SYNGO, CT BONE READING

K123584 · Siemens Medical Solutions USA, Inc. · JAK · Mar 12, 2013 · Radiology

Device Facts

Record IDK123584
Device NameSYNGO, CT BONE READING
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeJAK · Radiology
Decision DateMar 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The syngo.CT Bone Reading is image analysis software for CT volume data sets which has been continuously acquired with computed tomography (CT) imaging systems. The software combines following digital image processing and visualization tools: - multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT) - geometric measurement tools (distance line, polyline, marker, arrow, angle) - HU measurement tools (Pixel lens, ROI circle, ROI polygonal, ROI freehand, VOI sphere) - curved MPR visualization (unfolded ribs and spine views), cross-section MPRs - tools for creation and editing of anatomical centerline paths - tools for creation and editing of anatomical labels The specific visualizations of spine and rib structures allow for easy manual identification and marking of pathologies such as bone lesions or fractures. Reporting and documentation of results is facilitated by using of appropriate reporting tool, statistics and creation of ranges and snapshots.

Device Story

syngo.CT Bone Reading is post-processing software for CT volume data; operates on syngo.via VA20 platform. Inputs: CT volume data sets. Processing: provides multiplanar reconstruction (MPR), maximum intensity projection (MIP), inverted MIP, volume rendering (VRT), geometric/HU measurements, and curved MPR for unfolded ribs/spine. Output: visualized images, anatomical labels, and reports. Used in clinical settings by radiologists/clinicians. Clinician maintains full control over navigation, measurements, and final diagnosis. Benefits: facilitates manual identification and documentation of bone lesions or fractures through specialized visualization tools.

Clinical Evidence

Bench testing only. Verification and validation testing confirmed software specifications met acceptance criteria. Risk analysis and hazard mitigation were performed. No clinical data provided.

Technological Characteristics

Software-based image analysis tool. Operates on syngo.via VA20 platform. Features: MPR, MIP, VRT, geometric/HU measurement tools, curved MPR, and anatomical labeling. Connectivity: DICOM compliant. Standards: IEC 60601-1-6 (usability), IEC 62304 (software lifecycle), ISO 14971 (risk management).

