SYNGO DUAL ENERGY SOFTWARE PACKAGE

K132902 · Siemens Medical Solutions USA, Inc. · JAK · Apr 7, 2014 · Radiology

Device Facts

Record IDK132902
Device NameSYNGO DUAL ENERGY SOFTWARE PACKAGE
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeJAK · Radiology
Decision DateApr 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2

Intended Use

syngo® Dual Energy is designed to operate with Siemens Dual Source CT scanners. CT images taken at the same time, using two different kV levels, of the same anatomical region of a patient are used. Depending on the region of interest, contrast agents may be used. The differences in the energy dependence of the attenuation coefficient of the different materials provide information about the chemical composition of body materials. The images are combined to visualize and analyze information about anatomical and pathological structures. The functionality of the syngo® Dual Energy applications are as follows: Lung Nodules; Xenon (Lung Ventilation); Monoenergetic; Brain Haemorrhage; Gout Evaluation; Lung Vessels; Heart PBV; Dual Energy Bone Removal; Cartilage, Tendon, Ligament; Lung Perfusion; Liver; Hard Plaques in Vessels; Kidney Stones. Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis under consideration of all available information. The accuracy of identification is decreased in obese patients.

Device Story

Post-processing software package for Siemens Dual Source CT scanners; inputs two series of CT images acquired at different kV levels (e.g., 80/140kV or 100/140kV). Performs material decomposition based on energy-dependent attenuation coefficients; outputs color-coded overlays on orthogonal anatomical slices (sagittal, coronal, axial). Used by radiologists in clinical settings to visualize tissue/material composition (e.g., kidney stones, gout, lung perfusion). Enables interactive adjustment of window values, color look-up tables, and masking. Assists clinicians in differentiating stone types; final diagnosis requires integration with patient history and lab results.

Clinical Evidence

Bench testing only. Phantom-based testing conducted using kidney stones of varying chemical compositions and sizes within cylindrical phantoms of varying diameters. Classification performance measured against CTDIval; performance limitations for small or multicomponent stones identified.

Technological Characteristics

Post-processing software embedded in workstation framework. Utilizes dual-energy material decomposition algorithms. DICOM compliant. Standards: IEC 60601-1-6 (usability), IEC 62304 (software lifecycle), ISO 14971 (risk management).

