Philips CT Big Bore Sliding Gantry Configuration

K181797 · Phillips Medical Systems (Cleveland), Inc. · JAK · Aug 3, 2018 · Radiology

Device Facts

Record IDK181797
Device NamePhilips CT Big Bore Sliding Gantry Configuration
ApplicantPhillips Medical Systems (Cleveland), Inc.
Product CodeJAK · Radiology
Decision DateAug 3, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesPediatric

Intended Use

The Philips CT Big Bore Sliding Gantry Configuration is a Computed Tomography X-Ray System intended to produce images of the head and body to be used for diagnostic imaging in radiology and in oncology as part of treatment preparation and radiation therapy planning. These systems are indicated for head and whole body X-ray Computed Tomography applications in patients of all ages and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.

Device Story

Computed Tomography (CT) system for diagnostic imaging and radiation therapy planning; produces images via computer reconstruction of X-ray transmission data. System comprises a gantry (stator, rotor, X-ray tube, generator, detector array), a stationary patient support (couch), and a console with host computer and image reconstruction system. Unlike standard configurations, the gantry rides on a carriage to move over a stationary patient. Operated by clinicians in radiology/oncology settings. Output consists of axial images and post-processed views used for diagnostic assessment and treatment planning. Benefits include precise patient positioning and imaging for oncology workflows.

Clinical Evidence

Bench testing only. Performance verified via phantom testing (MTF, CTDIvol, CT number accuracy, noise, slice thickness) and system-level verification/validation protocols. No clinical study required for substantial equivalence.

Technological Characteristics

Computed tomography X-ray system. Components: Gantry (stator, rotor, X-ray tube, generator, detector array), stationary patient couch, console. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-44, ANSI AAMI IEC 62304, IEC 62366, ISO 14971. Software lifecycle per IEC 62304. Connectivity: Workstation for data acquisition, display, storage, and DICOM-compatible post-processing.

