Deep Learning Image Reconstruction for Gemstone Spectral Imaging

K201745 · Ge Medical Systems, LLC · JAK · Dec 10, 2020 · Radiology

Device Facts

Record IDK201745
Device NameDeep Learning Image Reconstruction for Gemstone Spectral Imaging
ApplicantGe Medical Systems, LLC
Product CodeJAK · Radiology
Decision DateDec 10, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesAI/ML

Intended Use

The Deep Learning Image Reconstruction for Gemstone Spectral Imaging option is intended for whole body, vascular, and contrast enhanced head CT applications.

Device Story

Deep Learning Image Reconstruction for Gemstone Spectral Imaging (DLIR-GSI) is a software-based CT reconstruction method. It takes dual-energy X-ray transmission raw data as input. A dedicated Convolutional Neural Network (CNN) transforms this data into cross-sectional images (monochromatic, material decomposition, virtual unenhanced). Used in clinical settings on GE Revolution CT systems; operated by radiologists. Output is displayed as 'TrueFidelity' images. The device improves image noise texture and maintains/improves performance metrics (LCD, CNR, spatial resolution) compared to ASiR-V. Benefits include diagnostic-quality images for routine and challenging clinical cases, aiding physician decision-making.

Clinical Evidence

Clinical reader study using 40 retrospective cases evaluated by 5 board-certified radiologists. Images reconstructed with ASiR-V vs. DLIR-GSI were assessed using a 5-point Likert scale. Results confirmed DLIR-GSI produces diagnostic quality images with preferred noise texture. Additional evaluation of 7 challenging cases with small, low-contrast objects confirmed adequate visualization for diagnostic use.

Technological Characteristics

Software-based reconstruction using a dedicated Convolutional Neural Network (CNN). Integrated into CT scanner raw data-based reconstruction chain. Compatible with dual-energy scan modes. Supports three user-selectable reconstruction strengths (Low, Medium, High). DICOM compatible output. Developed under ISO 13485 and 21 CFR 820.

