K133705 · Ge Medical Systems, LLC · JAK · Apr 11, 2014 · Radiology
Device Facts
Record ID
K133705
Device Name
REVOLUTION CT
Applicant
Ge Medical Systems, LLC
Product Code
JAK · Radiology
Decision Date
Apr 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles. The system has the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc.. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results. The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Device Story
Revolution CT is a 256-detector row, multi-slice CT scanner. Inputs: X-ray transmission projection data acquired via rotating gantry (up to 0.28s/rotation). Transformation: Computer reconstruction of data into cross-sectional images; utilizes Volumetric High Definition (VHD) algorithm to reduce cone-beam artifacts and ASiR-V iterative reconstruction for noise reduction/dose management. Output: Transaxial and reformatted images, post-processed analysis results. Used in clinical settings by radiologists/technicians. Features 160mm z-coverage for whole-organ imaging in single rotation; 80cm bore for patient comfort. SnapShot Freeze technology improves temporal resolution for cardiac imaging. Output aids diagnosis, therapy planning, and monitoring. Benefits include reduced radiation dose, faster scan speeds (potentially reducing contrast volume), and improved image quality.
Clinical Evidence
Clinical evaluation involved 49 subjects at one US site. Data covered cardiac (CTA, perfusion, function, calcium, gated chest, TAVI), body/extremity (abd/pelvis, chest, ankle, shoulder, knee, spine), and neuro (sinus, brain, angiography, perfusion) scans. Multiple radiologists evaluated images for clinical acceptance and quality using a 5-point Likert scale. Results demonstrated acceptable diagnostic imaging performance. Non-clinical testing included phantom studies, model observer studies for low contrast detectability, and performance verification for temporal resolution and dose.
Technological Characteristics
Multi-slice (256-row) CT scanner; 160mm z-coverage; 80cm bore. Components: Gantry, patient table, operator console, PDU. Detector: Gemstone Clarity with 3D collimator. Tube: Performix HDw. Reconstruction: ASiR-V (iterative) and VHD (Volumetric High Definition). Standards: AAMI/ES 60601-1, IEC 60601-1 (3rd ed), 21CFR Subchapter J, NEMA XR-25. Software: Moderate level of concern. Connectivity: Networked system.
Indications for Use
Indicated for head, whole body, cardiac, and vascular X-ray CT applications in patients of all ages. Used for diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy. Images may be obtained with or without contrast.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
## GE Healthcare
510(k) Premarket Notification Submission- Revolution CT
Image /page/0/Picture/3 description: The image shows a handwritten document with the text "K133705 Page 1 of 6" along with the GE logo. The text appears to be part of a document header or footer, possibly indicating a document identification number, page number, and the company associated with the document. The GE logo is a circular emblem with the letters "GE" inside.
ut. ady
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
Date:
November 20th, 2013
Submitter:
GE Medical Systems, LLC (GE Healthcare) 3000 N. Grandview Blvd. Waukesha, WI 53188
Primary Contact:
Helen Peng Regulatory Affairs Leader, RAC GE Medical Systems, LLC Tel: 262-424-8222 e-mail: hong.peng@ge.com
Secondary Contact:
John Jaeckle Chief Requlatory Affairs Strategist GE Medical Systems, LLC Tel: 262-424-9547 e-mail: john.jaeckle@ge.com
#### PRODUCT IDENTIFICATION
Device Trade Name:
Classification Name:
Revolution CT
90-JAK
Common/Usual Name:
Computed Tomography X-ray System
Computed Tomography X-ray System per 21CFR892.1750
Product Code:
Manufacturer / Design Location:
Manufacturing Location(s):
Distributor:
GE Medical Systems, LLC (GE Healthcare) 3000 N. Grandview Blvd. Waukesha. WI 53188
GE Medical Systems, LLC (GE Healthcare) 3000 N. Grandview Blvd. Waukesha, WI 53188
GE Medical Systems, LLC (GE Healthcare) 3000 N. Grandview Blvd. Waukesha, WI 53188
5-3
{1}------------------------------------------------
## GE Healthcare
510(k) Premarket Notification Submission- Revolution CT
Image /page/1/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular frame. The letters and the frame are black, creating a bold and recognizable symbol.
## Marketed Devices:
The Revolution CT is a new CT device built upon the existing technologies of the predicate device Discovery CT750 HD (K12083). It is of comparable type and substantially equivalent to its predicate device Discovery CT750 HD and GE's other currently marketed Computed Tomography X-ray Systems that comply with the same standards. In addition, the system has the same intended use as that of the predicate device. The proposed device's indications for use have been revised to match the system capabilities as substantiated in the engineering and clinical testing provided. The system is labeled as the Revolution CT.
