HQS INTRODUCER (Model 2064-HQS) is intended to be inserted in the vasculature to provide a conduit for the insertion of endovascular devices while minimizing blood loss associated with such insertions.
Device Story
HQS Introducer is a catheter introducer system comprising a radio-opaque introducer sheath with a haemostasis valve, a radio-opaque dilator, a centering wire device, and an extension line with a 3-way stopcock. Available in sizes 18F to 26F, it facilitates the insertion of large-caliber endovascular tools into the vasculature while maintaining a seal to minimize blood loss. The device is designed for one-handed operation of the sheath and adjustable valve. The centering device allows guidewire insertion through the valve while maintaining sealing. The extension line connects to the lateral port for fluid injection during procedures. Used by clinicians in clinical settings to provide vascular access for endovascular interventions.
Clinical Evidence
Bench testing only. In-vitro design verification testing confirmed compliance with ISO 11070:1999. Testing included tensile strength, overpressure/suction resistance, dimensional accuracy, valve sealing, kink resistance, and compatibility with vascular tools/guidewires. Biocompatibility testing per ISO 10993-1:2009 included cytotoxicity, sensitization, intracutaneous toxicity, systemic toxicity, pyrogenicity, and hemocompatibility (hemolysis, prothrombin time, coagulation, platelet, complement activation, and thrombogenicity).
Technological Characteristics
System includes radio-opaque introducer sheath, dilator, centering wire, and 3-way stopcock extension line. Sizes: 18F-26F. Materials meet ISO 10993-1:2009 requirements. Mechanical design features haemostasis valve for sealing. Sterilization and aging tests performed. No software or electronic components.
Indications for Use
Indicated for patients requiring endovascular device insertion via the vasculature; provides conduit while minimizing blood loss.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo with the text "AlSeel" written in a stylized, cursive font. The logo features a dark, oval shape with a white circle in the center, resembling a stylized eye or a lens. The text "AlSeel" is positioned below and to the right of the oval shape, with the letters slightly overlapping each other.
JUL 3 1 2012
113849
# 510(k) SUMMARY
As required by section 807.92(c)
| Submitter: | ALSEAL |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contacts: | M. DELFORGE Jean-François<br>VALPARC - ESPACE VALENTIN<br>6 C Rue De Franche-Comté<br>25048 BESANCON Cedex - France<br>Phone : 00333.81.61.69.93 |
| Preparation Date: | November 16, 2011 |
| Trade Name: | HQS Introducer (Model 2064-HQS) |
| Common Name: | Catheter Introducer |
| Classification Name: | Catheter Introducer |
| Regulation Number: | 870.1340 |
| Product Code: | DYB |
| Legally Marketed Predicate Devices: | Gore Dryseal Sheath (K093791) Manufactured By W.L. Gore &<br>Associates, Inc |
| Device Description: | The High Quality Sealing (HQS) Introducer comprises 4 elements, a radio-<br>opaque introducer sheath equipped with a haemostasis valve, a radio-<br>opaque dilator, a centering wire device and an extension line with 3 ways<br>stopcock.<br>The range of HQS introducer (18F, 20F, 22F, 24F, 26F) permits the<br>insertion, preserving sealing, of large caliber tools, from OF, up to the<br>nominal size of-the-introducer-sheath..<br>The introducer and its adjustable valve are easily handled with one hand.<br>The distal tip is designed to perform an efficient introduction of the HQS<br>introducer with the dilator in the vessel.<br>The centering device permits to introduce a guidewire through the valve<br>and to keep a perfect sealing.<br>The extension line is connected to the lateral port of the introducer for the<br>injection during the procedure. |
| Intended Use: | HQS INTRODUCER (Model 2064-HQS) is intended to be inserted in the<br>vasculature to provide a conduit for the insertion of endovascular devices<br>while minimizing blood loss associated with such insertions. |
. . . . . . .
..
·
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the text "At Seal" written in a stylized font. The logo also includes an oval shape with a white circle inside, positioned above the text. The logo appears to be for a company or brand named "At Seal."