Indications for Use

Indicated for patients undergoing dual-source CT imaging for visualization and analysis of anatomical and pathological structures, including chemical composition of kidney stones (uric acid vs. non-uric acid). Contraindicated/limited in obese patients due to decreased accuracy.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SIEMENS Traditional 510(k) Submission: syngo® Dual Energy Software Package N32902 Page 1 of 5 ### 510(K) Summary APR 0 7 2014 ## FOR ## syngo® Dual Energy Software Package Submitted by: Siemens Medical Solutions USA. Inc. 51 Valley Stream Parkway Malvern, PA 19355 Date Prepared: August 9, 2013 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. #### 1. General Information lmporter/Distributor: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Establishment Registration Number: 2240869 Manufacturing Site: Siemens AG Medical Solutions Henkestrasse 127 D-91052 Erlangen, Germany Establishment Registration Number: 3002808157 #### 2. Contact Person: Mrs. Kimberly Manqum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Vallev Stream Parkway D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly.mangum@siemens.com {1}------------------------------------------------ SIEMENS Traditional 510(k) Submission: syngo® Dual Energy Software Package | 3. | Device Name and Classification | | | |----|----------------------------------------------|------------------------------------------------|--| | | Product Name: | syngo® Dual Energy Software Package | | | | Propriety Trade Name: | syngo® Dual Energy Software Package | | | | Classification Name: | Computed Tomography X-ray System | | | | Classification Panel: | Radiology | | | | CFR Section: | 21 CFR §892.1750 | | | | Device Class: | Class II | | | | Product Code: | 90JAK | | | | Legally Marketed Predicate Devices | | | | | Trade Name: | syngo® Dual Energy | | | | 510(k)#: | K062351 | | | | Clearance Date: | October 5, 2006 | | | | Classification Name: | System, X-Ray, Tomography, Computed | | | | Classification Panel: | Radiology | | | | Classification Regulation:21 CFR & 892.1750 | | | | | Device Class: | | | | | Product Code: | JAK | | | | Trade Name: | syngo® Dual Energy with extended functionality | | | | 510(k)#: | K083524 | | | | Clearance Date: | April 01, 2009 | | | | Classification Name: | System, X-Ray, Tomography, Computed | | | | Classification Panel: | Radiology | | | | Classification Regulation: 21 CFR & 892.1750 | | | | | Device Class: | | | | | Product Code: | JAK | | | | | | | #### 4. Substantial Equivalence: The subject device Kidney Stones application class included as part of Siemens syngo® Dual Energy software package is substantially equivalent to the Kidney Stone application classes included as part of the following medical devices in commercial distribution: | Predicate Device Name | FDA Clearance Number | FDA Clearance Date | |----------------------------------------|----------------------|--------------------| | syngo® Dual Energy Software<br>Package | K083524 | April 1, 2009 | | syngo® Dual Energy | K062351 | October 5, 2006 | #### ട്. Device Description:. syngo® Dual Energy Software Package is a post processing application package consisting of the following applications that can be used to improve visualization of various energy dependent materials in the human body: {2}------------------------------------------------ # SIEMENS Traditional 510(k) Submission: syngo® Dual Energy Software Package - . Lung Nodules - Lung Nodules Xenon (Lung Ventilation) . - Monoenergetic . - Brain Haemorrhage . - Gout Evaluation . - Luna Vessels . - Heart PBV . - . Dual Energy Bone Removal - Cartilage, Tendon, Ligament . - Lung Perfusion . - Liver ● - Hard Plaques in Vessels . - Kidney Stones . The output of the syngo® Dual Energy Software Package applications are shown within three orthogonal slices (sagittal, coronal, and axial) as overlay to anatomical grayscale information. Typically the anatomical information is built from a combination of the original dual energy scans (for example 80kV and 140kV or 100 kV and 140kV and/or by using different filters). Window values, color look-up tables (LUTs) and masking operations for both datasets (anatomical and the material decomposition information) can be interactively defined and user modified. An additional side-by-side display of Dual Energy results is also supported, and navigation and rotation of datasets are provided. After two series of dual energy images have been loaded from the patient browser (80kV and 140kV or 100 kV and 140kV), three orthogonal slices (sagittal, coronal, and axial) of both spectra are shown within the respective segments on the monitor in grayscale mode. The middle axial, sagittal and coronal view is used for the initial display. The Kidney Stones application class is a post processing application that uses dual energy information to visualize the chemical composition in kidney stones. The basis for this approach is a material decomposition into tissue, uric acid, and oxalate stones. The result of material decomposition is shown as a color overlay to anatomical gray scale information. Summary of Technical Characteristics of the Subject Device as 6. Compared with the Predicate Device: The Kidney Stones application class is a post processing application included as part of syngo® Dual Energy Software Package. syngo® Dual Energy Software Package is a post processing application package embedded in the frame-work of a single user HW/SW architecture based on advanced workstations. The predicate device Kidney Stones application class is a post processing application which was cleared as part of syngo® Dual Energy Software (K062351, clearance date October 5, 2006) and syngo® Dual Energy with extended functionality (K083524, clearance date April 1, 2009). The subject device Kidney Stones application class provides similar evaluation, {3}------------------------------------------------ Traditional 510(k) Submission: syngo Dual Energy Software Package reporting and visualization tools, and functionality as the predicate devices Kidney Stones application classes. This includes image processing and visualization tools such as basic visualization of various energy dependent materials in the human body and VRT visualization. The subject device Kidney Stones application class does not have significant changes in technological characteristics when compared to the predicate devices Kidney Stones application class. The Indication for Use, operating principle, and the scientific technology are similar; therefore, Siemens believes that Kidney Stones application class marketed as part of the syngo® Dual Energy Software Package is substantially equivalent to the predicate devices. No other post processing applications cleared as part of the predicate devices have been modified with this submission. #### 7. Nonclinical Testing: SIEMENS syngo® Dual Energy Software Package is designed to fulfill the requirements of following standards: - IEC 60601-1-6 : 2007; Medical electrical equipment Part 1-6; General . requirements for basic safety and essential performance - Collateral Standard: Usability - . IEC 62304 Ed. 1.0, "Medical Device Software - Software Lifecycle Processes" - t ISO 14971:2007; Medical devices - Application of risk management to medical devices - DICOM (Digital Imaging and Communications in Medicine) Standard: . 2009 DICOM conformity is fully covered by syngo.via implementations. Non clinical tests were conducted for the Kidney Stones Application class during product development. Phantom based test were conducted using kidney stones of various chemical composition and sizes. The stones were put into cylindrical phantoms with varying diameters and scanned. Classification performance was measured as a function of CTDIval, and performance limitations for small or multicomponent stones are a result of the analysis. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence. Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for {4}------------------------------------------------ K132902 Page 5 of 5 # SIEMENS Traditional 510(k) Submission: syngo Dual Energy Software Package Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. ### 8. Indications for Use: syngo® Dual Energy is designed to operate with Siemens Dual Source CT scanners. CT images taken at the same time, using two different kV levels, of the same anatomical region of a patient are used. Depending on the region of interest, contrast agents may be used. The differences in the energy dependence of the attenuation coefficient of the different materials provide information about the chemical composition of body materials. The images are combined to visualize and analyze information about anatomical and pathological structures. The functionality of the syngo® Dual Energy applications are as follows: - Lung Nodules - Xenon (Lung Ventilation) - Monoenergetic . - Brain Haemorrhage ● - Gout Evaluation . - Lung Vessels - Heart PBV - Dual Energy Bone Removal - Cartilage, Tendon, Ligament - Lung Perfusion ● - Liver ● - Hard Plaques in Vessels . - Kidney Stones Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis under consideration of all available information. The accuracy of identification is decreased in obese patients. #### 9. General Safety and Effectiveness Concerns: The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development. verification and validation testing. To minimize electrical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards. #### 10. Conclusion as to Substantial Equivalence In summary. Siemens is of the opinion that the syngo® Dual Energy Software Package does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices. {5}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three overlapping, curved lines that resemble a person with outstretched arms. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 7, 2014 Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Technical Specialist 51 Valley Stream Parkway MALVERN PA 19355 Re: K132902 Trade/Device Name: syngo® Dual Energy Software Package Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 5, 2014 Received: March 7, 2014 Dear Ms. Mangum: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class 11 (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below, 510(k) Number (if known) K132902 ### Device Name syngow Dual Energy ### Indications for Use (Describe) syngo® Dual Energy is designed to operate with Siemens Dual Source CT images taken at the same time, using two different kV levels, of the same anatomical region of a patient are used. Depending on the region of interest, contrast agents may be used. The differences in the energy dependence of the attenuation coefficient of the different materials provide information about the chemical composition of body materials. The images are contined to visualize and analyze information about anatomical and pathological structures. The functionality of the syngo® Dual Energy applications are as follows: - · Lung Nodules - · Dual Energy Bone Removal - · Monoenergetic · Cartilage, Tendon, Ligament · Lung Perfusion - . Brain Haemorrhage - · Gout Evaluation • Liver - · Hard Plaques in Vessels - · Lung Vessels • Heart PBV - · Kidney Stones Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid stones. For full identification of the kidney stone additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis under consideration of all available information. The accuracy of identification is decreased in obese patients. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) J Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {8}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *An agency may not conduct or sponsor, and a person is not required to respond to. a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...