Precise Image
K210760 · Philips Medical Systems Nederland B.V. · JAK · Jan 14, 2022 · Radiology
Device Facts
| Record ID | K210760 |
| Device Name | Precise Image |
| Applicant | Philips Medical Systems Nederland B.V. |
| Product Code | JAK · Radiology |
| Decision Date | Jan 14, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1750 |
| Device Class | Class 2 |
| Attributes | AI/ML |
Intended Use
The Precise Image is a reconstruction software application for a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipment supports, components and accessories. Precise Image has been evaluated and available on preselected reference protocols for adult subjects. Precise Image is not indicated for use in pediatric subjects. The CT system with Precise Image is indicated for head, whole body and vascular X-ray Computed Tomography applications. These scanners are intended to be used for diagnostic imaging. Precise Image uses an Artificial Intelligence powered reconstruction algorithm that is designed for low radiation dose, provides low noise, and improves low contrast detectability.
Device Story
Precise Image is a reconstruction software application for Philips whole-body CT systems. It processes raw x-ray transmission data acquired at various angles and planes to generate diagnostic images. The device utilizes an AI-powered reconstruction algorithm based on Convolutional Neural Networks (CNN) to reduce radiation dose, lower image noise, and improve low-contrast detectability while maintaining an image appearance similar to traditional Filtered Back Projection (FBP). The software includes five user-adjustable settings allowing radiologists to balance dose reduction and image quality preferences. It is intended for use by clinicians in diagnostic imaging environments. By enabling lower radiation doses without compromising image quality, the device aims to benefit patients through reduced exposure while providing radiologists with high-quality images for diagnostic decision-making.
Clinical Evidence
No external clinical study was required. A comparative image evaluation study was conducted on 55 image set pairs from the subject device and predicate devices (Incisive CT and Brilliance iCT). Six board-certified radiologists evaluated images on a five-point Likert scale for diagnostic confidence, sharpness, noise level, image texture, and artifacts. All images were rated as having good image quality.
Technological Characteristics
Software-based reconstruction application utilizing Artificial Intelligence and Convolutional Neural Networks (CNN). Operates on Philips CT systems. Features five user-adjustable settings for dose/quality optimization. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-44, IEC 62304, ISO 10993-1, and ISO 14971.
Indications for Use
Indicated for head, whole body, and vascular X-ray Computed Tomography diagnostic imaging in adult subjects. Not indicated for pediatric subjects.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Reference Devices
Related Devices
- K230807 — Deep Learning Image Reconstruction · Ge Healthcare Japan Corporation · Apr 20, 2023
- K193073 — Deep Recon · Shanghai United Imaging Healthcare Co., Ltd. · Jul 6, 2020
- K213999 — Deep Learning Image Reconstruction · Ge Medical Systems, LLC · Feb 18, 2022
- K212067 — Deep Learning Image Reconstruction · Ge Healthcare Japan Corporation · Sep 17, 2021
- K183460 — ClariCT.AI · Claripi, Inc. · Jun 13, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
January 14, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Medical Systems Nederland, B.V. % Michael Chilbert Regulatory Affairs Engineer Veenpluis 4-6 Best Noord-Brabant, 5684 PC NETHERLANDS
Re: K210760
Trade/Device Name: Precise Image Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: December 10, 2021 Received: December 13, 2021
Dear Mr. Michael Chilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210760
Device Name Precise Image
#### Indications for Use (Describe)
The Precise Image is a reconstruction software application for a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment supports, components and accessories. Precise Image has been evaluated and available on preselected reference protocols for adult subjects. Precise Image is not indicated for use in pediatric subjects.
The CT system with Precise Image is indicated for head, whole body and vascular X-ray Computed Tomography applications. These scanners are intended to be used for diagnostic imaging.
Precise Image uses an Artificial Intelligence powered reconstruction that is designed for low radiation dose, provides lower noise, and improves low contrast detectability.
Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 5 510(K) SUMMARY
HA300-006-05 Revision: 01 Status: Approved Page: 1 of 5
## CT/AMI
## SECTION 5 510(K) SUMMARY
# 510(k) Summary for K210760
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared:<br>Manufacturer: | January 13, 2022<br>Philips Medical Systems Nederland B.V.<br>Veenpluis 6,<br>5684 PC BEST<br>The Netherlands<br>Establishment Registration Number: 3015777306 | |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Primary Contact<br>Person: | Michael Chilbert, Ph.D., P.E.<br>Regulatory Affairs Engineer<br>Phone: 440-869-4576<br>E-mail: Michael.chilbert@philips.com | |
| Secondary<br>Contact Person: | Douglas Kentz<br>Director of Regulatory Affairs<br>Phone: (262) 389-7369<br>E-mail: douglas.kentz@philips.com | |
| Device: | Trade Name: | Precise Image |
| | Common name: | Computed Tomography x-Ray System |
| | Classification Name: | Computed Tomography x-Ray System |
| | Classification Regulation: | 21CFR 892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | II |
| | Primary Product Code: | JAK |
| | Secondary Product Code: | Not Applicable |
| Predicate<br>Device: | Trade Name: | Incisive CT |
| | Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd. |
| | 510(k) Clearance: | K180015 |
| | Classification Name: | Computed Tomography X-Ray System |
| | Classification Regulation: | 21CFR §892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | JAK |
| Reference<br>Device: | Trade Name: | Brilliance iCT |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K162838 |
| | Classification Name: | Computed Tomography X-Ray System |
| | Classification Regulation: | 21CFR §892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | JAK |
{4}------------------------------------------------
| PHILIPS<br>BG Diagnostic Imaging | SECTION 5<br>510(K) SUMMARY | HA300-006-05<br>Revision: 01<br>Status: Approved<br>Page: 2 of 5 |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Device<br>Description: | The proposed Precise Image is a reconstruction software application that may be used<br>on a Philips whole-body computed tomography (CT) X-Ray System. Precise Image is a<br>robust reconstruction software application, utilizing technological advancements in<br>Artificial Intelligence and a Convolutional Neural Networks (CNN), When used, Precise<br>Image generates CT images that provides an image appearance similar to traditional<br>FBP images while reducing dose and improving image quality.<br><br>The implemented algorithm includes 5 user-adjustable settings to match the Radiologist's<br>preference for dose reduction and image quality. | |
| | The proposed Precise Image reconstruction has been trained on and may be used on the<br>currently marketed predicate device Philips Incisive CT System (K180015). | |
| Indications for<br>Use / Intended<br>Use: | The Precise Image is a reconstruction software application for a Computed Tomography<br>X-Ray System intended to produce images of the head and body by computer<br>reconstruction of x-ray transmission data taken at different angles and planes. These<br>devices may include signal analysis and display equipment, patient and equipment<br>supports, components and accessories. Precise Image has been evaluated and available<br>on preselected reference protocols for adult subjects. Precise Image is not indicated for<br>use in pediatric subjects. | |
| | The CT system with Precise Image is indicated for head, whole body and vascular X-ray<br>Computed Tomography applications. These scanners are intended to be used for<br>diagnostic imaging. | |
| | Precise Image uses an Artificial Intelligence powered reconstruction algorithm that is<br>designed for low radiation dose, provides low noise, and improves low contrast<br>detectability. | |