11 3
| Bench/ Performance data | The following in-vitro testing was performed on the HQS Introducer<br>(Model 2064-HQS) in accordance with ISO standards and/or<br>internal procedures to assure reliable design and performance. In-<br>vitro design verification testing data demonstrate that the device is<br>in compliance with ISO 11070:1999 Sterile, single use<br>intravascular catheter introducers and product labeling. |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 1. Tensile strength |
| | 2. Overpressure resistance |
| | 3. Suction resistance |
| | 4. Introducer useful dimensions (lengths, Inside and Outside diameters) |
| | 5. Dilator useful dimensions (lengths, Inside and Outside diameters) |
| | 6. Sealing of the valve |
| | 7. Operating mechanism resistance |
| | 8. Visual control (atraumatic surface) |
| | 9. Resistance to kinking |
| | 10. Dilator compatibility (introduction / withdrawal) |
| | 11. Extension line connection |
| | 12. Centering device compatibility |
| | 13. Introducer/dilator distal transition |
| | 14. Compatibility with vascular tools and guidewire |
| | 15. Radiodetectability test |
| | 16. Packaging resistance |
| | 17. Sterilization tests<br>18. Aging tests |
| Non-Clinical Tests: | Biocompatibility tests<br>In accordance with ISO 10993-1:2009, the following<br>biocompatibility tests were conducted on the HQS Introducer<br>(Model 2064-HQS): |
| | • Cytotoxicity (ISO 10993-5:2009) |
| | • Sensitization (ISO 10993-10:2006) |
| | • Intracutaneous Toxicity (ISO 10993-10:2006) |
| | • Systemic Toxicity (ISO 10993-11:2006) |
| | • Pyrogenicity (ISO 10993-11:2006) |
| | • Hemocompatibility |
| | o Hemolysis (ASTM Guideline F756:2008) |
| | o Prothrombin Time (ISO 10993-4:2006) |
Page 18 / 66
.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
{2}------------------------------------------------
:
:
:
.
Image /page/2/Picture/1 description: The image shows a logo with a stylized oval shape in the upper left, and the text "AlSool" in a cursive font below the oval. The oval is tilted and appears to be a solid, dark color. The text "AlSool" is also in a dark color, contrasting with the background. The logo has a simple, modern design.
:
.
KII 3849 D. Zof 3
| | <ul><li>Coagulation UPTT (ISO 10993-4:2006)</li><li>Platelet (ISO 10993-4:2006)</li><li>Complement Activation (ISO 10993-4:2006)</li><li>Thrombogencity (ISO 10993-4:2006)</li></ul> Results for all biocompatibility testing demonstrate that the materials used meet the requirements of ISO 10993-1:2009. |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial equivalence: | HQS INTRODUCER is compared to predicate legally marketed device GORE DRYSEAL SHEATH (K093791). HQS INTRODUCER is substantially equivalent to its predicate devices in terms of intended use, function and technological characteristics. Any minor differences between these two devices do not raise new questions of safety and effectiveness. Performance data included within this submission demonstrates safety, effectiveness and substantial equivalence. |
| Conclusion: | The studies conducted on the HQS Introducer (Model 2064-HQS) demonstrate that the device is substantially equivalent to the predicate devices currently in commercial distribution. The proposed device meets the performance criteria of design verification as specified by ISO standards and test protocols. Any differences between the devices do not raise any significant issues of safety or effectiveness. |
:
:
:
:
and the comments of the comments of
:
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 3 1 2012
ALSEAL c/o Intertek Testing Services Ms. Paula Wilkerson 2307 E. Aurora Rd. Unit B7 Twinsburg, OH 44087
Re: K113849
Trade/Device Name: HQS Introducer (Model 2064 - HQS) Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: July 2, 2012 Received: July 10, 2012
Dear Ms. Wilkerson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 – Ms. Paula Wilkerson
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
CMR Hillen
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with the text "AtSeel" written in a stylized font. The logo also includes a circular graphic element above the text. The graphic is a solid black circle with a white oval shape cut out of the center. The text is in a cursive style and appears to be slightly tilted.
#### INDICATIONS FOR USE
K113849 510(k) Number (if known): _ Device Name: HQS INTRODUCER (Models 2064-HQS) Indications for Use:
The HQS introducer (Model 2064-HQS) is intended to be inserted in the vasculature to provide a conduit for the insertion of endovascular devices while minimizing blood loss associated with such insertions
Prescription Use | ★ (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Unz Keel
ardiovascular Devices
િની ડેક્કે 510(k) Number
Page 15 / 66
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.