SLS-CT
Device Facts
| Record ID | K960985 |
|---|---|
| Device Name | SLS-CT |
| Applicant | Bio-Imaging Research, Inc. |
| Product Code | JAK · Radiology |
| Decision Date | Oct 25, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1750 |
| Device Class | Class 2 |
Intended Use
The SLS-CT System manufactured by Bio-Imaging Research (BIR) is a passive add-on option to Philips Medical Systems, Inc.'s existing, legally marketed, Radiotherapy Simulators. The SLS-CT does not control the Simulator x-ray and mechanical functions. The SLS-CT is a data gathering device that allows the Radiographic Simulators to process data into tomographic images. The SLS-CT is not a diagnostic tool.
Device Story
SLS-CT System; passive add-on for Philips Radiotherapy Simulators. Device gathers data from simulator; processes data into tomographic images. Does not control simulator x-ray or mechanical functions. Used in clinical radiotherapy settings by trained personnel. Output used for radiotherapy planning/simulation; not for diagnostic purposes. System verification/validation performed via Beta testing in clinical and manufacturer settings.
Clinical Evidence
Bench testing and two Beta tests performed; Beta Test 1 at Philips Medical Systems, Beta Test 2 in a clinical setting. In-house verification and validation conducted by BIR.
Technological Characteristics
Passive add-on hardware/software module. Complies with IEC601, UL 2601, CSA 601. Software developed per IEEE Software Engineering Standard. ISO9001 certified manufacturing.
Indications for Use
Indicated for use as a data gathering add-on for existing Philips Radiotherapy Simulators to enable tomographic image processing. Not for diagnostic use.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Related Devices
- K953233 — SIMVIEW 3000 CT OPTION · Siemens Medical Solutions USA, Inc. · Nov 29, 1996
- K981056 — VARIAN XIMATRON SCAN VISION CT · Varian Medical Systems, Inc. · Jun 12, 1998
- K063730 — DEXTROSCOPE, MODEL MK10; DEXTROBEAM, MODEL MK3, MK4; RADIODEXTER, MODEL 1.0 · Volume Interactions Pte, Ltd. · Jan 31, 2007
- K980837 — SHIMADZU WHOLE BODY X-RAY COMPUTED TOMOGRAPHY SCANNER SCT-7000TS · Shimadzu Medical Systems · May 20, 1998
- K984156 — SHIMADZU WHOLE BODY X-RAY COMPUTED TOMOGRAPHY SCANNER SCT-7000TX W/FLURO OPTION · Shimadzu Medical Systems · Apr 21, 1999