| Technological<br>Characteristics | The proposed Precise Image reconstruction software application, has the same<br>fundamental design characteristics and are based on comparable technologies found in<br>the currently marketed predicate device Philips Incisive CT (K180015) | |
| | This 510(k) submission addresses the following: | |
| | Precise Image is an Artificial Intelligence powered reconstruction algorithm that is<br>designed for low radiation dose, provides low noise, and improves low contrast<br>detectability as compared to iDose, as well as improve image appearance that<br>more closely resembles Filtered Back Projection (FBP) at higher doses.The currently marketed predicate Philips Incisive CT (K180015) is being used for<br>diagnostic imaging in radiology, interventional radiology and cardiology and in<br>oncology as part of treatment preparation and radiation therapy planning. The<br>fundamental scientific technology and design of the proposed Precise Image<br>reconstruction software application is equivalent to the currently marketed<br>predicate device Philips Incisive CT (K180015). | |
{5}------------------------------------------------
### PHILIPS BG Diagnostic Imaging
## SECTION 5 510(K) SUMMARY
HA300-006-05 Revision: 01 Status: Approved Page: 3 of 5
## CT/AMI
| Technological Characteristics Comparison<br>Design and Fundamental Scientific Technology | | | |
|------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Device | Predicate Device:<br>Incisive CT System<br>(K180015) | Proposed Device:<br>Precise Image software<br>application | Conclusion |
| Application | Head, Body, Vascular<br>and Cardiac | Head, Body and Vascular | Proposed device has scan<br>types found in the<br>predicate, |
| Scan Regime | Continuous Rotation | Continuous Rotation | Scan data used for<br>reconstruction - Identical, |
| Scan Field of<br>View | Up to 500 mm | Up to 500 mm | Identical |
| Scan modes | Surview<br>Axial-after-Axial<br>Dynamic Scan<br>Helical Scan | Axial-after-Axial<br>Helical Scan | Proposed device has scan<br>types found in the<br>predicate |
| Low Contrast<br>Resolution (20 cm<br>Catphan<br>phantom) | 4 mm @ 0.3% @ 22 mGy<br>CTDIvol | 5 mm @ 0.3% @ 5.5<br>mGy CTDIvol | Improved, better low<br>contrast resolution at lower<br>dose levels. |
| Minimum Scan<br>Time | 0.35 sec for 360° rotation | 0.35 sec for 360° rotation | Identical |
| Noise in Standard<br>Mode (as<br>measured on<br>21.6 cm water-<br>equivalent) | 0.27% at 27 mGY | 0.27% at 27 mGY | Identical |
| Noise Reduction<br>and Low Contrast<br>Detectability | N/A - Standard mode<br>(baseline for claim) | achieving up to 85% lower<br>noise at 80% lower dose<br>and 60% better low<br>contrast detectability | Improved, better noise<br>reduction and low contrast<br>detectability |
| Noise Power<br>Spectrum | N/A - Standard mode<br>(baseline for claim) | Where noise is reduced<br>by at least 50%, the<br>system shall shift the<br>noise power spectrum of<br>images by no more than<br>6% as compared to the<br>same data reconstructed<br>without Precise Image | Will not shift NPS more<br>than 6% |
| Application | Head, Body, Vascular<br>and Cardiac | Head, Body and Vascular | Proposed device has scan<br>types found in the<br>predicate. |
| Scan Regime | Continuous Rotation | Continuous Rotation | Scan data used for<br>reconstruction - Identical. |
Based on the information provided above, the proposed Precise Image reconstruction software application for Philips CT Systems is considered substantially equivalent to the currently marketed predicate device Philips Incisive CT (K180015), in terms of fundamental scientific technology.