Indications for Use

Indicated for all ages. Deep learning-based CT reconstruction method for cross-sectional images via computer reconstruction of dual energy X-ray transmission data acquired with Gemstone Spectral Imaging. Used for whole body, vascular, and contrast-enhanced head CT applications.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 10, 2020 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. GE Medical Systems, LLC % Ms. Katelyn Rowley Regulatory Affairs Leader 3000 N. Grandview Blvd. WAUKESHA WI 53188 ## Re: K201745 Trade/Device Name: Deep Learning Image Reconstruction for Gemstone Spectral Imaging Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: November 13, 2020 Received: November 16, 2020 Dear Ms. Rowley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ Page 2 combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are also blue, and they are surrounded by a white, decorative border. The logo is simple and recognizable, and it is often used to represent the company. Section 4: Indications for Use Deep Learning Image Reconstruction for Gemstone Spectral Imaging (GSI_DLIR) {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K201745 Device Name Deep Learning Image Reconstruction for Gemstone Spectral Imaging Indications for Use (Describe) The Deep Learning Image Reconstruction for Gemstone Spectral Imaging option is a deep learning based CT reconstruction method intended to produce cross-sectional images by computer reconstruction of dual energy X-ray transmission data acquired with Gemstone Spectral Imaging, for all ages. Deep Learning Image Reconstruction for Gemstone Spectral Imaging can be used for whole body, vascular, and contrast enhanced head CT applications. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by a swirling, wave-like design, also in blue, which gives the impression of motion or energy. ## 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h): | Date: | June 25, 2020 | |-------------------------------------|------------------------------------------------------------------| | Submitter: | GE Medical Systems, LLC | | | 3000 North Grandview Blvd | | | Waukesha, WI 53188 | | Primary Contact: | Katelyn Rowley | | | Regulatory Affairs Leader | | | Phone 262-309-5888 | | | Email: Katelyn.rowely@ge.com | | Secondary Contacts: | Helen Peng | | | Senior Regulatory Affairs Director | | | GE Healthcare | | | Tel: 262-424-8222 | | | Email: hong.peng@med.ge.com | | Device Trade Name: | Deep Learning Image Reconstruction for Gemstone Spectral Imaging | | Device Classification | Class II | | Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK | | Predicate Device Information | | |-------------------------------------|--------------------------------------------------------| | Device Name: | Deep Learning Image Reconstruction | | Manufacturer: | GE Medical Systems, LLC | | 510(k) Number: | K183202 cleared on April 12, 2019 | | Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK | ## Reference Devices Information | Device Name: | ASiR-V | |----------------|-----------------------------------| | Manufacturer: | GE Medical Systems, LLC | | 510(k) Number: | K134640 cleared on March 25, 2014 | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are also some white swirls around the letters. The logo is simple and recognizable. Regulation Number/ 21 CFR 892.1750 Computed tomography x-ray system / JAK Product Code: #### Reference Devices Information | Device Name: | Revolution CT | |-------------------------------------|--------------------------------------------------------| | Manufacturer: | GE Medical Systems, LLC | | 510(k) Number: | K163213 cleared on December 16, 2016 | | Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK | #### Device Description Deep Learning Image Reconstruction for Gemstone Spectral Imaging (DLIR-GSI) is the next step in CT reconstruction advancement. Like its predicate device (DLIR), DLIR-GSI is an image reconstruction method that uses a dedicated Convolution Neural Network (CNN) that has been designed and trained specifically to reconstruct CT GSI Images to give an image appearance similar to traditional FBP images while maintaining or improving the performance of ASiR-V. The DLIR-GSI can generate monochromatic images (MC), material decomposition images (MD), and virtual unenhanced images (VUE). Multiple MD images such as lodine, Water, Calcium, Hydroxyapatite (HAP), Fat, Uric Acid can be prescribed by the user and generated by the subject device. DLIR-GSI demonstrates same or better Imaging performance as compared to ASiR-V in the following areas: low contrast detectability (LCD), image noise, contrast to noise ratio (CNR), high contrast spatial resolution, CT number accuracy, MD quantification accuracy and metal artifact reduction. Reconstruction times with DLIR-GSI support a normal throughput for routine CT. The device is marketed as Deep Learning Image Reconstruction for Gemstone Spectral Imaging and the images produced are branded as "TrueFidelity™ CT Images". Deep Learning Image Reconstruction for Gemstone Spectral Imaging is compatible with dual energy scan modes using the standard kernel and was trained specifically on the Revolution CT family of systems (K163213, K133705, K19177). The deep learning technology is integrated into the scanner's existing raw data-based image reconstruction chain to produce DICOM compatible "TrueFidelity™ CT Images". The system allows user selection of three strengths of DLR-GSI: Low, Medium, or High. The strength selection will vary with individual users' preference for the specific clinical need. As compared to the predicate device, the intended use of Deep Learning Image Reconstruction for Gemstone Spectral Imaging does not change (head and whole-body CT image reconstruction). Both algorithms are designed to produce cross-sectional images of the head and body by computer reconstruction of X-ray transmission data, for all ages. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle has small white flourishes around its perimeter, giving it a dynamic and slightly vintage appearance. ## Intended Use The Deep Learning Image Reconstruction for Gemstone Spectral Imaging option is intended for whole body, vascular, and contrast enhanced head CT applications. ## Indications for Use The Deep Learning Image Reconstruction for Gemstone Spectral Imaging option is a deep learning based CT reconstruction method intended to produce cross-sectional images by computer reconstruction of dual energy X-ray transmission data acquired with Gemstone Spectral Imaging, for all ages. Deep Learning Image Reconstruction for Gemstone Spectral Imaging can be used for whole body, vascular, and contrast enhanced head CT applications. ## Comparisons The Deep Learning Image Reconstruction for Gemstone Spectral Imaging (DLIR-GSI) employs the same fundamental technology as that of the predicate device. In fact, the DLIR-GSI deep learning neural network was modified from that of the predicate. The DLIR-GSI option is substantially equivalent to the predicate device, DLIR reconstruction option. Because both the DLIR-GSI and the predicate device DLIR reconstruction for single energy are implemented on reference device, the Revolution CT family of Systems with GSI (K163213) they utilize the same hardware and software platform technology on which substantial equivalency is demonstrated. The table below summarizes the substantive feature/technological similarities and differences between the predicate device and the proposed device: {7}------------------------------------------------ # GE Healthcare 510(k) Premarket Notification Submission – DLIR-GSI Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The