Edwards eSheath Introducer Set

K152225 · Edwards Lifesciences · DYB · Nov 24, 2015 · Cardiovascular

Device Facts

Record IDK152225
Device NameEdwards eSheath Introducer Set
ApplicantEdwards Lifesciences
Product CodeDYB · Cardiovascular
Decision DateNov 24, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

Entry of interventional devices into the vascular system

Device Story

The Edwards eSheath Introducer Set facilitates vascular access for the Edwards SAPIEN 3 Transcatheter Heart Valve and Commander Delivery System. The system includes a sheath and two introducers. The sheath features a dual-layer shaft (HDPE/TecoFlex and PTFE liner) with a folded seam design, allowing temporary expansion during device insertion. A proximal housing contains three hemostasis valves (duckbill, disc, and cross-slit) and a flushport. Radiopaque markers and a hydrophilic coating assist in navigation and placement. Two radiopaque introducers with tapered tips and guidewire lumens aid vessel entry and dilation. Used by physicians in clinical settings, the device provides a stable conduit for heart valve delivery. The expandable design minimizes the required sheath profile, potentially reducing vascular trauma. The healthcare provider monitors the sheath tip via radiopacity and manages hemostasis through the integrated valve housing.

Clinical Evidence

Bench testing only. No clinical data presented. Testing included visual inspection, dimensional analysis, radiopacity, guidewire compatibility, hemostasis, lubricity, durability, kink resistance, seam return after expansion, bond strength, device interaction, hydrophilic coating characterization, USP particulate testing, sterilization validation, and comprehensive biocompatibility (cytotoxicity, hemocompatibility, systemic toxicity, pyrogenicity, irritation, sensitization, chemical acceptability, and thrombogenicity).

Technological Characteristics

Sheath shaft: HDPE/TecoFlex coextruded outer layer, PTFE liner. Expandable distal region via folded seam. Hemostasis: 3-valve housing (duckbill, disc, cross-slit). Radiopaque markers. Hydrophilic coating. Introducers: radiopaque, tapered tip, guidewire lumen. Available in 14F and 16F diameters. Sterilization: validated method.

Indications for Use

Indicated for the introduction and removal of devices used with the Edwards SAPIEN 3 Transcatheter Heart Valve.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus-like symbol with three figures in profile, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion. The seal is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 24, 2015 Edwards Lifesciences Karen Reynolds Regulatory Affairs Associate III 1 Edwards Way Irvine, California 92614-5686 Re: K152225 Trade/Device Name: Edwards eSheath Introducer Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: October 26, 2015 Received: October 27, 2015 Dear Karen Reynolds: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. M.R. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K152225 Device Name Edwards eSheath Introducer Set Indications for Use (Describe) The Edwards eSheath Introducer Set is indicated for the introduction and removal of devices used with the Edwards SAPIEN 3 Transcatheter Heart Valve. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary | Submitter: | Edwards Lifesciences, LLC<br>One Edwards Way<br>Irvine, CA 92663 | |-----------------------|--------------------------------------------------------------------------------------------------------------| | Contact: | Karen Reynolds Phone: 949-250-2500, Fax: 949-756-4408 | | Prepared: | August 5, 2015 | | Trade Name: | Edwards eSheath Introducer Set | | Common Name: | Catheter, Introducer | | Classification: | Catheter Introducer<br>21 CFR 870.1340, Product Code DYB | | Predicate<br>Devices: | RetroFlex 3 Introducer Sheath Set (K093877)<br>Solopath Balloon Expandable Transfemoral Introducer (K100819) | ## Device Description: The Edwards eSheath Introducer Set consists of a sheath and 2 introducers. It is available with inner sheath diameters of 14 French (model 914ES) and 16 French (model 916ES). The 14 French introducer set is used to facilitate introduction of the 23mm and 26mm SAPIEN 3 THV and Commander Delivery System into the vasculature, and the 16 French introducer set is used to facilitate introduction of the 29mm SAPIEN 3 THV and Commander Delivery System into the vasculature. The sheath shaft is comprised of 2 layers of material (HDPE/TecoFlex coextruded outer layer and PTFE liner). The outer and inner layer are folded, creating a seam which allows the distal region of the sheath to temporarily expand in diameter when a device is inserted. A tapered strain relief at the proximal end of the sheath shaft helps provide hemostasis when the sheath is inserted. A radiopaque marker on the distal end indicates the location of the sheath tip in the body and a hydrophilic coating on the sheath tubing exterior facilitates introduction into the vessel. The sheath tubing mates with a housing, which holds three seals (valves) to provide hemostasis: a duckbill seal, a disc seal, and a cross slit seal. The housing also includes flushport tubing and a stopcock to allow for flushing of the sheath. Two introducers are provided to aid in the introduction of the target vessel, and can also be used for dilation of the vessel prior to sheath insertion. The introducers are radiopaque and feature a tapered tip and guidewire lumen. ## Intended Use: Entry of interventional devices into the vascular system {4}------------------------------------------------ #### Indication: The Edwards eSheath Introducer Set is indicated for the introduction and removal of devices used with the Edwards SAPIEN 3 Transcatheter Heart Valve. #### Comparison to Predicate: The Edwards eSheath Introducer Set (models 914ES) is substantially equivalent in function, performance, and design to the RetroFlex 3 Introducer Sheath Set (K093877), Solopath Balloon Expandable Transfemoral Introducer (K100819), and Cook Medical Check-Flo Performer Extra Large Introducer (K142829). The Edwards eSheath and the Solopath Introducer have an expandable shaft. The Edwards eSheath is expanded in the vasculature by the device that is passed through the inner diameter and the Solopath Introducer is expanded in the vasculature via inflation. The eSheath device compatibility testing was completed using the Edwards SAPIEN 3 Transcatheter Heart Valve and Commander delivery system (commercially available per PMA P140031). The RetroFlex 3 Introducer Sheath Set (K093877) has the same proximal end (housing, hemostasis control, and flush tube) and introducer as the Edwards eSheath but does not include the expansion feature. #### Summary of Non-Clinical Testing: Non-clinical testing was completed to demonstrate that the performance characteristics of the Edwards eSheath Introducer Set are equivalent to the predicates, and to verify that design requirements are satisfied. Specifically, the following design verification and validation testing was successfully completed: - . Visual Surface Inspection - Dimensional Inspection . - . Radiopacity/Visualization - Guidewire Compatibility - . Hemostasis - . Lubricity and Durability of the Sheath - Kink Resistance ● - Seam Return After Expansion - Bond Strength ● - . Device Interaction - Hydrophilic Coating Characterization . - . USP Particulate Test - . Sterilization Validation - Biocompatibility Tests: . - . Cytotoxicity - Hemocompatibility ■ - I Systemic Toxicity - I Material Mediated Pyrogenicity - I Irritation/Intracutaneous Reactivity - Sensitization I - 트 Chemical Acceptability - Thrombogenicity - Packaging Integrity - Shelf Life Verification ● ## Conclusion: Based upon device testing and descriptive characteristics, the Edwards eSheath Introducer Set is substantially equivalent to the predicate device and performance testing has demonstrated that safety and efficacy are not adversely impacted.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...