AQUILION ONE TSX-301A/2 COMPUTED TOMOGRAPHY SYSTEM

K072606 · Toshiba America Medical Systems, In.C · JAK · Oct 2, 2007 · Radiology

Device Facts

Record IDK072606
Device NameAQUILION ONE TSX-301A/2 COMPUTED TOMOGRAPHY SYSTEM
ApplicantToshiba America Medical Systems, In.C
Product CodeJAK · Radiology
Decision DateOct 2, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

This device is intended to produce cross sectional volume sets of the anatomy. Additionally the system, when used by a trained professional, is capable of proving data for dynamic and perfusion studies of organs and extremities.

Device Story

Whole-body CT scanner; captures cross-sectional volume data sets via ionizing radiation. System components: gantry, patient couch, peripheral cabinets for data processing/display. Operated by trained professionals in clinical settings. Transforms raw X-ray attenuation data into cross-sectional images; supports dynamic and perfusion studies of organs and extremities. Physician reviews output to inform clinical decision-making regarding anatomy and organ function. Benefits include high-resolution volumetric imaging of entire organs in single rotations.

Clinical Evidence

No clinical data provided; substantial equivalence supported by technological characteristics and adherence to safety standards.

Technological Characteristics

Whole-body CT scanner; ionizing radiation source. Components: gantry, patient couch, data processing/display cabinets. Conforms to IEC 60601-1, IEC 60601-2-32, and IEC 60601-2-44. Operates under 21 CFR Subchapter J (Federal Diagnostic X-ray Equipment Standard).

Indications for Use

Indicated for acquisition and display of whole-body cross-sectional volumes, including head and whole organs (e.g., brain, heart, pancreas) in a single rotation. Enables specialized whole-organ dynamic and perfusion studies by qualified physicians.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Toshiba America Medical Systems, Inc. 510(k) Pre-market Notification; Aquilion ONE TSX-301A/2 ## 510(k) Summary | | | OCT 2` 2007 | |---------------------------------------|-------------------------------------------------------------------------------------------------------|-------------| | Date: | September 12, 2007 | | | Submitter's Name: | Toshiba America Medical Systems, Inc. | | | Submitter's Address: | P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068 | | | Submitter's Contact: | Paul Biggins, Director Regulatory Affairs<br>(714)730-5000 | | | Establishment Registration<br>Number: | 2020563 | | | Device Proprietary Name: | TSX-301A/1, Aquilion ONE CT System, k063189<br>TSX-101A/h, /I Aquilion 32/64 SP CT System,<br>k051833 | | | Common Name: | Scanner, Computed Tomography, X-Ray<br>[Fed. Reg. No. 892.1750, Pro. Code: 90JAK] | | | Regulatory Class: | II (per 21 CFR 892.1750) | | | Performance Standard: | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment Standard | | | Predicate Device(s): | TSX-301A Aquilion ONECT System | | | Reason For Submission | Modification to existing device | | ### Description of this Device: The TSX-301A/2 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a gantry, patient couch (table) and peripheral cabinets used for data processing and display. ### Summary of Intended Uses: This device is intended to produce cross sectional volume sets of the anatomy. Additionally the system, when used by a trained professional, is capable of proving data for dynamic and perfusion studies of organs and extremities. {1}------------------------------------------------ ## Toshiba America Medical Systems, Inc. 510(k) Pre-market Notification; Aquilion ONE TSX-301A/2 #### Technological Characteristics: This device employs similar materials and processes as found in the predicate device. The device produces ionizing radiation that is employed to generate cross sectional images of the anatomy. #### Safety and Effectiveness Concerns: This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR § 1020.30 and 1020.33, that apply to this upgrade, will be met and reported via a supplement to the initial report for the predicate device. Additionally this system is in conformance with the applicable parts of the IEC 60601-1 {applicable portions}; IEC 60601-2-32, and IEC 60601-2-44. - Medical Device Safety standards. #### Substantial Equivalence: This device is similar in materials and processes to that of the predicate device, Toshiba TSX-301A Aquilion ONE CT scanner. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure with three curved lines representing wings or feathers. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 Toshiba America Medical Systems, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313 2007 Re: K072606 Trade/Device Name: TSX-301A/2, Aquilion ONE CT System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: September 14, 2007 Received: September 17, 2007 #### Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1906-2006" at the top, followed by the letters "PA" in a large, bold font. Below the letters, the word "Centennial" is written in a smaller font. Three stars are located at the bottom of the logo. Proscions and Promoting Public Statte {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html. Sincerely yours, Nancy C. Hogdon Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ ## Toshiba America Medical Systems, Inc. Pre-Market Notification 510(k) for TSX-301A/2 Aquilion ONE CT System ## Indications for Use 510{k) Number (if known): Device Name: TSX-301A/2, Aquilion ONE CT System Indications for Use: , . . This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc. The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can then be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of Tomu M. Whing fon Sign-Off) Division of Reproductive, AbdomInal and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...