letters and the circle are in a blue color. There are also some white swirls around the letters. | Specification/ Attribute | Deep Learning Image<br>Reconstruction<br>(Predicate Device, K183202) | Deep Learning Image<br>Reconstruction for Gemstone<br>Spectral Imaging<br>(Proposed Device) | |-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Compatible Image Types | Compatible with single energy scan<br>modes and standard kernel. | Compatible with dual energy<br>scan modes and standard kernel. | | IQ performance vs dose | Low contrast detectability (LCD),<br>image noise, contrast to noise ratio<br>(CNR), and high contrast spatial<br>resolution are as good or better<br>than ASiR-V when substituted using<br>raw data from the same scan. | Same | | Technology | Utilizes a dedicated neural network<br>which is trained using single energy<br>acquired images on the CT Scanner<br>and designed specifically to<br>generate high quality CT images. | Utilizes a dedicated neural<br>network which is trained using<br>dual energy acquired images on<br>the CT Scanner and designed<br>specifically to generate high<br>quality CT images. | | System statistics - Noise<br>modeling of the data<br>collection imaging chain<br>(photon noise and<br>electronic noise) | Characterization of the photon<br>statistics as it propagates through<br>the preprocessing and calibration<br>imaging chain. | Same | | System statistics - Noise<br>characteristics of the<br>reconstructed images | Utilizes a trained neural network<br>which models the scanned object<br>using information obtained from<br>extensive phantom and clinical<br>data. | Same | | Clinical Workflow | Select recon type and strength<br>(Low, Medium, High). | Same | Deep Learning Image Reconstruction for Gemstone Spectral imaging does not introduce any additional risks/hazards, warnings, or limitations. {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. The logo is simple, recognizable, and represents the multinational conglomerate corporation. ## Determination of Substantial Equivalence ## Summary of Non-Clinical Testing Deep Learning Image Reconstruction for Gemstone Spectral Imaging has successfully completed the design control testing per our quality system. No additional hazards were identified, and no unexpected test results were observed. Deep Learning Image Reconstruction for Gemstone Spectral Imaging was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GE believes that the extensive bench testing and the physician evaluation are sufficient for FDA's substantial equivalence determination. The following quality assurance measures have been applied to the development of the system: - Requirement Definition ● - . Risk Analysis - Technical Design Reviews - . Formal Design Reviews - . Code Review - Software Unit Implementation - Software Integrations and Integration Testing - System Testing - o Safety Testing (Verification) - o Image Performance Testing (Verification) - O Simulating Use Testing (Validation) - Software Release . The testing and results did not raise different questions of safety and effectiveness than associated with predicate device. We consider the proposed device is substantially equivalent to the predicate device, DLIR. The substantial equivalence is also based on the software documentation for a "Moderate" level of concern. ## Additional Non-Clinical Testing Engineering bench testing was performed to support substantial equivalence and the product performance claims. The evaluation and analysis used the identical raw datasets obtained on GE's Revolution CT family of systems and then applies the Deep Learning Image Reconstruction for Gemstone Spectral Imaging or ASiR-V reconstruction. The resultant images were then compared for: - Low contrast detectability (LCD) - . Image noise (pixel standard deviation) - . High contrast spatial resolution (MTF) - Contrast to noise ratio (CNR) - CT Number accuracy - I CT Number uniformity - . Material Decomposition accuracy - 트 lodine detection {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle is surrounded by three white, curved shapes that resemble water droplets or stylized flames, adding a dynamic element to the design. - . Metal artifact reduction - . Pediatric test with anthropomorphic phantom ## Clinical Testing The reader study used a total of 40 retrospectively collected cases which represent typical and challenging clinical cases. The raw data from each of these cases was reconstructed with both ASIR-V and Deep Learning Image Reconstruction for Gemstone Spectral Imaging and presented to each reader independently. The results of the study support substantial equivalence and performance claims. These images were read by 5 board-certified radiologists with expertise in the specialty areas that align with the anatomical region of each case. Each image was read by 3 different radiologists who provided an assessment of image quality related to diagnostic use according to a 5-point Likert scale. Three readers read the cases primarily covering body and extremity anatomy, three readers read the cases primarily covering contrast enhanced head/neck anatomy. One reader was qualified to read both body and head/neck anatomy. Additionally, the readers were asked to compare directly the ASiR-V and Deep Learning Image Reconstruction for Gemstone Spectral Imaging images according to the key metric of image noise texture. The result of this reader study confirmed that the DLIR-GSI (the subject device) produce diagnostic quality images and have preferred noise texture than the reference device ASiR-V. To further challenge the DLR-GSI reconstruction algorithm, 7 additional retrospectively collected clinically challenging cases containing small, low contrast objects were evaluated. These images were read by a board-certified radiologist with expertise in the specialty area that aligns with all cases. The reader confirmed that all object(s) were adequately visualized for diagnostic use using DLIR-GSI. This additional clinical evaluation further confirmed that DLIR-GS! produces diagnostic quality images, even for extremely clinically challenging cases. ## Substantial Equivalence The changes associated with Deep Learning Image Reconstruction for Gemstone Spectral Imaging do not change the Intended Use from the predicate, and represent equivalent technological characteristics including the dedicated neural network, with no impact on control mechanism or operating principle. Deep Learning Image Reconstruction for Gemstone Spectral Imaging was developed under GE Healthcare's quality system. Design verification, along with bench testing and the clinical reader study provided in this submission demonstrates that Deep Learning Image Reconstruction for Gemstone Spectral Imaging is substantially equivalent and hence as safe and as effective as the legally marketed predicate device. GE's quality system's design, verification, and risk management processes did not identify any additional hazards, unexpected results, or adverse effects stemming from the changes to the predicate. {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are also some white swirls around the letters. The logo is simple and recognizable. Based on development under GE Healthcare's quality system, the successful verification and validation testing, including the additional engineering bench testing, and the clinical reader study, GE Healthcare believes that Deep Learning Image Reconstruction for Gemstone Spectral Imaging is substantially equivalent to the predicate device and hence is safe and effective for its intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...