RETROFLEX 3 INTRODUCER SHEATH SET, MODELS 9120S23 AND 9120S26
Applicant
Edwards Lifesciences, LLC
Product Code
DYB · Cardiovascular
Decision Date
Jul 1, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The RetroFlex 3™ Introducer Sheath Set is intended for entry of interventional devices into the vascular system.
Device Story
Hydrophilic delivery system for percutaneous introduction of intravascular devices; used over 0.035" guidewire. System comprises sheath, introducer (dilator), and loader. Sheath contains three valves (duckbill, cross-slit, disc) for hemostasis; includes three-way stopcock for fluid infusion/aspiration. Loader facilitates entry of larger devices by bypassing sheath valves; loader includes its own disc valve. Used in clinical settings by healthcare professionals to provide vascular access for interventional procedures. Single-use, sterile device.
Clinical Evidence
Bench testing only. Verification included visual examination, dimensional verification, tensile strength, guidewire compatibility, hydrophilic coating testing, packaging integrity, and sterilization validation. Biocompatibility testing included MEM, Agar Overlay, blood compatibility, mouse systemic injection, rabbit pyrogen, rabbit intracutaneous irritation, guinea pig maximization, complement activation, and porcine thrombogenicity. Chemical testing included material verification and USP physico-chemical tests.
Technological Characteristics
Hydrophilic-coated catheter introducer system. Includes sheath, dilator, and loader. Features three-valve hemostasis system (duckbill, cross-slit, disc) and three-way stopcock. Available in 25F and 28F outer diameters. Tested per ISO 10555-1:1997. Sterile, single-use.
Indications for Use
Indicated for the percutaneous introduction of interventional devices into the vascular system. No specific patient population, age, or gender restrictions are stated.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{1}------------------------------------------------
K093817
# 510(k) Summary
JUL· - 1 - 2010
Applicant:
Edwards Lifesciences, LLC One Edwards Way Irvine, CA 92614 USA Phone: 949.250-3837 Fax: 949.756-4408
Date:
May 21, 2010
Contact Persons:
Jason K. Lyon Principal Project Manager, Regulatory Affairs
Irene Parker Sr. Director of Regulatory Affairs
Proprietary Device Name:
RetroFlex 3™ Introducer Sheath Set
Common Device Name:
Catheter Introducer (21 CFR 870.1340, Product Code DYB), and;
Classification:
Predicate Devices:
Class II
Edwards Lifesciences, LLC - Edwards Introducer Sheath cleared under K031087
Guidant EndoVascular Surgery Group -ANCURE® Sheath cleared under K003889
Manufacturer:
Edwards Lifesciences, LLC One Edwards Way Irvine, CA 92614 USA
{2}------------------------------------------------
### Substantially Equivalent To:
The RetroFlex 3 Introducer Sheath Set was evaluated for in-vitro performance and biocompatibility tests, which are outlined below. The physical characteristics and mode of use of the RetroFlex 3 Introducer Sheath is similar to the Edwards Introducer Sheath and the Ancure® Sheath, and are common in mode of operation, intended use, and dimensions. The devices have only slight variations to technological design and materials. These differences, however, do not alter the fundamental use of the products, which is to facilitate entry into the arterial vasculature.
Based on the data presented in this 510(k) Premarket Notification, the RetroFlex 3 Introducer Sheath Set is substantially equivalent to the predicate devices.
## Description of the Device Subject to Premarket Notification:
The RetroFlex 3 Introducer Sheath Set is a hydrophilic delivery system used over a compatible 0.035" guidewire for the percutaneous introduction of intravascular devices. The RetroFlex 3 houses three (3) valves (a duckbill valve, a cross-slit valve, and a disc valve) to achieve hemostasis and is fitted with a three-way stopcock to allow infusion or aspiration of fluid. Each RetroFlex 3 Introducer Sheath Set contains a sheath, an introducer (or dilator), and a loader. The loader enables a device to enter into the sheath by bypassing all of the three valves within the sheath. The loader also contains a disc valve for hemostasis. The loader allows for larger devices to enter into the vascular system.
The RetroFlex 3 Introducer Sheath is for one-time use only, sold and packaged sterile. The table below identifies the model and sheath diameters. (Table 7.2).
| Model No. | Sheath O.D. | Sheath I.D. |
|-----------|-------------|-------------|
| 9120S23 | 25F | 22F |
| 9120S26 | 28F | 24F |
Table 7.2 - RetroFlex 3™ Introducer Sheaths
#### Indications For Use: 7.3
The RetroFlex 3 Introducer Sheath is intended for entry of interventional devices into the vascular system.
#### Performance Data: 7.4
The RetroFlex 3 Introducer Sheath was verified and tested according to performance testing standards ISO 10555-1:1997, Sec 4.5, Sterile Single Use Intravascular Catheters. The following tests have been conducted to demonstrate substantial equivalence to the predicate devices with respect to intended use, design, materials, and performance. The following non-clinical tests and biocompatibility tests were performed:
- . Visual examination
- Dimensional verification
7.2
{3}------------------------------------------------
- Tensile strength .
- Guidewire compatibility .
- Hydrophilic coating test .
- . Packaging Integrity
- . Sterilization Validation
- Biocompatibility .
- Medium Eluate Method (MEM) o
- Agar Overlay Method (AO) o
- Blood Compatibility Test Method o
- Mouse Systemic Injection O
- Rabbit Pyrogen (Chemical-Mediated) Test o
- Rabbit Intracutaneous Irritation o
- Guinea Pig Maximization Test o
- Complement Activation Test o
- Thrombogenicity (porcine model) o
- Chemical .
- Material Verification o
- USP Physico-Chemical Test o
#### 7.5 Conclusion:
Based upon the non-clinical testing noted above and the data presented in the 510(k) Premarket Notification, the RetroFlex 3™ Introducer Sheath Set has met the requirements threshold to establish substantial equivalence to the predicate devices for market clearance.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a series of interconnected lines, possibly representing human figures or abstract shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL - 1 2010
Edwards Lifesciences, LLC c/o Mr. Jason K. Lyon Principal Project Manager, Regulatory Affairs One Edwards Way Irvine, CA 92614
Re: K093877
Trade/Device Name: RetroFlex 3TM Introducer Sheath Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II (two) Product Code: DYB Dated: June 25, 2010 Received: June 28, 2010
Dear Mr. Lyon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act (117). general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbrandijng and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devine can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Jason K. Lyon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing
practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
e. h. hmer
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Traditional 510(k) - RetroFlex 3™ Introducer Sheath Set Edwards Lifesciences
## Statement of Indications for Use
Indications for Use
510(k) Number (if known): K093877
Device Name: RetroFlex 3™ Introducer Sheath Set
The RetroFlex 3™ Introducer Sheath Set is intended for entry of interventional devices into the vascular system.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Divina R. Valenzuela
.
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number Ko 93 877
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.