Indications for Use

Indicated for use by clinicians as an image analysis software tool for CT volume data sets to assist in the manual identification, marking, and reporting of bone pathologies, including lesions or fractures, in patients undergoing CT imaging.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Traditional 510(k) Submission: syngo.CT Bone Reading ## 510(K) SUMMARY FOR SYNGO.CT BONE READING ## MAR 1 2 2013 Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Date Prepared: February 26, 2013 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. #### 1. Contact Person: Mrs. Kimberly Mangum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly.mangum@siemens.com #### 2. Device Name and Classification Product Name: syngo.CT Bone Reading Propriety Trade Name: syngo.CT Bone Reading Classification Name: Computed Tomography X-ray System Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class Class II Product Code: 901AK {1}------------------------------------------------ KI23584 1 of 5 Image /page/1/Picture/1 description: The image shows the word "SIEMENS" in a stylized, outlined font. Each letter is clearly defined with a double-line border, giving it a bold and distinct appearance. The font style is consistent throughout the word, creating a uniform and recognizable logo or brand name. #### Substantial Equivalence: 3. Siemens syngo.CT Bone Reading post processing software package is substantially equivalent to the following medical devices in commercial distribution: | Predicate Device Name | FDA Clearance<br>Number | FDA Clearance Date | |------------------------------------------|-------------------------|--------------------| | Siemens syngo.CT Vascular Analysis | K112020 | 08/18/2011 | | Siemens syngo Multimodality<br>Workplace | K072728 | 04/22/2008 | | Siemens SOMATOM Definition Edge | K120579 | 05/23/2012 | #### 4. Device Description: syngo.CT Bone Reading is image analysis software for CT volume data sets which has been continuously acquired with computed tomography (CT) imaging systems. syngo.CT Bone Reading combines basic and advanced digital image processing and visualization tools for easy manual identification, marking and reporting of pathologies such as bone lesions or fractures. The software combines following digital image processing and visualization tools: - . multiplanar reconstruction (MPR) thin/thick, maximum intensity projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT) - geometric measurement tools (distance line, polyline, marker, ◆ arrow, angle) - HU measurement tools (Pixel lens, ROI circle, ROI polygonal, ROI . freehand, VOI sphere) - curved MPR visualization (unfolded ribs and spine views), cross-. section MPRs - . tools for creation and editing of anatomical centerline paths - . tools for creation and editing of anatomical labels Reporting and documentation of results is facilitated by using of appropriate reporting tool, statistics and creation of ranges and snapshots. Thereby, the user is always free to decide when or whether to apply particular tools. Accordingly, the user can operate the application in basic {2}------------------------------------------------ K123584 Page 3 of 5 # SIEMENS reading mode only and use conventional navigation on 2D and 3D views. Furthermore, during the whole reading process, the reading clinician has full control on the reported measurements, text and images. In the following sections we list all tools and explain in detail how different tools are operated, and which steps are to be followed in order to complete and report a clinical diagnosis. #### 5. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device: syngo.CT Bone Reading software package is designed to be operated on syngo via VA20 platform in a single or multi user environment. syngo.CT Bone Reading software package provides similar evaluation, reporting, and visualization tools and functionality as the predicate devices syngo MultiModality Workplace (K072728, clearance date April 22, 2008) and syngo.CT Vascular Analysis (K112020, clearance date August 18, 2011). syngo MultiModality Workplace contains visualization tools such as Curved Mode (in 3D application), Panoramic, and Paraxial Views. These tools can be used to generate reformatted MPR images which are equivalent to the curved and cross-section MPRs in syngo.CT Bone Reading. Predicate device svngo.CT Vascular Analysis also has equivalent visualization tools. Predicate device SOMATOM Definition Edge (K120579, clearance date May 23, 2012) includes similar image processing and visualization tools such as curved MPRs for ribs and spine as well as suggested labeling of ribs and spine. syngo.CT Bone Reading has similar indications for use and technological characteristics when compared to the predicate devices; therefore Siemens believes that syngo.CT Bone Reading is substantially equivalent to the predicate devices. #### 6. Nonclinical Testing: syngo.CT Bone Reading is designed to fulfill the requirements of following standards: - t IEC 60601-1-6 : 2006; Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability - IEC 62304 Ed. 1.0, "Medical Device Software Software Lifecycle . Processes" - ISO 14971:2007; Medical devices Application of risk management to . medical devices - DICOM (Digital Imaging and Communications in Medicine) Standard: . . 2008 DICOM conformity is fully covered by syngo.via implementations. {3}------------------------------------------------ Non clinical tests were conducted for syngo.CT Bone reading software package during product development. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence. #### 7. Indications for Use: The syngo.CT Bone Reading is image analysis software for CT volume data sets which has been continuously acquired with computed tomography (CT) imaging systems. The software combines following digital image processing and visualization tools: - multiplanar reconstruction (MPR) thin/thick, maximum intensity . projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT) - geometric measurement tools (distance line, polyline, marker, . arrow, angle) - HU measurement tools (Pixel lens, ROI circle, ROI polygonal, ROI . freehand, VOI sphere) - . curved MPR visualization (unfolded ribs and spine views), crosssection MPRs - tools for creation and editing of anatomical centerline paths . - tools for creation and editing of anatomical labels . The specific visualizations of spine and rib structures allow for easy manual identification and marking of pathologies such as bone lesions or fractures. Reporting and documentation of results is facilitated by using of appropriate reporting tool, statistics and creation of ranges and snapshots #### 8. General Safety and Effectiveness Concerns: The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards. {4}------------------------------------------------ K123584 Page 5 of 5 # SIEMENS ## Traditional 510(k) Submission: syngo.CT Bone Reading #### 9. Conclusion as to Substantial Equivalence In summary, Siemens is of the opinion that the syngo.CT Bone Reading software package does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 12, 2013 Siemens Medical Solutions, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 51 Valley Stream Parkway D-02 MALVERN PA 19355 Re: K123584 Trade/Device Name: Syngo.CT Bone Reading Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: Class II Product Code: JAK Dated: February 8, 2013 Received: February 15, 2013 Dear Ms. Mangum: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {6}------------------------------------------------ Page 2 - Ms. Mangum CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industrv/support/index.html. Sincerely yours, \$\ln .}) Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health for {7}------------------------------------------------ # Indications for Use 510(k) Number (if known): #### Device Name: syngo.CT Bone Reading ### Indications for Use: The syngo.CT Bone Reading is image analysis software for CT volume data sets which has been continuously acquired with computed tomography (CT) imaging The software combines following digital image processing and systems. visualization tools: - multiplanar reconstruction (MPR) thin/thick, maximum intensity ◆ projection (MIP) thin/thick, inverted MIP thin/thick, volume rendering technique (VRT) - geometric measurement tools (distance line, polyline, marker, ● arrow, angle) - HU measurement tools (Pixel lens, ROI circle, ROI polygonal, ROI freehand, VOI sphere) - curved MPR visualization (unfolded ribs and spine views), cross-. section MPRs - . tools for creation and editing of anatomical centerline paths - . tools for creation and editing of anatomical labels The specific visualizations of spine and rib structures allow for easy manual identification and marking of pathologies such as bone lesions or fractures. Reporting and documentation of results is facilitated by using of appropriate reporting tool, statistics and creation of ranges and snapshots. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) \$\sin^{-1}D\$ (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k) K123584
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...