{6}------------------------------------------------
CT/AMI
## SECTION 5 510(K) SUMMARY
| Summary of<br>Non-Clinical<br>Performance<br>Data: | Non-clinical performance testing has been performed on the proposed Precise Image<br>reconstruction software application for CT Systems.<br><br>Philips CT Systems, comply with the following International and FDA recognized<br>consensus standards and FDA guidance document(s).<br><br>IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012<br>(or IEC 60601-1: 2012 reprint) (FDA Recognition Number: 19-4) IEC 60601-1-2:2014: (FDA Recognition Number :19-8) IEC 60601-1-3:2008+A1:2013 (FDA Recognition Number: 12-269) IEC 60601-1-6:2010 +A1: 2013 (FDA Recognition Number: 5-89) IEC 60601-2-44:2009/AMD2:2016 (FDA Recognition Number: 12-302) IEC 62304:2006 + A1: 2015 (FDA Recognition Number: 13-79) ISO 10993-1:2009/Cor.1:2010 (FDA Recognition Number: 2-220) ISO 14971 2nd Edition. (FDA Recognition Number: 5-40) |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device Specific Guidance Document:<br><br>Guidance for Industry and FDA Staff – Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices" (issued May 11, 2005) International and FDA-recognized consensus IEC 62304 "Medical device software<br>- Software life cycle processes" (Edition 1.1, 2006). FDA/CDRH recognition<br>number 13-79. Proposed Regulatory Framework for Modifications to Artificial<br>Intelligence/Machine Learning (AI/ML)-Based Software as a Medical Device<br>(SaMD) Discussion Paper. (April 2, 2019) |
| | Design Verification planning and testing was conducted at the sub-system and at the<br>system level. The sub-systems are tested against the sub-system requirements<br>specifications (SSRS) and the system level verification is conducted against the system<br>requirement specifications (SRS). System and sub-system verification activities<br>demonstrate the system or sub-systems meet the established system and sub-system<br>level design input requirements. System and sub-system level requirements may be<br>verified by manual test, automated test, inspection/analysis, or any combination of the<br>three. Design verification also includes Image Quality verification and risk analysis/risk<br>mitigation testing. |
| | The traceability between the requirements, the hazard mitigations and the test protocols<br>are described in the Traceability Matrix. The Traceability Matrix also shows the overall test<br>results per requirement and per hazard mitigation. The summary and conclusion of<br>results are provided in the System Verification Test Report. |
| | Non-Clinical design validation testing covered the intended use and commercial claims<br>Validation testing included clinical image evaluation and workflow validation. |
| | All these tests were used to support substantial equivalence of the proposed Precise<br>Image reconstruction software application and demonstrate that the proposed Precise<br>Image reconstruction software application for CT Systems. |
- ç Maintains compliance with the aforementioned international and FDA-recognized consensus standards and/or FDA device specific guidance document, and;
- . Meets the acceptance criteria and is adequate for its intended use.
{7}------------------------------------------------
# SECTION 5 510(K) SUMMARY
Therefore, the proposed Precise Image reconstruction software application for Philips CT Systems is considered substantially equivalent to the currently marketed predicate device Philips Incisive CT (K180015) in terms of safety and effectiveness.
### Summary of Clinical Performance Data:
The proposed Precise Image reconstruction software application for CT Systems did not require any external clinical study. A comparative image evaluation study was performed on 55 image set pairs from Precise Image and the predicate Incisive CT and Brilliance iCT by 6 board certified radiologists to evaluate Diagnostic Confidence. Sharpness, Noise level. lmage texture and Artifacts on a five point Likert scale. The substantial equivalence to the currently marketed predicate device Philips Incisive CT (K180015) was demonstrated with the following attributes:
- Indication for use; .
- Technological characteristics; ●
- Non-clinical and clinical performance testing; and .
- . Safety and effectiveness.
Sample clinical images are provided with this submission, which were reviewed and evaluated by certified radiologists. All images were evaluated to have good image quality.
Substantial The proposed Precise Image reconstruction software application for Philips CT Systems is substantially equivalent to the currently marketed predicate device Philips Incisive CT Equivalence (K180015) in terms of design. features and fundamental scientific technology. The Precise Conclusion: Image reconstruction software application for Philips CT Systems does not introduce any new risk nor impact safety and effectiveness.
> Additionally, substantial equivalence was demonstrated by non-clinical and clinical (verification and validation) performance tests provided in this 510(k) premarket notification. These tests demonstrate that the proposed Precise Image reconstruction software application for CT Systems complies with the requirements specified by Philips Medical Systems Nederland B.V. and the international and FDA-recognized consensus standards and is as safe and effective as its currently marketed predicate device Philips Incisive CT (K180015) without raising any new safety and/or effectiveness concerns.
CT/AMI