ADVANCE VESSEL ANALYSIS II

K060779 · GE Healthcare · JAK · Apr 5, 2006 · Radiology

Device Facts

Record IDK060779
Device NameADVANCE VESSEL ANALYSIS II
ApplicantGE Healthcare
Product CodeJAK · Radiology
Decision DateApr 5, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Intended Use

Advanced Vessel Analysis II is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. Advanced Vessel Analysis II (AVA II) is a post processing application option for the Advantage Workstation (AW) platform, CT Scanner or PACS stations, which can be used in the analysis of 2D and 3D CT Angiography images/data derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular and vascular disease assessment. This software is designed to support the physician in assessment of stenosis analysis, pre/post stent planning and directional vessel tortuosity visualization. AVA II automatic visualization tools provide the users with the capabilities to facilitate segmentation of bony structures for accurate identification of the vessels. Once vessels are visualized, tools are available for sizing the vessel, analyzing calcified and non-calcified plaque to determine the densities of plaque within a coronary artery, measure area's of abnormalities within a vessel.

Device Story

AVA II is a post-processing software package for Advantage Workstation (AW), CT scanners, and PACS. It processes DICOM 3.0 compliant 2D/3D CT angiographic images. Key functions include automated segmentation of bony structures to isolate vessels; vessel sizing; plaque analysis (calcified/non-calcified) to determine coronary artery plaque density; and measurement of vascular abnormalities. Used by radiologists, cardiologists, and clinicians to visualize vascular anatomy, assess stenosis, and plan pre/post-stent procedures. Output consists of processed images, measurements, and visualizations, aiding clinical decision-making regarding treatment paths. Benefits include non-invasive assessment of vascular pathology.

Clinical Evidence

No clinical data provided. Safety and effectiveness are supported by software development, validation, and verification processes, and adherence to industry and international standards.

Technological Characteristics

Software post-processing package for Advantage Workstation (AW), CT scanners, and PACS. Operates on DICOM 3.0 compliant CT data. Features include automated bone segmentation, vessel sizing, and plaque density quantification. Software-based analysis.

Indications for Use

Indicated for clinicians to analyze vascular anatomy and pathology, including stenosis, plaque density, and vessel tortuosity, using 2D/3D CT angiographic images for cardiovascular and vascular disease assessment and stent planning.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K06079 Image /page/0/Picture/1 description: The image contains a lowercase letter 'g' with a horizontal line running through the middle of the letter. The 'g' is in a bold font and is black. The horizontal line is also black and appears to be slightly thicker than the lines that make up the letter 'g'. 5 2006 APR # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h). - Larry A. Kroger, Ph.D. Submitter Senior Regulatory Programs Manager Tel: (262) 544-3894 Fax: (262) 544-4768 GE Healthcare W-400 3000 North Grandview Blvd. Waukesha, WI 53188 USA Date Prepared: March 15, 2006. ## PRODUCT IDENTIFICATION | Name: | Advanced Vessel Analysis II | |-----------------------------------|---------------------------------------------------------------------------------------| | Classification Name: | Accessory to Computed Tomography System per 21 CFR 892-1750 | | Manufacturer : | General Electric Medical Systems<br>283, rue de la Minière<br>78533 Buc Cedex, FRANCE | | Distributor: | GE Healthcare, P.O. Box 414, Milwaukee, WI 53210 | | Marketed Devices<br>listed below: | The Advanced Vessel Analysis II is substantially equivalent to the devices | | Model: Model: | Smart Vessel Analysis Option, 510(k) # K993792 CardIQ Analysis III, 510(k) # K041267 | - Auto Bone, 510(k) # K031871 Model: ● - General Electric Medical Systems, Buc, France . Manufacturer: #### Device Description: CT Advanced vessel analysis (AVA II / AVA 2) is a post processing analysis software package designed to assist Radiologists, Cardiologists, and other clinicians in the evaluation and assessment of vascular anatomy. - Advanced vessel analysis is a software post-processing package for the Advantage . . Workstation (AW) platform, CT scanners and PACS reading stations. It is an additional tool for the analysis of 3D CT angiographic images/data providing a number {1}------------------------------------------------ of display, measurements and batch filming/archive features to study user-selected vessels which include but are not limited to stenosis analysis, thrombus, pre/post stent planning procedures and directional vessel tortuosity visualization. #### Indications for Use: Advanced Vessel Analysis II is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. Advanced Vessel Analysis II (AVA II) is a post processing application option for the Advantage Workstation (AW) platform, CT Scanner or PACS stations, which can be used in the analysis of 2D and 3D CT Angiography images/data derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular and vascular disease assessment. This software is designed to support the physician in assessment of stenosis analysis, pre/post stent planning and directional vessel tortuosity visualization. AVA II automatic visualization tools provide the users with the capabilities to facilitate segmentation of bony structures for accurate identification of the vessels. Once vessels are visualized, tools are available for sizing the vessel, analyzing calcified and non-calcified plaque to determine the densities of plaque within a coronary artery, measure area's of abnormalities within a vessel. #### Comparison with Predicate: | Device Name | FDA Clearance Number | |-----------------------|----------------------| | Smart Vessel Analysis | K993792 | | CardIQ Analysis III | K041267 | | Auto Bone | K031871 | AVA II is substantially equivalent to the predicate devices listed above : #### Adverse Effects on Health: The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by: - · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements. - · Adherence to industry and international standards. #### Conclusions: Advanced Vessel Analysis II does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the Advanced Vessel Analysis II to be equivalent to those of CardIO Analysis III (K041267), Auto Bone (K031871), Smart Vessel Analysis (K993792). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white circular logo. The logo features a stylized bird with three wing-like shapes, suggesting movement or flight. Encircling the bird is text that reads 'DEPARTMENT OF HEALTH & HUMAN SERVICES - USA'. The text is arranged along the perimeter of the circle, with the bird emblem positioned in the center. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 GE Healthcare % Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Rd., Unit B7 TWINSBURG OH 44087 ## Re: K060779 Trade/Device Name: Advanced Vessel Analysis II Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 20, 2006 Received: March 22, 2006 Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. 5 2006 APR {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Nancy C. brigdon Nanev C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): __ バイム △ フフ9 Device Name: Advanced Vessel Analysis II Indications for Use: Advanced Vessel Analysis II is intended to provide an optimized non-invasive application to analyze vascular anatomy and pathology and aid in determining treatment paths from a set of Computed Tomography (CT) Angiographic images. Advanced Vessel Analysis II (AVA II) is a post processing application option for the Advantage Workstation (AW) platform, CT Scanner or PACS stations, which can be used in the analysis of 2D and 3D CT Angiography images/data derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular and vascular disease assessment. This software is designed to support the physician in assessment of stenosis analysis, pre/post stent planning and directional vessel tortuosity visualization. AVA II automatic visualization tools provide the users with the capabilities to facilitate segmentation of bony structures for accurate identification of the vessels. Once vessels are visualized, tools are available for sizing the vessel, analyzing calcified and non-calcified plaque to determine the densities of plaque within a coronary artery, measure area's of abnormalities within a vessel. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy C Hogdon 510(k) Num Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...