K013561 · GE Medical Systems · JAK · Nov 8, 2001 · Radiology
Device Facts
Record ID
K013561
Device Name
LIGHTSPEED 4.0 CT SCANNER SYSTEM
Applicant
GE Medical Systems
Product Code
JAK · Radiology
Decision Date
Nov 8, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Intended Use
The LightSpeed 4 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications.
Device Story
LightSpeed 4.0 CT Scanner System is a diagnostic imaging device consisting of a gantry, patient table, console, and computer. It utilizes X-ray technology to acquire cross-sectional images of the head and whole body. The system features a multi-slice solid-state detector with thinner detector elements, enabling the acquisition of 16 slices per rotation. Operated by trained clinicians in a clinical setting, the device processes raw X-ray data to reconstruct images for diagnostic review. The output allows physicians to visualize internal anatomical structures, aiding in clinical decision-making and diagnosis. The system is designed to provide improved imaging capabilities compared to previous iterations while maintaining established safety and performance standards.
Clinical Evidence
Bench testing only. The device underwent system verification and validation to ensure performance to specifications, Federal Regulations, and user requirements. Compliance with industry and international standards (UL 2601-1, IEC 60601-1, 21 CFR 1020.30, 21 CFR 1020.33) was confirmed.
Technological Characteristics
Computed tomography X-ray system. Components: gantry, patient table, console, computer. Detector: multi-slice solid-state with thin elements. Standards: UL 2601-1, IEC 60601-1, 21 CFR Subchapter J. Energy source: X-ray. Connectivity: standard clinical imaging system architecture.
Indications for Use
Indicated for head and whole body X-ray Computed Tomography applications in patients requiring diagnostic imaging.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K033502 — LIGHTSPEED 6.0 CT SCANNER SYSTEM · Ge Medical Systems, LLC · Nov 20, 2003
K980176 — HISPEED QX/I CT SCANNER SYSTEM · General Electric Co. · Apr 13, 1998
K090173 — NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM · Neusoft Medical Systems Co., Ltd. · Feb 5, 2009
K052855 — STARSPEED SERIES CT SCANNER SYSTEMS · GE Healthcare · Oct 28, 2005
K092742 — NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM · Neusoft Medical Systems Co., Ltd. · Sep 22, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 0 82001
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The letters and the border are solid black, contrasting with the white background. The logo is simple and iconic, representing the well-known multinational conglomerate.
K013561
GE Medical Systems
General Electric Compa
Page 1 of 2
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
Larry A. Kroger, Ph.D. Submitter: Senior Regulatory Programs Manager TEL: 262-544-3894, FAX: 262-544-3863 Summary prepared: October 15, 2001
#### PRODUCT IDENTIFICATION
| Name: | LightSpeed 4.0 CT Scanner System |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Computed Tomography X-ray System |
| Manufacturer: | General Electric Medical Systems<br>16800 W. Ryerson Road<br>New Berlin, WI 53151 |
| Distributor: | Same as Manufacturer |
| Marketed Devices: | The LightSpeed 4.0 CT Scanner System is of comparable type and<br>substantially equivalent to currently marketed Computed<br>Tomography X-ray Systems that comply with the same or equivalent<br>standards and have the same intended uses. |
### DEVICE DESCRIPTION
The LightSpeed 4.0 CT Scanner System is composed of a gantry, patient table, console, computer, and associated accessories.
Materials: Materials and construction are equivalent to the LightSpeed 3.0 CT Scanner System (K002978) are compliant with UL 2601-1, IEC 60061-1, and 21CFR Subchapter J.
Design: The system is designed to be a head and whole body CT scanner utilizing a new solid state detector, an intuitive Operator Console, and the same tube and similar features to the LightSpeed 3.0 CT Scanner System (K002978), but now capable of imaging 16 slices per rotation. The system will use a new multi-slice detector having thinner detector elements.
{1}------------------------------------------------
#### Indications for Use:
The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Cor Tomography applications.
#### Comparison with Predicate:
It is the opinion of GE Medical Systems that the LightSpeed 4.0 CT Scanner System is of a type and substantially equivalent to currently marketed head and whole body X-ray computed tomography systems with respect to design, material composition, energy source, and radiation characteristics. It will comply with the X-ray requirements of 21CFR1020.30 and1020.33, as well as the safety requirements of UL2601-1, IEC 60601-1 and collateral standards.
#### Adverse Effects on Health:
Potential electrical, mechanical and radiation hazards are identified in a risk management summary (hazard analysis) and controlled by:
- System verification and validation to ensure performance to specifications, Federal . Requlations, and user requirements.
- Adherence to industry and international standards. (UL/CSA and IEC). .
#### CONCLUSIONS
The LightSpeed 4.0 CT Scanner Systems does not result in any new potential safety risks and performs as well as or better than devices currently on the market. GE considers the LightSpeed 4.0 CT System to be equivalent to other marketed devices with the same indications for use and meeting similar standards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with healthcare. The caduceus is depicted with three lines representing the snakes entwined around a staff.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 82001
General Electric Medical Systems % Mr. Reiner Krumme Division Manager, Medical Division TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470
Re: K013561
Trade/Device Name: LightSpeed 4.0 CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: October 22, 2001 Received: October 26, 2001
Dear Mr. Krumme:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for do bear of to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may , ators, and second the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
NOV 0 8 2001
## STATEMENT OF INTENDED USE
K013561
510(k) Number (if known): __ Kor 356 /
Device Name: LightSpeed 4.0 CT Scanner System
#### Indications For Use:
The LightSpeed 4 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) -OR- Over-The-Counter Use
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K013561
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.