PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM

K033326 · Philips Medical Systems, Inc. · JAK · Oct 29, 2003 · Radiology

Device Facts

Record IDK033326
Device NamePHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM
ApplicantPhilips Medical Systems, Inc.
Product CodeJAK · Radiology
Decision DateOct 29, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The "Philips Plus" is a Computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

Device Story

Philips Plus is a whole-body CT scanner; features continuously rotating X-ray tube and detector gantry; supports multi-slice acquisition up to 40 slices simultaneously. Device acquires X-ray transmission data; computer reconstructs data into cross-sectional images at various angles and planes. Includes signal analysis, display equipment, and patient support components. Used in clinical settings by trained medical professionals. Output consists of cross-sectional images for diagnostic review by physicians; aids in clinical decision-making by providing anatomical visualization.

Clinical Evidence

Bench testing only. Safety and effectiveness demonstrated through adherence to GMP practices, international standards (IEC 60601-1), and radiation safety compliance (21 CFR Subchapter J). No clinical data provided.

Technological Characteristics

Whole-body CT scanner; multi-slice capability (up to 40 slices); rotating X-ray tube and detector gantry. Complies with IEC 60601-1 (electrical/mechanical safety) and 21 CFR Subchapter J (radiation safety).

Indications for Use

Indicated for whole body computed tomography imaging for diagnostic purposes in patients requiring cross-sectional x-ray analysis.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ OCT 2 9 2003 # Philips Medical Systems Inc. 510(k) Summary # "Philips Plus" CT Scanner ### 1. Submitter Philips Medical Systems (Cleveland), Inc. 595 Miner Road Cleveland, OH 44143 (440) 483-3000 #### Contact: Robert L. Turocy Philips Medical Systems 595 Miner Road Cleveland, OH 44143 Telephone: 440 483 3528 FAX: 440 483 1116 Date of Summary: June 19, 2001 ### 2. Device Name and Classification Device Name: "Philips Plus" CT Scanner Classification Name: Computed Tomography X-Ray System The FDA has classified the Computed Tomography X-Ray System as Class II in 21 CFR 892.1750 (Product Code 90JAK) ### 3. Device Description The "Philips Plus" is a Whole Body Computed Tomography X-Ray System featuring a continuously rotating X-ray tube and detectors gantry and multi-slice capability of up to 40 slices simultaneously. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. This device also includes signal analysis and display equipment, patient and equipment supports, components and accessories. ### 4. Intended Use Of The Device The "Philips Plus" is a Whole Body Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. {1}------------------------------------------------ ## 5. Comparison to Predicate Devices In the opinion of Philips Medical Systems Inc., the "Philips Plus" CT scanner is of comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: - Mx8000 IDT, K012009 . - Mx8000 v5.0, K010817 ● ## 6. Safety and Effectiveness Considerations The safety of the device is assured by adherence to GMP practices and to International Standards. Potential hazards are identified in a hazard analysis and controlled in the following manner: Software safety is assured by the company procedures that conform to accepted practices. Quality assurance procedures were adhered to, and test results demonstrate that the option specifications and functional requirements were met. Electrical and Mechanical safety is assured by adherence to IEC 60601-1 standards. Radiation safety is assured by compliance with 21 CFR, Subchapter J performance standards. ## 7. Substantial Equivalency Statement Based on the above considerations, it is Philips's opinion that the "Philips Plus" CT scanner is substantially equivalent in safety and effectiveness to the predicate devices, Mx8000 IDT, K012009 and Mx8000 v5.0, K010817. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 2 9 2003 Philips Medical Systems % Mr. Juergen Welte Program Manager, Third Party Review Program TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470 Re: K033326 Trade/Device Name: "Philips Plus" Computed Tomography System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: October 14, 2003 Received: October 16, 2003 Dear Mr. Welte: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sir.cerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): Ka3 33 2 6 Page 1 of 1 "Philips Plus" Device Name: Indications for Use: The "Philips Plus" is a Computed Tomography X-Ray System intended to produce images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) David G. Larson fological Devices 510(k) Numbe Philips Plus 510(k) 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...