#### Predicate Device:
#### Discovery CT750 HD- K120833
## DEVICE DESCRIPTION
The Revolution CT is a multi-slice (256 detector row) CT scanner consisting of a gantry, patient table, scanner desktop (operator console), system cabinet, power distribution unit (PDU), and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software, and associated accessories.
The system generates images through the computer reconstruction of data acquired at different andles and planes of the rotating gantry. The gantry rotates at up to 0.28 seconds per rotation, and can acquire up to 512 slices of image data per rotation with a maximum total coverage of 160 mm in the z direction. The gantry however is designed to be able to rotate at 0.20 second per rotation. The system can be operated in Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisition modes.
The Revolution CT system is a powerful Volume High Definition CT scanner that is designed to provide best-in-class technologies for whole organ coverage, high image quality and responsible dose performance with the following characteristics:
- . 160 mm detector coverage
- 140ms temporal resolution (0.28s rot. Speed) combined with intelligent motion . correction with SnapShot Freeze for excellent cardiac imaging at any heart rate.
- 0.23 mm spatial resolution .
- A wide bore (80-cm bore size) to image all patients allowing better patient . positioning & access.
- The next-generation of iterative reconstruction technology, ASiR-V, designed to . deliver ultra-low noise levels, improved low contrast detectability and may enable a reduction in dose for all clinical applications
Built upon the existing technologies the Revolution system is designed to use less radiation dose than the previous generation product while maintaining the same diagnostic level of image quality. Further, the fast speed of the scan could potentially reduce contrast volumes. The hardware platform is also capable of supporting Gemstone spectral imaging and 0.2s rotation speed.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
The Revolution CT is intended to be a head and whole body CT system incorporating the same basic fundamental operating principles and the same intended for use as the predicate device. Materials and construction are equivalent to our existing marketed products, which are compliant with AAMI/ES 60601-1, IEC 60601-1(30 ed) and associated collateral and particular standards, 21CFR Subchapter J, and NEMA XR-25. It has been developed under the same GE quality system and has successfully completed design controls activities, including risk management, verification, and validation.
## Intended Use
The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
### Indications for Use:
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles. The system has the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver , kidney, pancreas, etc.. The system may acquire data using Axial, Cine, Helical, Cardiac , and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results
The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy..
## Technology:
The Revolution CT is built on the existing technologies of the predicate device Discovery CT750 HD. The vast majority of the software features and functions are common between the two products. The software user interface has been redesigned to provide more simplified workflow and user experience. Hardware components including the imaging chain however have been upgraded to support the new and improved capabilities.
The most notable changes in the imaging chain as compared to the predicate device Discovery CT750 HD is described below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. Three swirling shapes surround the circle, adding a dynamic element to the design. The logo is presented in black and white.
The Revolution CT system features the new "Gemstone Clarity" detector consisting of 256 rows at 0.625mm row thickness and a full 160mm z-coverage, the redesigned gantry with a 80cm bore and improved iterative reconstruction technology, allowing the system to deliver excellent image quality at full 160mm coverage to enable whole organ imaging.
The Gemstone Clarity detector features a unique focally aligned layout of the detector submodules and a 3D collimator (post patient) to minimize scatter artifacts, ensure HU uniformity & reduce beam hardening artifacts associated with wide coverage systems.
To house the wide Gemstone Clarity detector, the Revolution gantry has been redesigned with high specifications of strength and rigidity that support fast rotation speeds for acquisitions with high temporal resolution. while providing an 80cm bore opening to accommodate patients with positioning flexibility and comfort. With an improved center mount design that balances the mechanical forces generated by the rotating components, the gantry and detector assembly is capable of supporting accelerations that ensures reliable operation at today's 0.28 seconds per rotation, and can be extended to support imaging at 0.20 seconds per rotation in the future.
On the rotating gantry, the new Performix HDw tube tailored for the wide angle geometry generates the X-ray beam for patient imaging.
To address the challenges of the wide cone-beam geometry of the new scanner, a new advanced image reconstruction algorithm called Volumetric High Definition (VHD) has been designed specifically to reduce cone-beam artifacts and maintain CT number uniformity.
The image reconstruction engine further includes ASiR-V, the next generation of GE's ASiR iterative recon technology for noise reduction and dose management. Finally, GE's SnapShot Freeze technology is supported on all cardiac acquisitions to improve the temporal properties of the reconstructed images when visualizing the coronaries.
Despite the changes described above, the control mechanism, operating principle, energy type, and intended use have not changed from the predicate device Discovery CT750 HD.
#### Adverse Effects on Health:
Potential electrical, mechanical, and radiation hazards are identified in risk management including hazard analysis and controlled by:
- · System verification and validation to ensure performance to specifications, Federal Regulations, and user requirements.
