SOMATOM EMOTION 6 CT SYSTEMS

K023687 · Siemens Medical Systems, Inc. · JAK · Nov 22, 2002 · Radiology

Device Facts

Record IDK023687
Device NameSOMATOM EMOTION 6 CT SYSTEMS
ApplicantSiemens Medical Systems, Inc.
Product CodeJAK · Radiology
Decision DateNov 22, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1750
Device ClassClass 2

Intended Use

The SOMATOM Emotion 6 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

Device Story

Whole-body X-ray computed tomography scanner; utilizes continuously rotating tube-detector system; operates on fan beam principle. Inputs: X-ray transmission data acquired via axial or spiral scanning. Transformation: Computer reconstruction of transmission data into cross-sectional images. Output: Cross-sectional body images. Used in clinical settings by trained medical professionals. Healthcare providers use images for diagnostic evaluation. Benefits: Enables non-invasive visualization of internal body structures for clinical decision-making.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and adherence to established safety standards for diagnostic X-ray equipment.

Technological Characteristics

Whole-body CT scanner; fan beam geometry; continuously rotating tube-detector system. Operates with SOMARIS/5 software. Complies with 21 CFR Subchapter J, IEC 60601-1, and UL 187 safety standards.

Indications for Use

Indicated for whole-body diagnostic imaging via cross-sectional X-ray computed tomography reconstruction in patients requiring axial or spiral plane scanning.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the word "SIEMENS" in bold, black letters. The letters are large and evenly spaced. The background is white, providing a strong contrast to the text. The font appears to be a sans-serif typeface. NOV 22 2002 #### Attachment 1 ## Summary of Safety and Effectiveness This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### General Information I. | Device Name: | SOMATOM Emotion 6 Computed Tomography X-ray<br>Systems | |-------------------------|------------------------------------------------------------------------------------| | Classification Name: | § 892.1750:<br>Computed tomography X-ray system | | Proprietary Trade Name: | SOMATOM Emotion 6 | | Classification: | Class II | | Performance Standard: | 21 CFR Subchapter J,<br>Federal Diagnostic X-ray Equipment Standard | | Registration Number: | 2240869 | | Address: | Siemens Medical Systems, Inc.<br>186 Wood Avenue South<br>Iselin, N.J. 08830 | | Contact Person: | Praveen Nadkarni<br>Technical Specialist, Regulatory Submissions<br>(732) 321-4950 | Date of Summary Preparation: 10/29/02 {1}------------------------------------------------ ### II. Safety and Effectiveness Information Supporting the Substantial Equivalence Determination #### Device Description: The Siemens SOMATOM Emotion 6 systems are a whole body X-ray computed tomography scanners, which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is a command-based program used for patient management, data management, X-ray scan control, image reconstruction, and image archive/evaluation. #### Intended Use: The SOMATOM Emotion 6 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) #### Technological Characteristics: The SOMATOM Emotion 6 systems are whole body X-ray computed tomography scanners, which features a continuously rotating tube-detector system and functions according to the fan beam principle. The systems are based on the existing SOMATOM Emotion systems (for further details see chapter 2). The system will operate with SOMARIS/5 software. #### General Safety and Effectiveness Concerns: All components of the SOMATOM Emotion 6 systems subject to the Federal Diagnostic Equipment Performance Standard and applicable regulations of 21CFR § 1020.30 and § 1020.33 are certified to meet those requirements; and an initial report as per 21 CFR § 1002.10 will be filed with the Center for Devices and Radiological Health (CDRH). To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice. The SOMATOM Emotion Duo is designed to meet the ELECTRICAL AND MECHANICAL SAFETY STANDARD IEC 60601-1 and UL 187 X-RAY EQUIPMENT STANDARD FOR SAFETY. #### Substantial Equivalence: Siemens believes that within the meaning of the Safe Medical Device Act of 1990, the SOMATOM Emotion 6 systems operating with SOMARIS/5 software are substantially equivalent to the Siemens SOMATOM Emotion Duo and Somatom Sensation 4 CT scanners in commercial distribution. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract symbol that resembles three stylized human profiles or birds in flight, stacked on top of each other. ood and Drug Administration 0 Corporate Boulevard Mr. Praveen Nadkarni Technical Specialist, Regulatory Submissions SIEMENS Medical Systems, Inc. 186 Wood Avenue South ISELIN NI 08830 Re: K023687 Trade/Device Name: SOMATOM Emotion 6 CT System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: October 29, 2002 Received: November 1, 2002 Dear Mr. Nadkarni: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|-----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692) | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Snogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SIEMEN #### Attachment 2 Indication for use 510(k) Number (if known): K023687 Device Name: SOMATOM Emotion 6 systems Indication for use: The SOMATOM Emotion 6 systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.) *Prescription Use* Daniel A. Layne Attachment 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...