iSLEEVE Introducer Set

K180785 · Boston Scientific Corporation · DYB · Jun 22, 2018 · Cardiovascular

Device Facts

Record IDK180785
Device NameiSLEEVE Introducer Set
ApplicantBoston Scientific Corporation
Product CodeDYB · Cardiovascular
Decision DateJun 22, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system.

Device Story

Sterile, single-use introducer catheter; provides percutaneous femoral artery access for valve system delivery. Comprises dilator, introducer sheath with three-way stopcock, product mandrel, and flushing tube. Sheath features expandable distal end for transient expansion during valve introduction; hydrophilic coating enhances lubricity. Used by clinicians in vascular procedures. Device facilitates valve system introduction/removal; expandable design minimizes vessel expansion time. Benefits patient by providing stable, lubricated access path for transcatheter valve delivery.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing (dimensional verification, deployment/retraction, bond strength, flexibility, torque, radiopacity, coating integrity, particulate, burst pressure, insertion force, valve leakage, package integrity, sterility, biocompatibility, endotoxin) and acute animal assessment in simulated use environment.

Technological Characteristics

Sterile, single-use intravascular introducer. Materials conform to EN ISO 10993-1 (biocompatibility). Standards: EN ISO 10555-1 (general requirements), EN ISO 11070 (introducers/dilators). Features hydrophilic coating, expandable distal sheath, three-way stopcock. Sizes: 14F and 15F. Sterilization: Ethylene Oxide (EO).

Indications for Use

Indicated for femoral vascular access to introduce/remove ACURATE TF Valve System (14F, vessels ≥ 5.5 mm) or LOTUS Edge Valve System (15F, vessels ≥ 5.9 mm) and ancillary devices.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 22, 2018 Boston Scientific Corporation Toni Guidarelli Sr. Regulatory Affairs Specialist Three Scimed Place Maple Grove, Minnesota 55446 Re: K180785 Trade/Device Name: iSLEEVE™ Introducer Set Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: March 23, 2018 Received: March 26, 2018 Dear Ms. Guidarelli: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Jaime Raben -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K180785 Device Name iSLEEVE Introducer Set #### Indications for Use (Describe) #### 14F iSLEEVE Introducer Set: The 14F iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of the ACURATE TF Valve System and ancillary devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.5 mm. #### 15F iSLEEVE Introducer Set: The 15F iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of the LOTUS Edge Valve System and ancillary devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.9 mm. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary #### per 21 CFR §807.92 | Sponsor | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA | | | |---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------| | Contact Name and<br>Information | Toni Guidarelli<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-2870<br>Fax: 763-494-2222<br>e-mail: Toni.Guidarelli@bsci.com | | | | Prepared | March 23, 2018 | | | | Proprietary Name | iSLEEVE TM Introducer Set | | | | Common Name | Catheter introducer | | | | Product Code | DYB | | | | Classification | Class II, 21 CFR Part 870.1340 | | | | Primary Predicate<br>Device | Creganna-Tactx<br>LotusTM Introducer Set | K140338 | 27 August 2014 | | Reference<br>Device(s) | Edwards eSheath<br>Introducer Set | K152225 | 24 November 2015 | #### Device Description The iSLEEVE™ Introducer Set is a sterile, single-use introducer catheter that provides percutaneous access to the femoral artery for introduction and removal of compatible valve systems and ancillary devices into the vascular system. The iSLEEVE Introducer Set is composed of a dilator and an introducer sheath with a three-way stopcock. The distal end of the introducer sheath is expandable which allows for transient sheath expansion when the compatible valve system is introduced through it. This reduces the time the access vessel is expanded. There is a hydrophilic coating on the introducer sheath which increases the lubricity of the surface to aid in delivery when activated. The iSLEEVE Introducer Set is available in two sizes; 14F and 15F. The 14F iSLEEVE Introducer Set is for use with the ACURATE TF Valve System only and the 15F iSLEEVE Introducer Set is for use with the LOTUS Edge™ Valve System only. A product mandrel and flushing tube are also supplied with the iSLEEVE Introducer Set. {4}------------------------------------------------ #### Indications for Use / Intended Use Guide Wire #### 14F iSLEEVE Introducer Set: #### INTENDED USE: The iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system. #### INDICATIONS FOR USE: The 14F iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of the ACURATE TF Valve System and ancillary devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.5 mm diameter. #### 15F iSLEEVE Introducer Set: #### INTENDED USE: The iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system. #### INDICATIONS FOR USE: The 15F iSLEEVE Introducer Set is intended to facilitate femoral access to the vascular system for introduction and removal of the LOTUS Edge™ Valve System and ancillary devices. The iSLEEVE Introducer Set is suitable for use in vessels ≥ 5.9 mm. #### Comparison of Technological Characteristics The iSLEEVE™ Introducer Set incorporates substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate, Creganna-Tactx Lotus™ Introducer Set K140338. {5}------------------------------------------------ #### Performance Data The iSLEEVE Introducer Set was subjected to testing to support a determination of substantial equivalence according to Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010. The iSLEEVE Introducer Set also conforms to relevant sections of EN ISO 10555-1: 2013, Sterile, single-use intravascular catheters; Part 1: General Requirements and EN ISO 11070:2014 Sterile single-use intravascular introducers, dilators and quidewires. In addition biocompatibility testing was performed in accordance with EN ISO 10993-1 Biological evaluation of medical devices. A summary of testing completed is provided below; - Dimensional Verification ● - Device Deployment and Retraction ● - Catheter Bond Strength ● - Flexibility and Kink Test - Torque Strength - . Radiopacity - . Coating Integrity - Particulate Evaluation ● - Catheter Body Burst Pressure ● - System Insertion Force ● - Valve Leakage - Visual Appearance - Package Integrity - Sterility and EO Residuals . - Biocompatibility ● - Endotoxin ● In addition to the aforementioned testing, a preclinical study was executed to support a determination of substantial equivalence. The study evaluated the performance of the iSLEEVE device through an acute animal assessment in a simulated use environment and bench testing under simulated use conditions. As part of the acute animal assessment Tip and Dilator radiopacity performance was assessed. The results of the tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing; therefore this device may be considered substantially equivalent to the predicate device. #### Clinical Testing Performance testing from clinical studies is not required to demonstrate substantial equivalence of the iSLEEVE Introducer Set. #### Conclusion Based on the indications for use, technological characteristics, and performance testing the iSLEEVE™ Introducer Set has been shown to be appropriate for its' intended use and is considered to be substantially equivalent to the Creganna-Tactx Lotus™ Introducer Set K140338.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...