Indications for Use

Indicated for head and whole-body CT imaging in oncology, vascular, and cardiology applications for patients of all ages, including low-dose CT lung cancer screening for early detection of lung nodules.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 3, 2018 Philips Medical Systems (Cleveland) Inc. % Michael Chilbert, Ph.D., P.E. Regulatory Affairs Engineer 595 Miner Road CLEVELAND OH 44143 Re: K181797 Trade/Device Name: Philips CT Big Bore Sliding Gantry Configuration Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: June 22, 2018 Received: July 5, 2018 Dear Dr. Chilbert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Hole 2. Nils Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181797 Device Name Philips CT Big Bore Sliding Gantry Configuration #### Indications for Use (Describe) The Philips CT Big Bore is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipments and accessories. These systems are indicated for head and whole body X-ray Computed Tomography applications in oncology, vascular and cardiology, for patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. * Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text. # 510(k) Summary of Safety and Effectiveness [As required by 21 CFR 807.92(c)] | Applicant's Name:<br>Address: | Philips Medical Systems (Cleveland), Inc.<br>595 Miner Road<br>Cleveland, OH<br>44143 USA | |-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | Contact Person:<br>Title:<br>Address: | Michael Chilbert, Ph.D., P.E.<br>Regulatory Affairs Engineer<br>595 Miner Road<br>Cleveland, OH<br>44143 USA | | Telephone number:<br>E-mail: | +1 440 869-4576<br>michael.chilbert@philips.com | | 510(k) Summary Date of<br>Preparation: | 15-June-2018 | | Device Trade Name: | Philips CT Big Bore Sliding Gantry Configuration | | Common or Usual Name: | Computed Tomography X-ray system | | Classification<br>Name:<br>Regulation:<br>Class:<br>Product Code:<br>Panel: | Computed Tomography X-ray system<br>21 CFR 892.1750<br>II<br>JAK<br>Radiology | | Primary Predicate device | K171850 - Phillips CT Big Bore | | Classification<br>Name:<br>Regulation:<br>Class:<br>Product Code:<br>Panel: | Computed Tomography X-ray system<br>21 CFR 892.1750<br>II<br>JAK<br>Radiology | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. ## Indications for Use The Philips CT Big Bore Sliding Gantry Configuration is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. These systems are indicated for head and whole body X-ray Computed Tomography applications in oncology, vascular and cardiology, for patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of protocols that have been approved and published by either a governmental body or professional medical society. *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. ### Intended Use: The Philips CT Big Bore Sliding Gantry Configuration is a Computed Tomography X-Ray System intended to produce images of the head and body to be used for diagnostic imaging in radiology and in oncology as part of treatment preparation and radiation therapy planning. These systems are indicated for head and whole body X-ray Computed Tomography applications in patients of all ages and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. ### Device Description: The Philips CT Big Bore is available in three system configurations, the Oncology configuration the Radiology (Base) configuration, and the new Sliding Gantry configuration. The main components (detection system, the reconstruction algorithm, and the x-ray system) that are used in the Philips CT Big Bore have the same fundamental design characteristics and are based on comparable technologies as the predicate. The main system modules and functionalities are: - 1. Gantry. The Gantry consists of 4 main internal units: - a. Stator a fixed mechanical frame that carries HW and SW - b. Rotor A rotating circular stiff frame that is mounted in and supported by the stator. - c. X-Ray Tube (XRT) and Generator fixed to the Rotor frame - d. Data Measurement System (DMS) a detector array, fixed to the Rotor frame - 2. Patient Support (Couch) supports the patient in a stationary position while the gantry moves in and out on a carriage. - 3. Console A two part subsystem containing a Host computer and display that is the primary user interface and the Common Image Reconstruction System (CIRS) - a dedicated, powerful image reconstruction computer {5}------------------------------------------------ In addition to the above components and the software operating them, each system includes workstation hardware and software for data acquisition, display, manipulation, storage and filming as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient. The Sliding Gantry Configuration provides the following features. - . The patient table is stationary during the scan. - . Patient is located on the stationary table. - To perform imaging, CT gantry rides on a carriage and moves over the patient. # Substantial Equivalence: = Primary Predicate Device: Phillips CT Big Bore Manufacturer: Philips Medical Systems (Cleveland), Inc. Predicate Device k#: K171850 The design, intended use and technology provided with the proposed Philips CT Big Bore Sliding Gantry Configuration is identical to the predicate device, Phillips CT Big Bore, and therefore is considered substantially equivalent. | Table 5-2: Substantial Equivalence Comparison<br>Comparison of the primary predicate device, Phillips CT Big Bore (K171850)<br>versus the proposed Philips CT Big Bore Sliding Gantry Configuration | | | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | Philips CT Big<br>Bore | Proposed<br>Philips CT<br>Big Bore Sliding<br>Gantry Configuration | Comments | | Gantry (Stator,<br>Rotor, X-Ray Tube<br>and Generator,<br>Data Measurement<br>System) | No change | No change | Identical | | Software | No change | No change | Identical | | Patient Support | Couch slides<br>horizontally, Gantry<br>is stationary | Couch is stationary in<br>the horizontal axis,<br>Gantry slides<br>horizontally | See table 10.1 for details. This<br>operational change does not<br>introduce new hazards and has no<br>effect on the safety or effectiveness<br>of the device. | | Console | No change | No change | Identical | | Environmental<br>Requirements | No change | No change | Identical | | Table 5-2: Substantial Equivalence Comparison<br>Comparison of the primary predicate device, Phillips CT Big Bore (K171850)<br>versus the proposed Philips CT Big Bore Sliding Gantry Configuration | | | | | Feature | Philips CT Big<br>Bore | Proposed<br>Philips CT<br>Big Bore Sliding<br>Gantry Configuration | Comments | | Horizontal<br>Movements,<br>minimum<br>increments | No change | No change | identical | | Horizontal position<br>precision planning | No change | No change | identical | | Horizontal position<br>repeatability | No change | No change | identical | | Horizontal speed | Maximum Speed =<br>185 mm/sec<br>Minimum Speed =<br>0.5 mm/sec | Maximum Speed = 150<br>mm/sec<br>Minimum Speed = 1<br>mm/sec | Clinical scans on Big Bore scanner<br>use speeds in the range of 1mm/sec<br>to 60mm/sec, thus there is no impact<br>to these differences. | | Horizontal (Linear)<br>movement when<br>emergency stop is<br>actuated | No change | No change | Identical | | Horizontal<br>obstruction<br>detection | Couch stops<br>horizontal motion<br>when obstructed. | Gantry stops horizontal<br>motion when obstructed. | The operational change does not<br>introduce new hazards. | | Collision envelope | 25 mm gap<br>requirement is met.