K041267 · General Electric Medical Systems · JAK · May 27, 2004 · Radiology
Device Facts
Record ID
K041267
Device Name
CARDIQ ANALYSIS III
Applicant
General Electric Medical Systems
Product Code
JAK · Radiology
Decision Date
May 27, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
CardIQ Analysis III is a CT image analysis software package, which allows the visualization of 2D and 3D medical image data of the heart derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular disease assessment. It provides functionally for 2D/3D rendering, assessment of calcified and non-calcified plaque to determine the densities of the plaque within a coronary artery, ventricular function of the heart, and measurement tools to detect coronary artery stenosis. This product can be used to aid a trained physician for to process, render, review, archive, print and visualizing cardiac anatomy and coronary vessels. CardIQ Analysis III will run on the AW workstation, scanner operator console and PACS system.
Device Story
CardIQ Analysis III is a post-processing software option for GE LightSpeed multi-slice CT scanners. It processes DICOM 3.0 compliant cardiac CT images to generate 2D/3D renderings (MIP, reformat, volume rendering). The software enables physicians to measure vessel diameters, assess plaque density (calcified/non-calcified), and calculate ventricular function parameters using end-systole and end-diastole images. Operated by trained physicians on AW workstations, scanner consoles, or PACS systems, the output aids in clinical decision-making regarding cardiovascular disease. Results are archived, printed, or saved to the workstation. The device benefits patients by providing quantitative data for cardiac anatomy and coronary vessel assessment.
Clinical Evidence
No clinical data provided. Safety and effectiveness are supported by software development, validation, and verification processes, and adherence to industry and international standards.
Technological Characteristics
Software-based post-processing package for CT angiographic images. Operates on GE AW workstations, scanner consoles, and PACS. Compatible with DICOM 3.0 data. Features include 2D/3D rendering, vessel diameter measurement, Hounsfield unit analysis for plaque density, and ventricular function calculation.
Indications for Use
Indicated for trained physicians to visualize and analyze 2D/3D cardiac CT image data for cardiovascular disease assessment, including plaque density assessment, ventricular function analysis, and coronary artery stenosis measurement.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K041267
MAY 2 7 2004
Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are black, creating a strong contrast against the white background.
GE Medical Systems
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
Larry A. Kroger, Ph.D. Submitter Senior Regulatory Programs Manager Tel: (262) 544-3894 Fax: (262) 544-4768 GE Medical Systems W-400 3000 North Grandview Blvd. Waukesha, WI 53188 USA Date Prepared: February 13, 2004.
## PRODUCT IDENTIFICATION
| Name: | CardIQ Analysis III |
|----------------------|---------------------------------------------------------------------------------------|
| Classification Name: | Accessory to Computed Tomography System |
| Manufacturer : | General Electric Medical Systems<br>283, rue de la Minière<br>78533 Buc Cedex, FRANCE |
| Distributor: | General Electric Medical Systems, Buc, France. |
The CardIQ Analysis III is substantially equivalent to the devices listed below: Marketed Devices
| Model: | CardIQ Analysis II |
|---------------|-------------------------------------------------------------|
| Manufacturer: | General Electric Medical Systems, Buc, France |
| 510(k) #: | K020796 |
| Model: | CardIQ Function |
| Manufacturer: | General Electric 3200 N. Grandview Blvd. Waukesha, WI 53188 |
| 510(k) #: | K013422 |
{1}------------------------------------------------
#### Device Description:
CardIO Analysis III is a post processing software option that can be used in the analysis of CT angiographic images to display structures of the heart in a MIP, reformat or volume rendering view. When the heart is displayed the software has the ability to measure the diameter of the vessel or hounsfield units within a coronary arteries to determine the size of a vessel or plaque density within a vessel. Functional parameters of the heart can also be determined when images of end systole and end diastole are present. Diarneters, densities, functional parameters and images can all be printed to reports or saved to the AW workstation. It is a software option for the GE family of LightSpeed multi-slice CT scanners.
### Indications for Use:
:
CardIQ Analysis III is a CT image analysis software package, which allows the visualization of 2D and 3D medical image data of the heart derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular disease assessment. It provides functionally for 2D/3D rendering, assessment of calcified and non-calcified plaque to determine the densities of the plaque within a coronary artery, ventricular function of the heart, and measurement tools to detect coronary artery stenosis. This product can be used to aid a trained physician for to process, render, review, archive, print and visualizing cardiac anatomy and coronary vessels. CardIQ Analysis III will run on the AW workstation, scanner operator console and PACS system.
#### Comparison with Predicate:
The functional features of the CardI() Analysis III software package are substantially equivalent to that of the following device:
| Device Name | FDA Clearance Number |
|--------------------|----------------------|
| CardIQ Analysis II | K020796 |
| CardIQ Function | K013422 |
#### Adverse Effects on Health:
The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by:
- · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements.
- · Adherence to industry and international standards.
#### Conclusions:
The CardIQ Analysis III does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features of the CardIQ Analysis III to be equivalent to those of CardIQ Analysis II (K020796) and CardIQ Function (K013422).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes a stylized depiction of an eagle with three lines representing its wings. The seal is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 27 2004
Mr. Tamas Borsai Program Manager 510(k) Review TUV Rheinland of North America, Inc. Medical Division, Newton Office 12 Commerce Road NEWTON CT 06470
Re: K041267
Trade/Device Name: CardIQ Analysis III Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography
x-ray system
Regulatory Class: II Product Code: 90 JAK Dated: May 3, 2004 Received: May 12, 2004
Dear Mr. Borsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device wo never reviewed your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass surve in the encrease 1976, the enactment date of the Medical Device Amendments, or to conniner of they to rialy 2011 accordance with the provisions of the Federal Food, Drug, de nocs that have boom require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, ate. En orovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
K041267
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are black, contrasting with the white background.
#### General Electric Medical Systems
# STATEMENT OF INTENDED USE
Device name: CardIQ Analysis III
#### Intended Use:
CardIQ Analysis III is a CT image analysis software package, which allows the visualization of 2D and 3D medical image data of the heart derived from DICOM 3.0 compliant CT scans for the purpose of cardiovascular disease assessment. It provides functionally for 2D/3D rendering, assessment of calcified and non-calcified plaque to determine the densities of the plaque within a coronary artery, ventricular function of the heart, and measurement tools to detect coronary artery stenosis. This product can be used to aid a trained physician for to process, render, review, archive, print and visualizing cardiac anatomy and coronary vessels. CardIQ Analysis III will run on the AW workstation, scanner operator console and PACS system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <div> -OR- </div> | Over-The-Counter Use |
|---------------------------------------|--------------------------|----------------------|
|---------------------------------------|--------------------------|----------------------|
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | 2041267 |
1/1PAGENOMFICHIER
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.