K150186 · Boston Scientific Corp · DYB · Jun 5, 2015 · Cardiovascular
Device Facts
Record ID
K150186
Device Name
Chariot Guiding Sheath
Applicant
Boston Scientific Corp
Product Code
DYB · Cardiovascular
Decision Date
Jun 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Intended Use
The Chariot™ Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the peripheral vasculature.
Device Story
Chariot Guiding Sheath is a catheter introducer designed for peripheral vascular access. Device features a coiled shaft, hydrophilic coating, and radiopaque tantalum marker band for placement visualization. It includes a dilator, a hemostatic valve (cross-cut or Tuohy-Borst), and a sidearm with a three-way stopcock for flushing and contrast medium delivery. The device accommodates guidewires up to 0.038 inches. Used by clinicians in interventional settings to facilitate the delivery of diagnostic or therapeutic devices into the peripheral vasculature. The sheath provides a stable conduit, while the hemostatic valve prevents blood loss during procedure. The device is provided sterile and is intended for single use.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, hemolysis, thrombogenicity) and in-vitro performance (dilator entry profile, valve leakage, flow rates, burst pressure, tensile strength, kink resistance, torque strength, and coating integrity).
Technological Characteristics
Coiled shaft design; polymer construction with PTFE inner liner; stainless steel coil; hydrophilic coating; tantalum radiopaque marker; copolyester elastomer hub; luer fitting; ethylene oxide sterilization. Dimensions accommodate guidewires ≤ 0.038 in. Maximum infusion pressure 309 psi.
Indications for Use
Indicated for the introduction of interventional and diagnostic devices into the peripheral vasculature.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES - USA
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 5, 2015
Boston Scientific Corp. Kurtis Hunsberger Principal Regulatory Affairs Specialist One Scimed Place Maple Grove, Minnesota 55311
Re: K150186
> Trade/Device Name: Chariot Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: May 6, 2015 Received: May 7, 2015
Dear Mr. Hunsberger,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hilleman
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150186
Device Name Chariot™ Guiding Sheath
Indications for Use (Describe)
The Chariot™ Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the peripheral vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
per 21 CFR §807.92
| Submitter's<br>Name and<br>Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311<br>Phone: 763-494-1700<br>Fax: 763-494-2222 | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|--------------|
| Contact Name<br>and<br>Information | Kurtis Hunsberger<br>Principal Regulatory Affairs Specialist<br>Phone: 763-494-1204<br>Fax: 763-494-2222<br>e-mail: kurtis.hunsberger@bsci.com | | |
| Date Prepared | April 23, 2015 | | |
| Proprietary<br>Name | Chariot™ Guiding Sheath | | |
| Common Name | Guiding Sheath | | |
| Product Code | DYB | | |
| Classification | Class II, 21 CFR Part 870.1340 | | |
| Predicate<br>Device | Terumo Pinnacle® Destination®<br>Peripheral Guiding Sheath | K091329 | May 29, 2009 |
| Device<br>Description | The Chariot™ Guiding Sheath is designed to perform as an introducer<br>sheath for delivering interventional and diagnostic devices into the<br>peripheral vasculature. The guiding sheath has a coiled shaft design and<br>comes with a straight (ST) or preformed multipurpose (MP) tip shape. It is<br>equipped with a cross-cut hemostatic valve or Tuohy-Borst adapter to<br>prevent bleeding and a sidearm with a three-way stopcock to allow for<br>flushing and introduction of contrast medium. It is also packaged with a<br>dilator to facilitate delivery over a guidewire. The guiding sheath can<br>accommodate guidewires with diameters less than or equal to 0.038 in<br>(0.97 mm). The outer surface of the guiding sheath has a hydrophilic<br>coating from the distal tip to approximately 9 cm from the hub. The distal<br>tip has a radiopaque marker band approximately 6 mm from the distal<br>edge, to help with guiding sheath placement.<br>A copolyester elastomer hub is over molded onto the proximal section of<br>the guiding sheath. It incorporates a luer fitting which serves as a junction | | |
| Intended Use /<br>Indications for<br>Use | to the hemostatic valve.<br>The Chariot™ Guiding Sheath is intended for the introduction of<br>interventional and diagnostic devices into the peripheral vasculature. | | |
{4}------------------------------------------------
| Comparison of<br>Technological<br>Characteristics | The Chariot™ Guiding Sheath incorporates substantially equivalent<br>device materials and design, packaging materials and design,<br>fundamental technology, manufacturing processes, sterilization process<br>and intended use as those featured in the Terumo Pinnacle®<br>Destination® Peripheral Guiding Sheath K091329 cleared May 29, 2009.<br>Similarities and differences in technological characteristics between the<br>predicate and subject device are listed below.<br>Similarities:<br>• Polymer material construction<br>• Stainless steel coil<br>• PTFE inner liner material<br>• Hydrophilic coating<br>• Radiopaque marker<br>• Sheath dimensions<br>• Dilator, Tuohy-Borst Valve, and Cross-Cut Valve accessories<br>• Ethylene Oxide sterilization<br>• Packaging design with same function<br>Differences:<br>• Shaft color: Blue (Chariot); Green (predicate)<br>• Radiopaque marker material: Tantalum (Chariot); Gold (predicate)<br>• Maximum infusion pressure: 309 psi (Chariot); not labeled (predicate) | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Performance<br>Data | The Chariot™ Guiding Sheath was subjected to testing according to the<br>requirements of Guidance for Industry and FDA Staff – Class II Special<br>Controls for Certain Percutaneous Transluminal Coronary Angioplasty<br>(PTCA) Catheters, September 8, 2010. Bench testing and<br>biocompatibility testing were performed to support a determination of<br>substantial equivalence. The results of these tests provide reasonable<br>assurance that the proposed device has been designed and tested to<br>assure conformance to the requirements for its intended use. No new<br>safety or performance issues were raised during the testing; therefore,<br>this device may be considered substantially equivalent to the predicate<br>device.<br><br>The following biocompatibility and chemical characterization tests were<br>completed on the Chariot™ Guiding Sheath and its accessories: | |
| Cytotoxicity | Hemolysis (Extract Method) | |
| Sensitization | Partial Thromboplastin Time | |
| Intracutaneous Reactivity | In Vitro Hemocompatibility | |
| Acute Systemic Toxicity | Complement Activation | |
| Materials Mediated Pyrogenicity | In Vivo Thromboresistance | |
| Hemolysis (Direct Contact) | USP <661> Physicochemical | |
| | The following in-vitro performance tests were completed on the Chariot™ Guiding Sheath: | |
| | Dilator Entry Profile | Valve Leakage |
| | Sheath Length | Dye Flow Rate |
| | Sheath Inner and Outer Diameter | Sheath Burst Pressure |
| | Dilator Inner and Outer Diameter | Device Visual Appearance |
| | Dilator Length | Radiopacity |
| | Sheath Tensile | Sheath Kink Resistance |
| | Sheath to Hub Tensile | Torque Strength |
| | Dilator to Hub Tensile | Particulates |
| | Sheath Tip Tensile | Coating Integrity |
| | Hub Function and Luer Compatibility | |
| Conclusion | Based on the indications for use, technological characteristics, and safety and performance testing, the Chariot™ Guiding Sheath has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Terumo Pinnacle Destination Peripheral Guiding Sheath. | |
{5}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.