The Medtronic Vascular Input® PS Introducer is a percutaneous introducer used to facilitate placing a catheter through the skin into a vein or artery. Percutaneous introducers are recommended for initial percutaneous introduction or the exchange of intravascular devices.
Device Story
Input® PS Introducer is a percutaneous catheter introducer set consisting of a sheath, locking dilator, needle, guidewire, and optional obturator. Used by clinicians to facilitate vascular access; needle punctures vessel; guidewire provides platform for sheath/dilator advancement. Sheath allows placement of various intravascular devices; hydrophilic coating on shaft reduces insertion trauma. Obturator maintains sheath patency during delays or catheter removal. Device is used in clinical settings for initial percutaneous introduction or exchange of intravascular devices.
Clinical Evidence
No clinical investigation performed. Substantial equivalence supported by bench testing and biocompatibility evaluation.
Technological Characteristics
Components: sheath, locking dilator, needle, guidewire, obturator. Sheath features hydrophilic coating, hub, strain relief, hemostatic valve, and sidearm tubing with stop-cock. Available in 11cm and 23cm lengths; 5F-9F diameters. Guidewire: 0.038" max outer diameter, 15cm length. Biocompatibility per ISO 10993-1:2009.
Indications for Use
Indicated for patients requiring percutaneous access to a vein or artery for the introduction or exchange of intravascular devices.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
K961745 — BRAUN PERCUTANEOUS INTRODUCER SET · B.Braun Medical, Inc. · Sep 17, 1996
K172331 — Type I, Type II, Type III, Type IV · Lepu Medical Technology (Beijing) Co., Ltd. · Mar 2, 2018
K971165 — INTRODUCER SHEATH AND DILATOR · Boston Scientific Corp · Jul 23, 1997
K222184 — Introducer Sheath Kits · Beijing Demax Medical Technology Co.,Ltd · Feb 21, 2023
K140768 — TYPE I BRILLIANT INTRODUCER KIT, TYPE II BRILLIANT INTRODUCER KIT, TYPE III BRILLIANT INTRODUCER KIT, TYPE IV BRILLIANT · Lepu Medical Technology (Bejing) Co., Ltd. · Nov 25, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circle with three figures inside, one standing upright and two in motion around it. Below the circle, the word "Medtronic" is written in a bold, sans-serif font.
. !
:
## Traditional 510(k) Summary
| Submitter: | Medtronic Vascular<br>37A Cherry Hill Drive<br>Danvers, MA 01923,<br>USA |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Nisarg Shah<br>Regulatory Affairs Specialist<br>37A Cherry Hill Drive<br>Danvers, MA 01923<br>USA<br>Phone: (978)-739-6632<br>Fax: (978)-750-8204<br>Email: nisarg.g.shah@medtronic.com |
| Date Prepared: | April 23, 2014 |
| Trade Name: | Input® PS Introducer |
| Common Name: | Catheter Introducer |
| Classification<br>Name: | Catheter introducer<br>Class II per 21 CFR §870.1340, Product Code DYB. |
| Predicate Device: | USCI Input™ Introducer [K940092 clearance received on<br>August 10th, 1995]; referred to as Input PS Introducer in this<br>premarket notification for identification purposes. |
| Device<br>Description: | The Medtronic Vascular Input® PS Introducer is comprised of<br>a sheath, a locking dilator, a Needle and a guide wire. This set<br>may also contain an introducer obturator which must be left in<br>dwelling post PTCA. Individual components within the<br>Input® PS Introducer set are described below:<br><br>1. Input PS Dilator: The function of a dilator is to dilate<br>the artery or vein to provide a less traumatic entry of<br>the introducer sheath to a vasculature. The dilator is<br>usually placed inside the lumen of the introducer<br>sheath during sheath insertion. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular emblem above the word "Medtronic". The emblem depicts a stylized human figure in motion, possibly representing health or activity. The word "Medtronic" is written in a bold, sans-serif font.
T
K132617 510(k) Summary Page 2 of 4
- 2. GuideWire: The guidewire is inserted through the needle into the vessel and provides a platform over which the dilator and introducer sheath can be advanced. The guidewire is available with a maximum outer diameter of 0.038" Double Distal "J" guidewire approximately 15cm in length.
