inserted through it providing a seal during laparoscopic surgery. Trocars can also be placed at the periphery of the Gel Seal Cap to allow for additional ports for instrument access during Hand Assisted Laparoscopy reducing the number of incisions required during surgery. The Applied GelPort Laparoscopic Hand Access Device is very simple in design. It is simple to set up, easy to use and has a lower profile to maximize comfort. It consists of a Gel Seal Cap, a Base Ring Assembly, and a Retracting Sheath. The simple design makes installation of the Applied GelPort Laparoscopic Hand Access Device during clinical use very easy. Using a sterile skin marker an incision line is marked at the surgery site. Once the incision is made the retracting sheath is placed in position. The base ring assembly is then attached to the sheath. To complete the set up the Gel Seal Cap is snapped onto the base ring assembly. The sterile lubricant is provided to lube the surgeon's glove and the top of the Gel Seal Cap to make insertion of the hand into the port easy. Instruments may also be placed through the center slit if desired during surgery. If additional ports of entry are required for instrument access to the surgical site, slits can be made at desired locations on the periphery of the Gel Seal Cap using a non-threaded trocar. Instruments can then be inserted through the trocar cannula placed at the peripheral slit. The Applied GelPort Laparoscopic Hand Access Device is a disposable, single-use device, packaged inside a PETG Tray with a Mylar peel cover, which is standard packaging material for medical products. The Applied GelPort Laparoscopic Hand Access Device is substantially equivalent to predicate devices in design methodology, principle of operation and clinical utility. The Applied GelPort Laparoscopic Hand Access Device introduces no new safety and effectiveness issues when used as instructed. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird symbol with three curved lines representing the bird's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. The text is in all capital letters and is evenly spaced around the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 8 2002 Mr. Anil Bhalani Vice President of Regulatory Affairs and Clinical Programs Applied Medical Resources Corporation 22872 Avenida Empresa Rancho Santa Margarita, CA 92688 Re: K014047 Trade/Device Name: Applied GelPort Laparoscopic Hand Access Device Regulation Number: 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: December 6, 2001 Received: December 7, 2001 Dear Mr. Bhalani: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Mr. Anil Bhalani This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. to Mark N. Millman Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE Applied Medical Resources is providing this separate cover page for the Applied GelPort Laparoscopic Hand Access Device's "Indications for Use" as required. 510(k) Number: K 014047 Device Name: Applied GelPort Laparoscopic Hand Access Device Indications for Use: The Applied GelPort Laparoscopic Hand Access Device is indicated for use in patients undergoing laparoscopic surgical procedures to obtain abdominal access by the surgeon's hand and instrumentation. Additional slits may be made using a trocar at the periphery of the Gel Seal Cap to allow instrument access to the surgical site. The Applied GelPort Laparoscopic Hand Access Device may be used in procedures such as nephrectomy, colectomy and splenectomy in colorectal, urological and general surgery to access the surgical site. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) ✓ OR Over-The -Counter Use **__** for (Optional Format 1-2-96) (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number **__**K014047 ર 8
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...