<br>Operator is<br>monitoring motion of<br>couch in relation to<br>the gantry. | 25 mm gap requirement<br>is met. Operator is<br>monitoring the motion of<br>the sliding gantry in<br>relation to the stationary<br>couch. | The operator is instructed to observe<br>the patient position while performing<br>the scan, as in the case of the<br>predicate device. This operational<br>change does not introduce new<br>hazards. | | Couch Vertical<br>Movements | Controlled by CT<br>(Gantry) | Patient support is<br>provided by a third party<br>couch, which operates<br>independently. | This operational change does not<br>introduce new hazards and has no<br>effect on the safety or effectiveness<br>of the device. | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The overall impression is clean and modern, reflecting the brand's identity. ## Summary of Non-Clinical Testing: This 510(k) premarket notification contains technical documentation, which demonstrates that the proposed Philips CT Big Bore Sliding Gantry Configuration is substantially equivalent to the primary predicate device, Phillips CT Big Bore in terms of safety and effectiveness. Testing was performed on the proposed Philips CT Big Bore Sliding Gantry Configuration according to the following international and FDA recognized consensus standards and FDA requlations and quidance documents: - . ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2: Ed 3 2007-03, Medical electrical equipment -- Part 1-2: General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests - . IEC 60601-1-3 Ed 2.1:2013-04 Medical electrical equipment -- Part 1-3: General requirements for basic safety - Collateral standard: Radiation protection in diagnostic Xray equipment - IEC 60601-1-6: Ed 3.1 2013-10 Medical electrical equipment -- Part 1-6: General . requirements for basic safety and essential performance - Collateral standard: Usability - . IEC 60601-2-44:Ed 3.1 2012 Medical electrical equipment -- Part 2-44: Particular requirements for the safety of X-rav equipment - . ANSI AAMI IEC 62304:2006 First Edition Medical device software -- Software life cycle processes - IEC 62366 Ed 1.0 2015-02 Medical devices -- Part 1: Application of usability ● engineering to medical devices - ISO 14971 Medical devices Application of risk management to medical devices (Ed. . 2.0, 2007-03-01) - . Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005, document number 337). - . Pediatric Information for X-ray Imaging Device Premarket Notifications - Guidance for Industry and Food and Drug Administration Staff - Code of Federal Requlations Title 21, Subchapter J Radiological Health . Bench testing included basic CT performance tests on phantoms, safety tests and software tests for functional and non-functional attributes of the proposed Philips CT Big Bore Sliding Gantry Configuration. | CT Performance Metric | Values and ranges measured on phantoms | |-----------------------|-----------------------------------------------------------------------| | MTF | Cut-off: High Mode 16±2lp/cm; Standard Mode: 13±2 lp/cm | | CTDIvol | Head: 10.61mGy/100mAs±25%; Body: 5.92mGy/100mAs±25%<br>at 120kV | | CT number accuracy | Water: 0±4HU | | Noise | 0.27% ± 0.04% at 120 kV, 250 mAs, 12 mm slice thickness, UA<br>filter | | Slice Thickness | 0.5mm - 1.5mm at nominal 0.75mm; 1.0mm - 2.0mm at nominal<br>1.5mm | {8}------------------------------------------------ The completed verification test protocols cover the main system level requirements of the System Requirements Specification and the subsystem requirement specifications as well as the identified hazard mitigations from the Safety Risk Management Report. The traceability between the requirements, the hazard mitigations and the test protocols are described in the Traceability Matrix. The Traceability Matrix also shows the overall test results per requirement and per hazard mitigation. The results of the functional and non-functional regression tests as well as the user interface verification are provided in the Traceability Matrix. The detailed results are provided in the Full System Verification Test Report. The completed validation test plan identified the tasks, deliverables, methodology, requirements and the resources for validation of the intended use and meets customer needs. The validation testing includes clinical validation and serviceability validation. The clinical validation covered requirements related to clinical workflows and features. The serviceability validation is planned to cover requirements related to installation, servicing and troubleshooting of the systems, which will be performed at system installation. The validation test plan was executed as planned and acceptance criteria met for each requirement. All deviations found were analyzed based on the defect management process and are being dispositioned per process. Conclusion: All verification and validation tests were executed according to the Philips Big Bore Sliding Gantry Configuration System Design Verification and Validation Plans, All deviations found were analyzed based on the defect management process and are being dispositioned per process. No unacceptable deviation exist. # Summary of Clinical Performance Testing The proposed Philips CT Big Bore Sliding Gantry Configuration did not require a clinical study since substantial equivalence to the legally marketed predicate device was proven with the verification/validation testing. # Conclusion: The test results demonstrate that the proposed Philips CT Big Bore Sliding Gantry Configuration meets the acceptance criteria and is adequate for its intended use. Additionally, the risk management activities show that all risks are sufficiently mitigated, that no new risks are introduced, and that the overall residual risks are acceptable. Based on the supporting data provided in this Special 510(k) submission, the proposed Philips CT Big Bore Sliding Gantry Configuration is considered substantially equivalent to the primary predicate device, Philips CT Big Bore in terms of safety and effectiveness. There are no significant differences that may raise new issues of safety or effectiveness. Bench tests and user validation have been performed to demonstrate that the proposed Philips CT Big Bore Sliding Gantry Configuration is as safe and effective as the predicate Philips CT Big Bore, without raising any new safety and/or effectiveness concerns.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...