3. Input PS Introducer Sheath: The Input PS Introducer sheath allows for placement of various devices into the vasculature. The sheath shaft is coated with hydrophilic coating to improve ease of insertion. The Input Introducer sheath is available in two different lengths: 11cm sheath and 23cm sheath. The l 1 cm sheath is available in five different diameters measured in French size (5F, 6F, 7F, 8F and 9F). The 23cm sheath is available in four different diameters measured in French size (6F, 7F, 8F and 9F).
- 4. Needle: The needle provided with the Percutaneous Catheter Introducer Set is a standard puncture needle. The needle is used to puncture the vessel to provide route into the vessel for the guidewire.
- 5. Obturator: The Medtronic Input Introducer Obturator is inserted into an introducer sheath to provide support and help maintain sheath patency when placement of an intravascular catheter is delayed or removed.
Statement of The Medtronic Vascular Input PS Introducer is a percutaneous Intended Use: introducer used to facilitate placing a catheter through the skin into a vein or artery. Percutaneous introducers are recommended for initial percutaneous introduction or the exchange of intravascular devices.
Summarv of The Medtronic Vascular Input PS Introducer sheath allows for Technological placement of various devices into the vasculature. The Input Characteristics: PS Introducer sheath assembly involves following feature:
- i. Shaft (with hydrophilic coating)
- ii. Hub
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular emblem above the word "Medtronic" in a bold, sans-serif font. The emblem features a stylized representation of a human figure in motion, possibly depicting a person with an implantable medical device. The overall design is clean and professional, reflecting the company's focus on medical technology and innovation.
7
K132617 510(k) Summary Page 3 of 4
- iii. Strain relief
- IV. Hemostatic valve
- v. Sidearm tubing with a stop-cock.
Summary of Non-The device performance qualification and the biocompatibility i clinical Data: testing for the material modification was conducted in accordance with the relevant recommendations from the relevant FDA guidance to demonstrate that the modified Input PS Introducer sheath have met the acceptance criteria and performed similar to the predicate device:
> Performance Qualification/ Bench Testing: Bench testing was performed specific to the material modification to the Input PS Introducer sheath. The following verification tests were performed to demonstrate the substantial equivalence of the modified Input PS Introducer Sheath to the predicate device:
- i. Visual Inspections
- ii. Coating length
- iii. Lubricity
- iv. Coating Durability
- v. Introducer Insertability
- vi. Particulate evaluation
Biocompatibility Testing: Pursuant to the ISO 10993-1: 2009- Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process; the biocompatibility testing for the modified Input PS Introducers was completed.
No new safety or effectiveness concerns were raised during the testing on the modified device. The performance qualification testing along with biocompatibility testing demonstrated that the modified Input PS Introducer sheath is safe, effective and substantially equivalent to the predicate device.
### Summary of Clinical Data:
No clinical investigation has been performed on the modified device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo consists of an image of a person standing upright, with another person lying horizontally across their shoulders, inside of a circle. Below the image is the word "Medtronic" in a bold, sans-serif font.
K132617 510(k) Summary Page 4 of 4
Conclusion from Data:
Medtronic Vascular has demonstrated that the modified Input PS Introducers are substantially equivalent to the predicate device.
・
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. The logo is black and white and appears to be a scanned image.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002
April 30, 2014
Medtronic, Inc. % Nisarg Shah Regulatory Affairs Specialist 37A Cherry Hill Drive Danvers, Massachusetts 01923
Re: K132617
> Trade/Device Name: Medtronic Input® PS Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Percutaneous Catheter Introducer Regulatory Class: II Product Code: DYB Dated: February 27, 2014 Received: March 4, 2014
Dear Mr. Shah,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{5}------------------------------------------------
Page 2 - Nisarg Shah
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K132617
Device Name: Medtronic Vascular Input® PS Introducer
Indications for Use:
A percutaneous introducer is used to facilitate placing a catheter through the skin into a vein or artery. Percutaneous introducers are recommended for initial percutaneous introduction or the exchange of intravascular devices.
Prescription Use M (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
PAGE 1 OF
Bram D. Zuckerman -S 2014.04.30 05:50:28 -04'00'
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.