K221655 · Vizaramed, Inc. · DYB · Nov 14, 2022 · Cardiovascular
Device Facts
Record ID
K221655
Device Name
Multiflex Steerable Sheath
Applicant
Vizaramed, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Nov 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Intended Use
The VizaraMed Multiflex Steerable Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
Device Story
Multiflex Steerable Sheath is a sterile, single-use, percutaneous introducer; facilitates placement of cardiovascular catheters into the heart. Device features steerable shaft capable of 3D shapes; handle with lockable steering controls; Tuohy-Borst adapter with hemostasis valve; side port; atraumatic soft distal tip. Shaft is radiopaque for fluoroscopic visualization. Used in clinical settings by physicians; provides access path for catheters. Benefits include controlled navigation to target cardiac sites. Device is mechanical; no software or algorithms involved.
Clinical Evidence
No clinical data. Evidence consists of bench testing including dimensional verification, tensile testing, leak testing, and biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, hemocompatibility). Usability assessment conducted via physician questionnaire in a simulated heart model.
Technological Characteristics
Deflectable, percutaneous steerable sheath with dilator. Available in 12.5F and 15.5F sizes. Features lockable steering handle, Tuohy-Borst adapter, hemostasis valve, and radiopaque shaft. Sterilization via ethylene oxide (EO) per ISO 11135. Limited contact (≤24 hrs) externally communicating device.
Indications for Use
Indicated for patients requiring cardiovascular catheter introduction into the heart, including access to the left side of the heart via the interatrial septum.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
St. Jude Medical Agilis Steerable Introducer (K081645)
Reference Devices
Cook, Inc. Extra Large Check-Flow Introducer (K203670)
Related Devices
K260626 — Multiflex Steerable Sheath 17.5F · Vizaramed, Inc. · Mar 26, 2026
K211100 — AcQGuide MAX Steerable Sheath · Acutus Medical, Inc. · May 14, 2021
K233248 — FARADRIVE Steerable Sheath · Boston Scientific Corporation · Dec 11, 2023
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2022
VizaraMed, Inc. Jack P Douglas, PhD Vice President, Regulatory Affairs 1914 O'Toole Way San Jose, California 95131
Re: K221655
Trade/Device Name: Multiflex Steerable Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB Dated: June 3, 2022 Received: June 7, 2022
Dear Dr. Jack Douglas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
S
Ankurita Datta - Digitally signed by Ankurita Datta -S Date: 2022.11.14 14:35:55 -05'00'
for Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221655
Device Name Multiflex Steerable Sheath
Indications for Use (Describe)
The VizaraMed Multiflex Steerable Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 1. Date prepared | June 1, 2022 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Manufacturer | VizaraMed, Inc. |
| 3. Contact | Jack P Douglas, PhD<br>Vice President, Regulatory Affairs<br>VizaraMed, Inc.<br>1914 O'Toole Way<br>San Jose, CA 95131<br>(510) 792-7477 |
| 4. Contract Manufacturer Name | Novel Cath<br>(A Cirtec Company)<br>90 Great Oaks Blvd.<br>San Jose, CA 95119 |
| 5. Device Identification | Trade Name: Multiflex Steerable Sheath<br>Common Name: Introducer, Catheter<br>Classification Name: Catheter Introducer<br>Class: Class II, 21 CFR 870.1340<br>Product Code: DYB<br>Classification Panel: Cardiovascular |
# 510(k) SUMMARY
#### 6. Device Description
The VizaraMed Multiflex Steerable Sheath is a deflectable, sterile, single-use, percutaneous steerable sheath with dilator used to facilitate placement of various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
The sheath is available in two sizes and consists of a steerable shaft capable of threedimensional shapes, a handle with lockable steering controls, a Tuohy-Borst adapter with a hemostasis valve, a side port, and atraumatic soft distal tip. The shaft is radiopaque for visualization under fluoroscopy. The accompanying dilator is packaged with the sheath in a pouch for removal.
The 12.5F sheath can accommodate devices of 3F(0.039") - 12.5F (0.162"). The 15.5F sheath can accommodate devices sized 3F (0.039") - 15.5F (0.203").
{4}------------------------------------------------
- 7. Intended Use
The VizaraMed Multiflex Steerable Sheath is intended for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
- 8. Predicate Device
- St. Jude Medical Agilis Steerable Introducer [K081645]
- 9. Reference Device
- Cook, Inc. Extra Large Check-Flow Introducer [K203670] ●
- 10. Characteristics of Substantial Equivalence
The VizaraMed Multiflex Steerable Sheath is substantially equivalent to the predicate device based on the following:
- Indications for Use
- Specifications ●
- Product features ●
- Technological characteristics ●
- Product design
- Materials
- Sterilization method
- Mechanical testing
- Biocompatibility testing ●
The subject device was subjected to applicable testing to ensure reliable design and performance under testing parameters. The testing performed is described below:
{5}------------------------------------------------
## 11. Mechanical Properties
Mechanical testing, including dimensional verification, tensile testing, and leak testing, were conducted to ensure the Multiflex Steerable Sheath meets design requirements to support substantial equivalence to the predicate.
## 12. Sterilization and Packaging
Sterilization validation via ethylene oxide (EO) per ISO 11135 was conducted to ensure design requirements and to support substantial equivalence to the predicate. Sterilization validation used fractional and full cycles, and appropriate sample size of final preconditioned, packaged product that incorporated the use of biological indicators (BI) for the inoculated product (IP).
### 13. Usability Assessment
Subjective and objective evidence was obtained via questionnaire from physicians to assess critical tasks in a simulated heart model. Training was provided followed by user evaluation while manipulating the device.
### 14. Biocompatibility
The Vizaramed Sheath is considered to be a limited contact (≤24 hrs.), externally communicating device used in circulating blood. Testing was conducted for cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity and hemocompatibility. All tests were found to meet acceptance criteria, supporting substantial equivalence to the predicate.
#### 15. Conclusions
The results of the tests support the conclusion that the VizaraMed Steerable Sheath meets design input requirements based on the intended use and that it is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.