- · Adherence and certification to industry and international standards. (AAMI/ES and IEC60601-1 Ed.3 and associated collateral and particular standards for CT).
- · Compliance to applicable CDRH 21CFR subchapter J requirements.
- · Compliance to NEMA XR-25
The device is designed and manufactured under the Quality System Regulations of 21CFR 820.
#### Determination of Substantial Equivalence:
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular shape. The letters and the circle are white, set against a black background.
The Revolution CT has completed testing and is in compliance with IEC 60601-1 Ed. 3 and its associated collateral and particular standards, 21CFR Subchapter J, and NEMA XR-25. The device has successfully completed all testing per our quality system as well as comparison testing to the predicate device. It was designed and is manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures were applied to the development of the system:
- · Risk Analysis
- · Required Reviews
- Design Reviews .
- · Testing on unit level (Module verification)
- · Integration testing (System verification)
- · Performance testing (Verification)
- · Safety testing (Verification)
- · Simulated use testing (Validation)
GE believes the Revolution CT system is of comparable type and substantially equivalent to our currently marketed system Discovery CT750 HD.
The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
## Summary of Additional Testing
In addition to the verification and validation testing successfully completed as required by GE Healthcare's quality system, additional enqineering (non-Clinical testing) and clinical performance testing was performed to provide the requisite data to substantiate performance claims, the revised indications, and ultimately substantial equivalence.
## Non-Clinical Testing
The performance evaluation testing used a variety of test methods, phantoms, and clinical datasets. Various mathematical, physics and statistical analysis were performed to demonstrate that each performance specification was successfully verified and substantiated. The areas additionally evaluated for the non-clinical testing included cardiac, cardiovascular, and thoracic imaging; temporal resolution, dose performance, and image quality. The image quality evaluation included evaluation for artifacts, scatter, spatial resolution, and low contrast detectability. For the LCD evaluation, a model observer study was used along with the MITA LCD phantom. Acceptance testing per IEC 61223-3-5 was also conducted.
## Clinical Testing
Sample clinical data was collected from 49 subjects at one site in the US with the approval of appropriate ethics committee and in accordance with 21 CFR Parts 812, 50 and 56, as well as GE Healthcare's quality system's procedures for such evaluations.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a handwritten note with the text "K133705 page 6 of 6" along with the GE logo. The text appears to be a document identifier and page number. The GE logo is a circular emblem with the letters "GE" intertwined in the center.
The intent of the protocol was to obtain a sample set of clinical images across different patient populations, clinical scenarios, and scanning protocols/techniques. Patients were selected for potential recruitment to meet these needs. Any patient who met these criteria stated in the Protocol and who voluntarily signed the Informed Consent Form was recruited.
This sample image data was representative of a wide range of anatomical coverage and patient indications and was categorized into the following types of scans:
- Cardiac Coronary CTA, Stress/Rest Perfusion, Cardiac Function, Calcium . Burden, Gated Chest (Triple Rule Out), TAVI
- Body & Extremity Abd/Pelvis, Chest, Ankle, Shoulder, Knee, Spine .
- . Neuro - Sinus, General Brain, Contrast Enhanced Brain, Neuroangiography, Neuro Perfusion, Neuro 4D CTA
The images were evaluated by multiple readers who are qualified radiologists at different institutions in the United States of America for clinical acceptance and image quality using a 5 point Likert scale.
The results of this clinical assessment demonstrate the acceptable diagnostic imaging performance of the GE Healthcare Revolution CT scanner.
### Substantial Equivalence Conclusion:
Based on the conformance to standards, development under our quality system, and the engineering and clinical testing provided, GE Medical Systems believes that the Revolution CT is as safe and effective, and performs in a substantially equivalent manner to the predicate device Discovery CT750 HD (K120833).
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the logo.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-00002
April 11. 2014
GE Medical Systems. LLC % Ms. Helen Peng Regulatory Affairs Manager 3000 North Grandview Blvd. WAUKESHA WI 53188
Re: K133705
Trade/Device Name: Revolution CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 10, 2014 Received: March 11, 2014
Dear Ms. Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (sec above) into cither class 11 (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Ms. Peng
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetw/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
#### K133705
Device Name Revolution CT
#### Indications for Use (Describe)
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different anyles. The system has the creans in a single roation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc.. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal and display equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformation planes. Further, the images can be post processed to produce additional imaging planes or analysis results
The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Type of Use (Select one or both, as applicable)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE CHART FOR THE OF CHARDED FOR FDA USE QUES EN LE CONTRACTOR . . . . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
> Prescription Use (Part 21 CFR 801 Subpart D)
Michael D. O'Hara
FORM FDA 3881 (9/13)
**PAC Publishing Services (301) 443-6740 EP**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.